Comparison of the Abbott m2000 RealTime CT Assay and the Cepheid GeneXpert CT/NG Assay to the Roche Amplicor CT Assay for Detection of Chlamydia trachomatis in Ocular Samples from Tanzania

被引:32
|
作者
Dize, Laura [1 ]
West, Sheila [1 ]
Williams, James A. [2 ]
Van der Pol, Barbara [2 ,3 ]
Quinn, Thomas C. [1 ,4 ]
Gaydos, Charlotte A. [1 ]
机构
[1] Johns Hopkins Univ, Baltimore, MD 21218 USA
[2] Indiana Univ Sch Med, Indianapolis, IN USA
[3] Indiana Univ, Sch Publ Hlth, Bloomington, IN USA
[4] NIAID, Div Intramural Res, NIH, Bethesda, MD 20892 USA
基金
美国国家卫生研究院;
关键词
MASS TREATMENT; INFECTION;
D O I
10.1128/JCM.00519-13
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The GeneXpert CT/NG assay (GeneXpert) and the Abbott m2000 RealTime CT (m2000) assay were compared to Amplicor for detecting ocular Chlamydia trachomatis. Discordant specimens were tested by the Aptima CT assay. The m2000 assay sensitivity was 100% (95% confidence interval [CI], 90% to 100%), and specificity was 98.46% (95% CI, 95.2% to 99.2%); GeneXpert sensitivity was 100% (95% CI, 90% to 100%), and specificity was 100% (95% CI, 98.1% to 100%). The m2000 and GeneXpert assays appear to perform as well as the Amplicor assay.
引用
收藏
页码:1611 / 1613
页数:3
相关论文
共 50 条
  • [1] COMPARISON OF THE ABBOTT M2000 REALTIME CT ASSAY FOR CHLAMYDIA TRACHOMATIS MONITORING IN TANZANIA COMPARED TO THE ROCHE AMPLICOR CT ASSAY
    Agreda, P.
    Dize, L.
    West, S.
    Williams, J. A.
    Van der Pol, B.
    Gaydos, C. A.
    [J]. SEXUALLY TRANSMITTED INFECTIONS, 2011, 87 : A272 - A272
  • [2] Evaluation of the Abbott RealTime CT/NG Assay in Comparison to the Roche Cobas Amplicor CT/NG Assay
    Cheng, Annie
    Qian, Qinfang
    Kirby, James E.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2011, 49 (04) : 1294 - 1300
  • [3] Comparison of analytical performances of the Roche Cobas 6800 CT/NG assay with the Abbott m2000 Real Time CT/NG assay for detecting Chlamydia trachomatis and Neisseria gonorrhoeae
    Cherkaoui, Abdessalam
    Renzi, Gesuele
    Mombelli, Michele
    Jaton, Katia
    Yerly, Sabine
    Vuilleumier, Nicolas
    Schrenzel, Jacques
    [J]. JOURNAL OF MEDICAL MICROBIOLOGY, 2019, 68 (02) : 197 - 200
  • [4] Evaluation of the Abbott realtime Ct/Ng assay using the M2000 system and comparison with the GenProbe Aptima Combo 2 assay
    Quigley, D. I.
    Ellefson, A.
    Steed, L. L.
    Wolff, D. J.
    [J]. JOURNAL OF MOLECULAR DIAGNOSTICS, 2007, 9 (05): : 680 - 680
  • [5] Detection of Neisseria gonorrhoeae or Chlamydia trachomatis from oral wash specimens using the Abbott RealTime CT/NG assay and the Cobas 4800 CT/NG assay: A prospective study
    Matsumoto, Masahiro
    Hamasuna, Ryoichi
    Le, Phuong Thi
    Fujimoto, Naohiro
    Matsumoto, Tetsuro
    [J]. JOURNAL OF INFECTION AND CHEMOTHERAPY, 2022, 28 (04) : 480 - 485
  • [6] Evaluation of pooled ocular and vaginal swabs by the Cepheid GeneXpert CT/NG assay for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae compared to the GenProbe Aptima Combo 2 Assay
    Dize, L.
    West, S. K.
    Mkocha, H.
    Quinn, T. C.
    Gaydos, C. A.
    [J]. DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2015, 81 (02) : 102 - 104
  • [7] Analytical Evaluation of the Abbott RealTime CT/NG Assay for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae in Rectal and Pharyngeal Swabs
    Adamson, Paul C.
    Pandori, Mark W.
    Doernberg, Sarah B.
    Komarow, Lauren
    Sund, Zoe
    Tran, Thuy Tien T.
    Jensen, David
    Tsalik, Ephraim L.
    Deal, Carolyn D.
    Chambers, Henry F.
    Fowler, Vance G., Jr.
    Evans, Scott R.
    Patel, Robin
    Klausner, Jeffrey D.
    [J]. JOURNAL OF MOLECULAR DIAGNOSTICS, 2020, 22 (06): : 811 - 816
  • [8] Evaluation of the Abbott RealTime™ CT assay with the BD ProbeTec™ ET assay for the detection of Chlamydia trachomatis in a clinical microbiology laboratory
    Walsh, Anne
    Rourke, Fiona O.
    Lao, Bairbre Ni
    Crowley, Brendan
    [J]. DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2009, 64 (01) : 13 - 19
  • [9] Chlamydia trachomatis and Neisseria gonorrhoeae (CT/NG) Detection in SurePath and ThinPrep Specimens Using Abbot RealTime CT/NG Assay
    Lucic, D.
    Eisenhut, C.
    Cloherty, G.
    [J]. JOURNAL OF MOLECULAR DIAGNOSTICS, 2013, 15 (06): : 896 - 896
  • [10] Comparison of GMT presto assay and Roche cobase® 4800 CT/NG assay for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in dry swabs
    de Waaij, Dewi J.
    Dubbink, Jan Henk
    Peters, Remco P. H.
    Ouburg, Sander
    Morre, Servaas A.
    [J]. JOURNAL OF MICROBIOLOGICAL METHODS, 2015, 118 : 70 - 74