A novel approach to determine primary stability of acetabular press-fit cups

被引:12
|
作者
Weissmann, Volker [1 ,2 ]
Boss, Christian [3 ]
Bader, Rainer [2 ]
Hansmann, Harald [1 ]
机构
[1] Univ Appl Sci Technol Business & Design, Fac Engn, Philipp Muller Str 14, D-23966 Wismar, Germany
[2] Rostock Univ Med, Dept Orthoped, Biomech & Implant Technol Res Lab, Doberaner Str 142, D-18057 Rostock, Germany
[3] Inst Polymer Technol eV, Alter Holzhafen 19, D-23966 Wismar, Germany
关键词
Ti6A14V; Selective laser melting; Press-fit; Primary stability; TOTAL HIP-ARTHROPLASTY; IMPROVED PRIMARY FIXATION; POROUS TITANIUM; MECHANICAL-PROPERTIES; SURFACE ARCHITECTURE; INITIAL STABILITY; DESIGN; STEM; IMPLANTS; LASER;
D O I
10.1016/j.jmbbm.2018.01.015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Today hip cups are used in a large variety of design variants and in increasing numbers of units. Their development is steadily progressing. In addition to conventional manufacturing methods for hip cups, additive methods, in particular, play an increasingly important role as development progresses. The present paper describes a modified cup model developed based on a commercially available press-fit cup (Allofit 54/JJ). The press-fit cup was designed in two variants and manufactured using selective laser melting (SLM). Variant 1 (Ti) was modeled on the Alio& cup using an adapted process technology. Variant 2 (Ti-S) was provided with a porous load bearing structure on its surface. In addition to the typical (complete) geometry, both variants were also manufactured and tested in a reduced shape where only the press-fit area was formed. To assess the primary stability of the press-fit cups in the artificial bone cavity, pull-out and lever-out tests were carried out. Exact fit conditions and two-millimeter press-fit were investigated. The closed-cell PU foam used as an artificial bone cavity was mechanically characterized to exclude any influence on the results of the investigation. The pull-out forces of the Ti-variant (complete-526 N, reduced-468 N) and the Ti-S variant (complete-548 N, reduced-526 N) as well as the lever-out moments of the Ti-variant (complete-10 Nm, reduced-9.8 Nm) and the Ti-S variant (complete-9 Nm, reduced-7.9 N) show no significant differences in the results between complete and reduced cups. The results show that the use of reduced cups in a press-fit design is possible within the scope of development work.
引用
收藏
页码:1 / 10
页数:10
相关论文
共 50 条
  • [1] A novel method to assess primary stability of press-fit acetabular cups
    Crosnier, Emilie A.
    Keogh, Patrick S.
    Miles, Anthony W.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2014, 228 (11) : 1126 - 1134
  • [2] STABILITY OF PRESS-FIT ACETABULAR CUPS
    KUMMER, FJ
    STUCHIN, SA
    ADLER, A
    BULLETIN OF THE HOSPITAL FOR JOINT DISEASES ORTHOPAEDIC INSTITUTE, 1991, 51 (01): : 116 - 116
  • [3] Effect of cup medialization on primary stability of press-fit acetabular cups
    Morosato, Federico
    Cristofolini, Luca
    Castagnini, Francesco
    Traina, Francesco
    CLINICAL BIOMECHANICS, 2020, 80
  • [4] Effect of bearing friction torques on the primary stability of press-fit acetabular cups: A novel in vitro method
    Jahnke, Alexander
    Schroeder, Stefan
    Fonseca Ulloa, Carlos A.
    Ahmed, Gafar A.
    Ishaque, Bernd A.
    Rickert, Markus
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2018, 36 (10) : 2745 - 2753
  • [5] Initial stability of press-fit acetabular cups - An in-vitro study
    Wetzel, R
    Simnacher, M
    Scheller, G
    BIOMEDIZINISCHE TECHNIK, 2005, 50 (12): : 400 - 403
  • [6] Deformation characteristics and eigenfrequencies of press-fit acetabular cups
    Hothan, Arne
    Huber, Gerd
    Weiss, Cornelius
    Hoffmann, Norbert
    Morlock, Michael
    CLINICAL BIOMECHANICS, 2011, 26 (01) : 46 - 51
  • [7] Initial stability of press-fit acetabular cups. In vitro lever-out trials
    Kuhn, A
    Scheller, G
    Schwarz, M
    BIOMEDIZINISCHE TECHNIK, 1999, 44 (12): : 356 - 359
  • [8] In vitro examination of the primary stability of three press-fit acetabular cups under consideration of two different bearing couples
    Jahnke, Alexander
    Bott, Catharina Chiara
    Ulloa, Carlos Alfonso Fonseca
    Jahnke, Gerhard Walter
    Rickert, Markus
    Ishaque, Bernd Alexander
    Ahmed, Gafar Adam
    MEDICAL ENGINEERING & PHYSICS, 2019, 67 : 49 - 54
  • [9] Deformation of acetabular press-fit cups: Influence of design and surgical factors
    Messer-Hannemann, Philipp
    Campbell, Graeme M.
    Morlock, Michael M.
    CLINICAL BIOMECHANICS, 2019, 69 : 96 - 103
  • [10] Initial stability of cementless acetabular cups: press-fit and screw fixation interaction—an in vitro biomechanical study
    Tabata T.
    Kaku N.
    Hara K.
    Tsumura H.
    European Journal of Orthopaedic Surgery & Traumatology, 2015, 25 (3) : 497 - 502