Validation of a Monte Carlo simulation of the Philips Allegro/GEMINI PET systems using GATE

被引:117
|
作者
Lamare, F [1 ]
Turzo, A [1 ]
Bizais, Y [1 ]
Le Rest, CC [1 ]
Visvikis, D [1 ]
机构
[1] Univ Bretagne Occidentale, CHU Morvan, INSERM U650, Lab Traitement Informat Med, F-29609 Brest, France
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2006年 / 51卷 / 04期
关键词
D O I
10.1088/0031-9155/51/4/013
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A newly developed simulation toolkit, GATE (Geant4 Application for Tomographic Emission), was used to develop a Monte Carlo simulation of a fully three-dimensional (3D) clinical PET scanner. The Philips Allegro/GEMINI PET systems were simulated in order to (a) allow a detailed study of the parameters affecting the system's performance under various imaging conditions, (b) Study the optimization and quantitative accuracy of emission acquisition protocols for dynamic and static imaging, and (c) further validate the potential of GATE for the Simulation of clinical PET systems. A model of the detection system and its geometry was developed. The accuracy of the developed detection model was tested through the comparison of simulated and measured results obtained with the Allegro/GEMINI systems for a number of NEMA NU2-2001 performance protocols including spatial resolution, sensitivity and scatter fraction. In addition, in approximate model of the system's dead time at the level of detected single events and coincidences was developed in an attempt to simulate the count rate related performance characteristics of the scanner. The developed dead-time model was assessed under different imaging conditions using the Count rate loss and noise equivalent Count rates performance protocols of standard and modified NEMA NU2-2001 (whole body imaging conditions) and NEMA NU2-1994 (brain imaging conditions) comparing simulated with experimental measurements obtained with the Allegro/GEMINI PET systems. Finally,a reconstructed image quality protocol was used to assess the overall performance of the developed model. An agreement of < 3% was obtained in scatter fraction, with a difference between 4% and 10% in the true and random coincidence Count rates respectively, throughout a range of activity concentrations and under various imaging conditions, resulting in < 8% differences between simulated and measured noise equivalent Count rates performance. Finally, the image quality validation Study revealed a good agreement in signal-to-noise ratio and contrast recovery coefficients fora number of different volume spheres and two different (clinical level based) tumour-to-background ratios. In conclusion, these results support the accurate modelling of the Philips Allegro/GEMINI PET systems using GATE in combination with a dead-time model for the signal flow description, which leads to an agreement of < 10% in coincidence count rates under different imaging conditions and clinically relevant activity concentration levels.
引用
收藏
页码:943 / 962
页数:20
相关论文
共 50 条
  • [1] Validating a GATE Monte Carlo Simulation of a Philips Gemini PET/CT Scanner
    Golby, C.
    Partridge, M.
    Seco, J.
    Bidmead, M.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S260 - S261
  • [2] Validation of a Monte Carlo simulation of the Inveon PET scanner using GATE
    Lu, Lijun
    Zhang, Houjin
    Bian, Zhaoying
    Ma, Jianhua
    Feng, Qiangjin
    Chen, Wufan
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 828 : 170 - 175
  • [3] Monte-Carlo simulation of the photodetection systems prototype PET scanner using GATE: A validation study
    Groiselle, CJ
    Kudrolli, HA
    Glick, SJ
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3130 - 3132
  • [4] Simulation of the Allegro PET system using GATE
    Lamare, F
    Turzo, A
    Bizais, Y
    Visvikis, D
    [J]. MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 890 - 897
  • [5] Monte Carlo simulation in PET and SPECT instrumentation using GATE
    Assié, K
    Breton, V
    Buvat, I
    Comtat, C
    Jan, S
    Krieguer, M
    Lazaro, D
    Morel, C
    Rey, M
    Santin, G
    Simon, L
    Staelens, S
    Strul, D
    Vieira, JM
    Van de Walle, R
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 527 (1-2): : 180 - 189
  • [6] Validation of GATE Monte Carlo Simulations of the Philips GEMINI TF and TruFlight Select PET/CT Scanners Based on NEMA NU2 Standards
    Trindade, Andreia
    Rodrigues, Pedro
    Perkins, Amy E.
    Miller, Michael A.
    Narayanan, Manoj
    Griesmer, Jerome
    Tung, Chi-Rua
    Zhang, Bin
    Shao, Lingxiong
    Laurence, Thomas
    Solf, Torsten
    Wieczorek, Rerfried
    [J]. 2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2546 - 2549
  • [7] Validation of GATE Monte Carlo simulations for Siemens PET/CT and PET/MR systems
    Kertesz, H.
    Renner, A.
    Rausch, I.
    Beyer, T.
    Cal-Gonzalez, J.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S298 - S299
  • [8] Validation of a Monte Carlo Model of the PET Component of the Gemini GXL PET/CT
    Lavoie, G.
    Martel, C.
    Despres, P.
    Lacroix, F.
    Chen, Y.
    Carrier, J.
    [J]. MEDICAL PHYSICS, 2008, 35 (06) : 2675 - +
  • [9] A Monte Carlo Simulation of PET of a Real Patient with GATE
    Gerganov, G.
    Kang, H.
    Madzhunkov, Y.
    Mitev, K.
    Schmidtlein, C. R.
    Kawrakow, I.
    Kirov, A.
    [J]. NUCLEAR PHYSICS METHODS AND ACCELERATORS IN BIOLOGY AND MEDICINE, 2009, 1204 : 219 - +
  • [10] PET digitization chain for Monte Carlo simulation in GATE
    Salvadori, Julien
    Merlet, Antoine
    Presles, Benoit
    Cabello, Jorge
    Su, Kuan-Hao
    Cochet, Alexandre
    Etxebeste, Ane
    Vrigneaud, Jean-Marc
    Sarrut, David
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (16):