The wear pattern in metal-on-metal hip prostheses

被引:48
|
作者
Anissian, HL [1 ]
Stark, A
Good, V
Dahlstrand, H
Clarke, IC
机构
[1] Karolinska Hosp, Dept Orthopaed Surg, S-17176 Stockholm, Sweden
[2] Loma Linda Univ, Med Ctr, Dept Orthopaed Surg, Loma Linda, CA 92350 USA
来源
关键词
implants; hip-joint simulator; wear; run-in phase; steady-state phase; metal-on-metal prosthesis;
D O I
10.1002/jbm.1068
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The accumulated data suggest that there is a strong correlation between polyethylene wear and osteolysis, which ultimately leads to prosthetic loosening. Second-generation metal-on-metal prostheses have been introduced, with an eye toward resolving this wear-induced osteolysis problem. The metal particles and ions are biologically active and can affect the cell homeostasis. Thus, defining the wear pattern and ratio of a given metal-on-metal prosthesis system is desirable. An early high-wear or run-in phase followed by a low-wear phase or steady state has been suggested for metal-on-metal hip prostheses. The aim of this study was to define the wear pattern of metal-on-metal bearings. The prosthesis systems were tested in a joint simulator. An early accelerated wear phase transformed to a slower wear phase after 700,000 cycles. The run-in and steady-state wear rates for combined head and liner averaged 2.22 mm(3)/Mc and 1.0 mm(3)/ million cycles, respectively. The metal-on-metal prosthesis featured a biphasic wear trend with the wear ratio in run in being more than twofold higher than the steady-state phase. (C) 2001 John Wiley & Sons, Inc.
引用
收藏
页码:673 / 678
页数:6
相关论文
共 50 条
  • [41] Predicting Wear and Blood Metal Ion Levels in Metal-on-Metal Hip Resurfacing
    Matthies, Ashley K.
    Henckel, Johann
    Cro, Suzie
    Suarez, Alexander
    Noble, Philip C.
    Skinner, John
    Hart, Alister J.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2014, 32 (01) : 167 - 174
  • [42] Characterization of protein degradation in serum-based lubricants during simulation wear testing of metal-on-metal hip prostheses
    Maskiewicz, Victoria K.
    Williams, Paul A.
    Prates, Sarah J.
    Bowsher, John G.
    Clarke, Ian C.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2010, 94B (02) : 429 - 440
  • [43] Metal release from hip prostheses: biological monitoring of cobalt and chromium in patients with metal-on-metal hip arthroplasty
    Torra, Merce
    Sandalinas, Silvia
    Martin, Mayte
    Nogue, Santiago
    TRACE ELEMENTS AND ELECTROLYTES, 2017, 34 (01) : 1 - 4
  • [44] Comparison of two cobalt-based alloys for use in metal-on-metal hip prostheses: Evaluation of the wear properties in a simulator
    St John, KR
    Poggie, RA
    Zardiackas, LD
    Afflitto, RM
    COBALT-BASE ALLOYS FOR BIOMEDICAL APPLICATIONS, 1999, 1365 : 145 - 155
  • [45] The relationship between activity and ions in patients with metal-on-metal bearing hip prostheses
    Heisel, C
    Silva, M
    Skipor, AK
    Jacobs, JJ
    Schmalzried, TP
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (04): : 781 - 787
  • [46] Bioengineering reasons for the failure of metal-on-metal hip prostheses AN ENGINEER'S PERSPECTIVE
    Fisher, J.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (08): : 1001 - 1004
  • [47] Cobalt toxicity - an emerging clinical problem in patients with metal-on-metal hip prostheses?
    Mao, Xinzhan
    Wong, Andrew A.
    Crawford, Ross W.
    MEDICAL JOURNAL OF AUSTRALIA, 2011, 194 (12) : 649 - 651
  • [48] Carbon composition effects on wear behaviour and wear mechanisms of metal-on-metal hip prosthesis
    Ihaddadene, R.
    Affatato, S.
    Zavalloni, M.
    Bouzid, S.
    Viceconti, M.
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2011, 14 : 33 - 34
  • [49] Influence of cup orientation on the wear performance of metal-on-metal hip replacements
    Angadji, A.
    Royle, M.
    Collins, S. N.
    Shelton, J. C.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2009, 223 (H4) : 449 - 457
  • [50] Wear of ceramic coated metal-on-metal bearings used for hip replacement
    Gispert, M. P.
    Serro, A. P.
    Colaco, R.
    Pires, E.
    Saramago, B.