Co-registration of angiography and intravascular ultrasound images through image-based device tracking

被引:5
|
作者
Prasad, Megha [1 ]
Cassar, Andrew [1 ]
Fetterly, Kenneth A. [1 ]
Bell, Malcolm [1 ]
Theessen, Heike [2 ]
Ecabert, Olivier [2 ]
Bresnahan, John F. [1 ]
Lerman, Amir [1 ]
机构
[1] Mayo Clin, Sch Med, Div Cardiol, Rochester, MN USA
[2] Siemens Corp Res & Technol, Munich, Germany
关键词
imaging; integration; coronary artery disease; novel; PERCUTANEOUS CORONARY INTERVENTION; SHEAR-STRESS; TOMOGRAPHY; LESIONS; FUSION;
D O I
10.1002/ccd.26340
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ObjectivesTo determine the feasibility of automated co-registration of angiography and intravascular ultrasound (IVUS) to facilitate integration of these two imaging modalities in a synchronous manner. BackgroundIVUS provides cross-sectional imaging of coronary arteries but lacks overview of the vascular territory provided by angiography. Co-registration of angiography and IVUS would increase utility of IVUS in the clinical setting. MethodsForty-nine consecutive patients undergoing surveillance for cardiac allograft vasculopathy with angiography and IVUS of the left anterior descending artery (LAD) were enrolled. A pre-IVUS angiogram of the LAD was performed followed by an ECG-triggered fluoroscopy (ECGTF) during IVUS pullback at 0.5mm/s using an automatic pullback device. ECGTF was used to track the IVUS catheter during pullback and establish a spatial relationship to the pre-IVUS angiogram. Angio-IVUS co-registration was performed with a research prototype (Siemens Healthcare, Germany) and accuracy was evaluated by distance mismatch between angiography and IVUS images at vessel bifurcations. ResultsMedian age was 54 (44.5, 67) years. The population was 82.6% male with minimal risk factors. The median (IQR) co-registration distance mismatch measured at 108 bifurcations in 42 (85%) patients was 0.35 (0.00-1.16) mm. Seven patients were excluded due to inappropriate data acquisition (n=3) and failure of tracking (n=4), e.g., due to overlapping sternal wires. Estimated effective radiation dose for ECGTF was 0.09mSv. ConclusionThis study demonstrates the feasibility of angio-IVUS co-registration which may be used as a clinical tool for localizing IVUS cross-sections along an angiographic roadmap. (c) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:1077 / 1082
页数:6
相关论文
共 50 条
  • [21] Image registration in a tomographic ultrasound system: comparison between camera-tracking and image-based motion compensation
    Ranger, Bryan J.
    Feigin, Micha
    Herr, Hugh M.
    Anthony, Brian W.
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [22] Quantification of the focal progression of coronary atherosclerosis through automated co-registration of virtual histology-intravascular ultrasound imaging data
    Timmins, Lucas H.
    Molony, David S.
    Eshtehardi, Parham
    Rasoul-Arzrumly, Emad
    Lam, Adrian
    Hung, Olivia Y.
    McDaniel, Michael C.
    Oshinski, John N.
    Giddens, Don P.
    Samady, Habib
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2017, 33 (01): : 13 - 24
  • [23] Image-based Guidance System for Intravascular Microrobot: Fiducial Marker-based Registration using Biplanar Fluoroscopic Images & CTA Images
    Oh, Jinsik
    Park, Jong-Oh
    Park, Sukho
    Ko, Seong Young
    2015 15TH INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND SYSTEMS (ICCAS), 2015, : 919 - 922
  • [24] Quantification of the focal progression of coronary atherosclerosis through automated co-registration of virtual histology-intravascular ultrasound imaging data
    Lucas H. Timmins
    David S. Molony
    Parham Eshtehardi
    Emad Rasoul-Arzrumly
    Adrian Lam
    Olivia Y. Hung
    Michael C. McDaniel
    John N. Oshinski
    Don P. Giddens
    Habib Samady
    The International Journal of Cardiovascular Imaging, 2017, 33 : 13 - 24
  • [25] Image-Based Cardiac Phase Retrieval in Intravascular Ultrasound Sequences
    Hernandez-Sabate, Aura
    Gil, Debora
    Garcia-Barnes, Jaume
    Marti, Enric
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (01) : 60 - 72
  • [26] Heterogeneous Images' Co-registration of Oil Tanks Based on Double-reflection Arcs in SAR Image
    Sun, Bing
    Zhang, Xinliang
    Ma, Haizhong
    Yang, Wei
    Xu, Hailun
    11TH EUROPEAN CONFERENCE ON SYNTHETIC APERTURE RADAR (EUSAR 2016), 2016, : 475 - 478
  • [27] Automatic co-registration of volumetric images based on implanted fiducial markers
    Koch, Martin
    Maltz, Jonathan S.
    Gangadharan, Bijumon
    Bose, Supratik
    Shukla, Himanshu
    Bani-Hashemi, Ali R.
    Belongie, Serge J.
    2008 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, VOLS 1-4, 2008, : 1171 - +
  • [28] Fluorescence Lifetime Imaging and Intravascular Ultrasound: Co-Registration Study Using Ex Vivo Human Coronaries
    Gorpas, Dimitris
    Fatakdawala, Hussain
    Bec, Julien
    Ma, Dinglong
    Yankelevich, Diego R.
    Qi, Jinyi
    Marcu, Laura
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2015, 34 (01) : 156 - 166
  • [29] Tomography image co-registration algorithm based on principle of mechanics decomposition
    Liu, Jun
    Zhu, Shan-An
    He, Bin
    Zhejiang Daxue Xuebao (Gongxue Ban)/Journal of Zhejiang University (Engineering Science), 2008, 42 (09): : 1601 - 1605
  • [30] Research on SAR image co-registration method based on feature points
    Liu Z.-Y.
    Lai L.-Y.
    Zhang G.-B.
    Qi H.-C.
    Pan Y.-F.
    Peng L.-C.
    Ng A.H.-M.
    Dizhen Dizhi, 2021, 43 (05): : 1339 - 1350