Evaluation of 3D Point Spread Function of a Semi-Ellipsoidal Ultrasound Computer Tomography System

被引:2
|
作者
Ruiter, N. V. [1 ]
Dapp, R. [1 ]
Zapf, M. [1 ]
Gemmeke, H. [1 ]
机构
[1] Karlsruhe Inst Technol, Inst Data Proc & Elect, Karlsruhe, Germany
关键词
D O I
10.1109/ULTSYM.2011.0059
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The design of 3DUSCT II was optimized with the aim of nearly isotropic and approx. spatially invariant 3D point spread function (PSF). To evaluate experimentally the 3DPSF, a thin wire was twisted into an irregular helix and imaged with the USCT system. The PSF was measured at 55 points along the wire as 2D profiles of the full width at half maximum (FWHM) in a plane perpendicular to the local direction of the thread. The mean FWHM over all profiles was 0.66 mm. The standard deviation (STD) of the FWHM profiles as measure of the isotropy had a mean of 0.18 mm. The variability of the mean FWHM over the points shows no systematic trend over the imaged space.
引用
收藏
页码:242 / 245
页数:4
相关论文
共 50 条
  • [21] Handheld 3D Ophthalmic Ultrasound Tomography System
    To, Josiah
    Wang, Haocun
    Wei, Xiao
    Vu, Anderson
    Tang, William
    Browne, Andrew
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [22] Prototype of a new 3D ultrasound computer tomography system:: transducer design and data recording
    Stotzka, R
    Widmann, H
    Müller, T
    Schlote-Holubek, K
    Gemmeke, H
    Ruiter, N
    Göbel, G
    MEDICAL IMAGING 2004: ULTRASONIC IMAGING AND SIGNAL PROCESSING, 2004, 5373 : 70 - 79
  • [23] Quantitative assessment of cone-beam CT system by 3D point spread function
    Chen, ZK
    Ning, RL
    Conover, D
    Yu, Y
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 574 - 585
  • [24] First Results of a Clinical Study with 3D Ultrasound Computer Tomography
    Ruiter, N. V.
    Zapf, M.
    Dapp, R.
    Hopp, T.
    Kaiser, W. A.
    Gemmeke, H.
    2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 647 - 650
  • [25] Phase aberration correction for 3D ultrasound computer tomography images
    Ruiter, N. V.
    Schnell, R.
    Zapf, M.
    Gemmeke, H.
    2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 1808 - 1811
  • [26] 3D Ultrasound Computer Tomography: Results with a Clinical Breast Phantom
    Ruiter, N. V.
    Schwarzenberg, G. F.
    Zapf, M.
    Liu, R.
    Stotzka, R.
    Gemmeke, H.
    2006 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-5, PROCEEDINGS, 2006, : 989 - 992
  • [27] 3D Ultrasound Computer Tomography: Update from a clinical study
    Hopp, T.
    Zapf, M.
    Kretzek, E.
    Henrich, J.
    Tukalo, A.
    Gemmeke, H.
    Kaiser, C.
    Knaudt, J.
    Ruiter, N. V.
    MEDICAL IMAGING 2016: ULTRASONIC IMAGING AND TOMOGRAPHY, 2016, 9790
  • [28] Hardware Setup for the Next Generation of 3D Ultrasound Computer Tomography
    Gemmeke, H.
    Berger, L.
    Birk, M.
    Goebel, G.
    Menshikov, A.
    Tcherniakhovski, D.
    Zapf, M.
    Ruiter, N. V.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2449 - 2454
  • [29] Point Spread Function Engineering to Map 3D Particle Motion
    Bonin, Keith
    Prasad, Sudhakar
    Kefer, Paul
    Holzwarth, George M.
    Vidi, Pierre-Alexandre
    BIOPHYSICAL JOURNAL, 2020, 118 (03) : 464A - 464A
  • [30] 3D microscopy with a double-helix point spread function
    Pavani, Sri Rama Prasanna
    Piestun, Rafael
    THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING XVI, 2009, 7184