Assessment of cortical bone with clinical and ultrashort echo time sequences

被引:60
|
作者
Du, Jiang [1 ]
Hermida, Juan C. [2 ]
Diaz, Eric [1 ]
Corbeil, Jacqueline [1 ]
Znamirowski, Richard [1 ]
D'Lima, Darryl D. [2 ]
Bydder, Graeme M. [1 ]
机构
[1] Univ Calif San Diego, Dept Radiol, San Diego, CA 92103 USA
[2] Scripps Res Inst, Dept Mol & Expt Med, La Jolla, CA 92037 USA
关键词
ultrashort echo time; projection reconstruction; adiabatic inversion; cortical bone; free water; bound water; SPECTROSCOPIC IMAGING UTESI; RF EXCITATION; BOUND WATER; MRI; CONTRAST; POROSITY; DENSITY; MATRIX; SIGNAL; PORE;
D O I
10.1002/mrm.24497
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We describe the use of ultrashort echo time (UTE) sequences and fast spin echo sequences to assess cortical bone using a clinical 3T scanner. Regular two- and three-dimensional UTE sequences were used to image both bound and free water in cortical bone. Adiabatic inversion recovery prepared UTE sequences were used to image water bound to the organic matrix. Two-dimensional fast spin echo sequences were used to image free water. Regular UTE sequences were used together with bicomponent analysis to measure T*(2)s and relative fractions of bound and free water components in cortical bone. Inversion recovery prepared UTE sequences were used to measure the T*(2) of bound water. Saturation recovery UTE sequences were used to measure the T-1 of bone water. Eight cadaveric human cortical bone samples and a lower leg specimen were studied. Preliminary results show two distinct components in UTE detected signal decay, a single component in inversion recovery prepared UTE detected signal decay, and a single component in saturation recovery UTE detected signal recovery. Regular UTE sequences appear to depict both bound and free water in cortical bone. Inversion recovery prepared UTE sequences appear to depict water bound to the organic matrix. Two-dimensional fast spin echo sequences appear to depict bone structure corresponding to free water in large pores. Magn Reson Med 70:697-704, 2013. (c) 2012 Wiley Periodicals, Inc.
引用
收藏
页码:697 / 704
页数:8
相关论文
共 50 条
  • [41] An Update in Qualitative Imaging of Bone Using Ultrashort Echo Time Magnetic Resonance
    Jerban, Saeed
    Chang, Douglas G.
    Ma, Yajun
    Jang, Hyungseok
    Chang, Eric Y.
    Du, Jiang
    FRONTIERS IN ENDOCRINOLOGY, 2020, 11
  • [42] Quantitative Ultrashort Echo Time (UTE) Magnetic Resonance Imaging of Bone: An Update
    Ma, Ya-Jun
    Jerban, Saeed
    Jang, Hyungseok
    Chang, Douglas
    Chang, Eric Y.
    Du, Jiang
    FRONTIERS IN ENDOCRINOLOGY, 2020, 11
  • [43] Evaluation of gadolinium deposition in cortical bone using three-dimensional ultrashort echo time quantitative susceptibility mapping: A preliminary study
    Zhang, Xiaodong
    Zhou, Beibei
    Chen, Yanjun
    Cai, Zhenyu
    Guo, Yihao
    Wei, Zhao
    Li, Shisi
    Feng, Yanqiu
    Sedaghat, Sam
    Jang, Hyungseok
    NMR IN BIOMEDICINE, 2024, 37 (01)
  • [44] Detection of gadolinium deposition in cortical bone with ultrashort echo time T1mapping: an ex vivo study in a rabbit model
    Zhao, Kaixuan
    Li, Shisi
    Yi, Peiwei
    Guo, Yihao
    Yu, Qinqin
    Zhu, Cuiling
    Feng, Qianjin
    Du, Jiang
    Zhang, Xiaodong
    Feng, Yanqiu
    EUROPEAN RADIOLOGY, 2021, 31 (03) : 1569 - 1577
  • [45] Quantification of Human Cortical Bone Bound and Free Water in Vivo with Ultrashort Echo Time MR Imaging: A Model-based Approach
    Abbasi-Rad, Shahrokh
    Rad, Hamidreza Saligheh
    RADIOLOGY, 2017, 283 (03) : 861 - 871
  • [46] Detection of gadolinium deposition in cortical bone with ultrashort echo time T1 mapping: an ex vivo study in a rabbit model
    Kaixuan Zhao
    Shisi Li
    Peiwei Yi
    Yihao Guo
    Qinqin Yu
    Cuiling Zhu
    Qianjin Feng
    Jiang Du
    Xiaodong Zhang
    Yanqiu Feng
    European Radiology, 2021, 31 : 1569 - 1577
  • [47] MRI-based Attenuation Correction for Emission Tomography using Ultrashort Echo Time Sequences
    Keereman, Vincent
    Vanhove, Christian
    Vandenberghe, Stefaan
    EMAGRES, 2012, 1 (02): : 323 - 332
  • [48] Bone Imaging of the Knee Using Short-Interval Delta Ultrashort Echo Time and Field Echo Imaging
    Bae, Won C.
    Malis, Vadim
    Yamashita, Yuichi
    Mesa, Anya
    Vucevic, Diana
    Miyazaki, Mitsue
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (16)
  • [49] Detecting small pulmonary nodules with spiral ultrashort echo time sequences in 1.5 T MRI
    Huang, Yu-Sen
    Niisato, Emi
    Su, Mao-Yuan Marine
    Benkert, Thomas
    Hsu, Hsao-Hsun
    Shih, Jin-Yuan
    Chen, Jin-Shing
    Chang, Yeun-Chung
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2021, 34 (03) : 399 - 409
  • [50] MRI-Based Attenuation Correction for PET/MRI Using Ultrashort Echo Time Sequences
    Keereman, Vincent
    Fierens, Yves
    Broux, Tom
    De Deene, Yves
    Lonneux, Max
    Vandenberghe, Stefaan
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (05) : 812 - 818