Stability of Prototype Two-Piece Zirconia and Titanium Implants after Artificial Aging: An In Vitro Pilot Study

被引:54
|
作者
Kohal, Ralf-Joachim [1 ]
Finke, Hans Christian [1 ]
Klaus, Gerold [1 ]
机构
[1] Univ Freiburg, Sch Dent, Dept Prosthodont, Univ Clin Freiburg, D-79106 Freiburg, Germany
关键词
all-ceramic crowns; artificial mouth; fracture strength; two-piece zirconia implant; LOW-TEMPERATURE DEGRADATION; COMMERCIALLY PURE TITANIUM; ALL-CERAMIC RESTORATIONS; FRACTURE-RESISTANCE; FAILURE MODE; INTERFACE; STRENGTH; ALUMINA; TISSUES; SURFACE;
D O I
10.1111/j.1708-8208.2008.00116.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background: Zirconia oral implants are a new topic in implant dentistry. So far, no data are available on the biomechanical behavior of two-piece zirconia implants. Therefore, the purpose of this pilot investigation was to test in vitro the fracture strength of two-piece cylindrical zirconia implants after aging in a chewing simulator. Materials and Methods: This laboratory in vitro investigation comprised three different treatment groups. Each group consisted of 16 specimens. In group 1, two-piece zirconia implants were restored with zirconia crowns (zirconia copings veneered with Triceram (R); Esprident, Ispringen, Germany), and in group 2 zirconia implants received Empress (R) 2 single crowns (Ivoclar Vivadent AG, Schaan, Liechtenstein). The implants, including the abutments, in the two zirconia groups were identical. In group 3, similar titanium implants were reconstructed with porcelain-fused-to-metal crowns. Eight samples of each group were submitted to artificial aging with a long-term load test in the artificial mouth (chewing simulator). Subsequently, all not artificially aged samples and all artificially aged samples that survived the long-term loading of each group were submitted to a fracture strength test in a universal testing machine. For the pairwise comparisons in the different test groups with or without artificial loading and between the different groups at a given artificial loading condition, the Wilcoxon rank-sum test for independent samples was used. The significance level was set at 5%. Results: One sample of group 1 (veneer fracture), none of group 2, and six samples of group 3 (implant abutment screw fractures) failed while exposed to the artificial mouth. The values for the fracture strength after artificial loading with 1.2 million cycles for group 1 were between 45 and 377 N (mean: 275.7 N), in group 2 between 240 and 314 N (mean: 280.7 N), and in the titanium group between 45 and 582 N (mean: 165.7 N). The fracture strength results without artificial load for group 1 amounted to between 270 and 393 N (mean: 325.1 N), for group 2 between 235 and 321 N (mean: 281.8 N), and between 474 and 765 N (mean: 595.2 N) for the titanium group. The failure mode during the fracture testing in the zirconia implant groups was a fracture of the implant head and a bending/fracture of the abutment screw in the titanium group. Conclusions: Within the limits of this pilot investigation, the biomechanical stability of all tested prototype implant groups seems to be - compared with the possibly exerted occlusal forces - borderline for clinical use. A high number of failures occurred already during the artificial loading in the titanium group at the abutment screw level. The zirconia implant groups showed irreparable implant head fractures at relatively low fracture loads. Therefore, the clinical use of the presented prototype implants has to be questioned.
引用
收藏
页码:323 / 329
页数:7
相关论文
共 50 条
  • [1] Stability of screw-retention in two-piece zirconia implants: An in vitro study
    Joos, Marc
    Sailer, Irena
    Filippi, Andreas
    Mukaddam, Khaled
    Rosentritt, Martin
    Kuhl, Sebastian
    CLINICAL ORAL IMPLANTS RESEARCH, 2020, 31 (07) : 607 - 614
  • [2] Two-piece zirconia versus titanium implants after 80 months: Clinical outcomes from a prospective randomized pilot trial
    Koller, Martin
    Steyer, Elisabeth
    Theisen, Kerstin
    Stagnell, Sami
    Jakse, Norbert
    Payer, Michael
    CLINICAL ORAL IMPLANTS RESEARCH, 2020, 31 (04) : 388 - 396
  • [3] Mechanical Stability of Restorations Supported by Titanium Base, Zirconia, and Polyetherketoneketone Abutments on One- and Two-Piece Zirconia Implants
    Cantarella, Josef
    Pitta, Joao
    Mojon, Philippe
    Hicklin, Stefan P.
    Fehmer, Vincent
    Sailer, Irena
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2021, 36 (02) : 313 - 321
  • [4] Fracture Resistance of a Two-Piece Zirconia Implant System after Artificial Loading and/or Hydrothermal Aging-An In Vitro Investigation
    Kohal, Ralf-Joachim
    Schikofski, Tim
    Adolfsson, Erik
    Vach, Kirstin
    Patzelt, Sebastian Berthold Maximilian
    Nold, Julian
    Wemken, Gregor
    JOURNAL OF FUNCTIONAL BIOMATERIALS, 2023, 14 (12)
  • [5] Two-piece zirconia oral implants withstand masticatory loads: An investigation in the artificial mouth
    Spies, Benedikt Christopher
    Nold, Julian
    Vach, Kirstin
    Kohal, Ralf-Joachim
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 53 : 1 - 10
  • [6] Mechanical stability of all-ceramic abutments retained with three different screw materials in two-piece zirconia implants—an in vitro study
    Michael Stimmelmayr
    Anna Lang
    Florian Beuer
    Sonia Mansour
    Kurt Erdelt
    Gerald Krennmair
    Jan-Frederik Güth
    Clinical Oral Investigations, 2020, 24 : 1801 - 1806
  • [7] Pilot Clinical and Histologic Evaluations of a Two-Piece Zirconia Implant
    Nevins, Myron
    Camelo, Marcelo
    Nevins, Marc L.
    Schupbach, Peter
    Kim, David M.
    INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2011, 31 (02) : 157 - 163
  • [8] Bone response to functionally loaded, two-piece zirconia implants: A preclinical histometric study
    Janner, Simone F. M.
    Gahlert, Michael
    Bosshardt, Dieter D.
    Roehling, Stefan
    Milz, Stefan
    Higginbottom, Frank
    Buser, Daniel
    Cochran, David L.
    CLINICAL ORAL IMPLANTS RESEARCH, 2018, 29 (03) : 277 - 289
  • [9] Success Rate of Two-Piece Zirconia Implants: A Retrospective Statistical Analysis
    Jank, Siegfried
    Hochgatterer, Gregor
    IMPLANT DENTISTRY, 2016, 25 (02) : 193 - 198
  • [10] Mechanical stability of all-ceramic abutments retained with three different screw materials in two-piece zirconia implants-an in vitro study
    Stimmelmayr, Michael
    Lang, Anna
    Beuer, Florian
    Mansour, Sonia
    Erdelt, Kurt
    Krennmair, Gerald
    Gueth, Jan-Frederik
    CLINICAL ORAL INVESTIGATIONS, 2020, 24 (05) : 1801 - 1806