Asymmetric speed modulation of a rotary blood pump affects ventricular unloading

被引:42
|
作者
Pirbodaghi, Tohid [1 ]
Weber, Alberto [2 ]
Axiak, Shannon [3 ]
Carrel, Thierry [2 ]
Vandenberghe, Stijn [1 ]
机构
[1] Univ Bern, ARTORG Cardiovasc Engn, CH-3010 Bern, Switzerland
[2] Univ Hosp Bern, Inselspital, Dept Cardiac Surg, CH-3010 Bern, Switzerland
[3] Univ Bern, Dept Vet Anesthesia, CH-3010 Bern, Switzerland
基金
瑞士国家科学基金会;
关键词
Rotary blood pumps; Speed modulation; Systole and diastole; Pulsatile flow; In vivo study; ASSIST DEVICE SUPPORT; PULSATILE; FLOW; NONPULSATILE; PERFUSION; RECOVERY; FAILURE;
D O I
10.1093/ejcts/ezs299
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES: Rotary blood pumps (RBPs) running at a constant speed are routinely used for the mechanical support of the heart in various clinical applications, from short-term use in heart-lung machines to long-term support of a failing heart. Their operating range is delineated by suction and regurgitation events, leaving limited control on the cardiac workload. This study investigates whether different ratios of systolic/diastolic support are advantageous over a constant-speed operation. METHODS: In order to effectively control the load on the heart, this study aimed at developing a pulsatile control algorithm for rotary pumps to investigate the impact of pump speed modulation during systole and diastole on the left ventricle unloading. The CentriMag (TM) RBP with a modified controller was implanted in four sheep via a left thoracotomy and cannulated from the ventricular apex to the descending aorta. To modulate the pump speed synchronized with the heartbeat, custom-made real-time software detected the QRS complex of the electrocardiogram and controlled the pump speed during systole and diastole. Four different speed modulations with the same average speed but different systolic and diastolic speeds were compared with the baseline and the constant speed support. Left ventricular (LV) pressure and volume, coronary flow and pump flow were analysed to examine the influence of the pump speed modulation. RESULTS: Pulsatile setting reduces the cardiac workload to 64% of the baseline and 72% of the constant speed value. Maximum unloading is obtained with the highest speed during diastole and high-pulse amplitude. End-diastolic volume in the pulsatile modes varied from 85 to 94% of the baseline and 96 to 107% of the constant speed value. Consequently, the mechanical load on the heart can be adjusted to provide assuagement, which may lead to myocardial recovery. The higher pump speed during systole results in an increase in the pulse pressure up to 140% compared with the constant speed. CONCLUSIONS: The present study is an initial step to more accurate speed modulation of RBPs to optimize the cardiac load control. To develop future control algorithms, the concept of high speed during diastole having a maximal unloading effect on the LV and high speed during systole increasing the pulse pressure is worth considering.
引用
收藏
页码:383 / 388
页数:6
相关论文
共 50 条
  • [21] Modeling of a Rotary Blood Pump
    Nestler, Frank
    Bradley, Andrew P.
    Wilson, Stephen J.
    Timms, Daniel L.
    ARTIFICIAL ORGANS, 2014, 38 (03) : 182 - 190
  • [22] Flow Field Analysis of a Centrifugal Rotary Blood Pump under Active Speed Modulations
    Huang F.
    Wang C.
    Ruan X.
    1600, Chinese Mechanical Engineering Society (32): : 2915 - 2923
  • [23] Physiologic outcome of varying speed rotary blood pump support algorithms: a review study
    Bozkurt, Selim
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (01) : 13 - 28
  • [24] Detection of Ventricular Suction in an Implantable Rotary Blood Pump Using Support Vector Machines
    Wang, Yu
    Faragallah, George
    Divo, Eduardo
    Simaan, Marwan A.
    2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 3318 - 3321
  • [25] Right ventricular assist system feedback flow control parameter for a rotary blood pump
    Yoshikawa, M
    Nakata, K
    Nonaka, K
    Linneweber, J
    Kawahito, S
    Takano, T
    Shulte-Eistrup, S
    Maeda, T
    Glueck, J
    Schima, H
    Wolner, E
    Nosé, Y
    ARTIFICIAL ORGANS, 2000, 24 (08) : 659 - 666
  • [26] Physiologic outcome of varying speed rotary blood pump support algorithms: a review study
    Selim Bozkurt
    Australasian Physical & Engineering Sciences in Medicine, 2016, 39 : 13 - 28
  • [27] Novel control system to prevent right ventricular failure induced by rotary blood pump
    Arakawa, Mamoru
    Nishimura, Takashi
    Takewa, Yoshiaki
    Umeki, Akihide
    Ando, Masahiko
    Kishimoto, Yuichiro
    Fujii, Yutaka
    Kyo, Shunei
    Adachi, Hideo
    Tatsumi, Eisuke
    JOURNAL OF ARTIFICIAL ORGANS, 2014, 17 (02) : 135 - 141
  • [28] Novel control system to prevent right ventricular failure induced by rotary blood pump
    Mamoru Arakawa
    Takashi Nishimura
    Yoshiaki Takewa
    Akihide Umeki
    Masahiko Ando
    Yuichiro Kishimoto
    Yutaka Fujii
    Shunei Kyo
    Hideo Adachi
    Eisuke Tatsumi
    Journal of Artificial Organs, 2014, 17 : 135 - 141
  • [29] Effect of rotary blood pump failure on left ventricular energetics assessed by mathematical modeling
    Vandenberghe, S
    Segers, P
    Meyns, B
    Verdonck, PR
    ARTIFICIAL ORGANS, 2002, 26 (12) : 1032 - 1039
  • [30] Development of a Pump Flow Estimator for Rotary Blood Pumps to Enhance Monitoring of Ventricular Function
    Granegger, Marcus
    Moscato, Francesco
    Casas, Fernando
    Wieselthaler, Georg
    Schima, Heinrich
    ARTIFICIAL ORGANS, 2012, 36 (08) : 691 - 699