Evaluation of Quantification of HIV-1 RNA Viral Load in Plasma and Dried Blood Spots by Use of the Semiautomated Cobas Amplicor Assay and the Fully Automated Cobas Ampliprep/TaqMan Assay, Version 2.0, in Kisumu, Kenya

被引:21
|
作者
Ouma, Kenneth N. [1 ]
Basavaraju, Sridhar V. [2 ]
Okonji, Jully A. [1 ]
Williamson, John [3 ]
Thomas, Timothy K. [3 ]
Mills, Lisa A. [3 ]
Nkengasong, John N. [4 ]
Zeh, Clement [3 ]
机构
[1] Kenya Govt Med Res Ctr, Ctr Global Hlth Res, Kisumu, Kenya
[2] Ctr Dis Control & Prevent, HIV Prevent Branch, Div Global HIV AIDS, Ctr Global Hlth, Atlanta, GA USA
[3] Ctr Dis Control & Prevent, Div HIV AIDS Prevent, Kisumu, Kenya
[4] Ctr Dis Control & Prevent, Int Lab Branch, Div Global HIV AIDS, Ctr Global Hlth, Atlanta, GA USA
关键词
VIRUS TYPE-1 RNA; TAQMAN HIV-1; ANTIRETROVIRAL THERAPY; DIAGNOSIS; QUANTITATION; PERFORMANCE; INFECTION; SENSITIVITY;
D O I
10.1128/JCM.03048-12
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
In Kenya, HIV-1 viral load monitoring is commonly performed with the Cobas Amplicor using plasma specimens. Interest is growing in transitioning to real-time PCR (RT-PCR), such as the Cobas Ampliprep/Cobas TaqMan (CAP/CTM), using dried blood spots (DBS). Before implementation, direct evaluation of the two assays using DBS field specimens is required. This study compares the sensitivity, specificity, negative and positive predictive values (NPV and PPV, respectively), concordance, and agreement between HIV-1 viral load measurements using plasma and DBS specimens obtained from 512 HIV-1-infected pregnant females enrolled in the Kisumu Breastfeeding Study and tested with the Cobas Amplicor and CAP/CTM assays. The sensitivity and NPV of viral load detection in DBS specimens were higher with CAP/CTM (sensitivity, 100%; 95% confidence interval [CI], 99.1 to 100.0%; NPV, 100%; 95% CI, 59.0 to 100.0%) than the Cobas Amplicor (sensitivity, 96.6%; 95% CI, 94.3 to 98.1%; NPV, 58.8%; 95% CI, 40.7 to 75.4%). The PPVs were comparable between both assays when using DBS. The specificity of viral load detection in DBS specimens was lower with CAP/CTM (77.8%; 95% CI, 40.0 to 97.2%) than that of the Cobas Amplicor (95.2%; 95% CI, 76.2 to 99.9%). Good concordance and agreement were observed when paired plasma and DBS specimens were tested with both assays. Lower specificity with the CAP/CTM is likely due to proviral HIV-1 DNA amplification and lower detection limits with RT-PCR. However, the CAP/CTM has better sensitivity and higher throughput than the Cobas Amplicor. These findings suggest that DBS may be a suitable alternative to plasma when using RT-PCR, which could increase access to viral load monitoring in resource-limited settings.
引用
收藏
页码:1208 / 1218
页数:11
相关论文
共 50 条
  • [41] Comparison of HIV-1 detection in plasma specimens and dried blood spots using the Roche COBAS Ampliscreen HIV-1 test in Kisumu, Kenya
    Okonji, Jully A.
    Basavaraju, Sridhar V.
    Mwangi, Jane
    Shiraishi, Ray W.
    Odera, Matthew
    Ouma, Kenneth
    Pitman, John P.
    Marum, Lawrence H.
    Ou, Chin-Yih
    Zeh, Clement
    [J]. JOURNAL OF VIROLOGICAL METHODS, 2012, 179 (01) : 21 - 25
  • [42] Separation of Plasma from Whole Blood by Use of the cobas Plasma Separation Card: a Compelling Alternative to Dried Blood Spots for Quantification of HIV-1 Viral Load
    Carmona, Sergio
    Seiverth, Britta
    Magubane, Dieketseng
    Hans, Lucia
    Hoppler, Matthias
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2019, 57 (04)
  • [43] Initial evaluation of the Roche COBAS TaqMan HIV-1 v2.0 assay for determining viral load in HIV-infected individuals
    Taylor, Ninon
    Schmid, Ilse
    Egle, Alexander
    Greil, Richard
    Patsch, Wolfgang
    Oberkofler, Hannes
    [J]. ANTIVIRAL THERAPY, 2009, 14 (08) : 1189 - 1193
  • [44] Analytical characteristics and comparative evaluation of Aptima HIV-1 Quant Dx assay with Ampliprep/COBAS TaqMan HIV-1 test v2.0
    Angelos. Hatzakis
    Helen Papachristou
    Sangeetha J. Nair
    Jacqueline Fortunko
    Tracy Foote
    HeeCheol Kim
    Tashi L. Peling
    Andrew J. Worlock
    [J]. Virology Journal, 13
  • [45] Analytical characteristics and comparative evaluation of Aptima HIV-1 Quant Dx assay with Ampliprep/COBAS TaqMan HIV-1 test v2.0
    Hatzakis, Angelos.
    Papachristou, Helen
    Nair, Sangeetha J.
    Fortunko, Jacqueline
    Foote, Tracy
    Kim, HeeCheol
    Peling, Tashi L.
    Worlock, Andrew J.
    [J]. VIROLOGY JOURNAL, 2016, 13
  • [46] Limited Utility of Dried-Blood- and Plasma Spot-Based Screening for Antiretroviral Treatment Failure with Cobas Ampliprep/TaqMan HIV-1 Version 2.0
    Sawadogo, Souleymane
    Shiningavamwe, Andreas
    Chang, Joy
    Maher, Andrew D.
    Zhang, Guoqing
    Yang, Chunfu
    Gaeb, Esegiel
    Kaura, Harold
    Ellenberger, Dennis
    Lowrance, David W.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2014, 52 (11) : 3878 - 3883
  • [47] Comparative evaluation of the Aptima HIV-1 Quant Dx assay and COBAS TaqMan HIV-1 v2.0 assay using the Roche High Pure System for the quantification of HIV-1 RNA in plasma
    Schalasta, Gunnar
    Boerner, Anna
    Speicher, Andrea
    Enders, Martin
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (03) : 493 - 499
  • [48] Performance evaluation of the Bioneer AccuPower® HIV-1 Quantitative RT-PCR kit: Comparison with the Roche COBAS® AmpliPrep/COBAS TaqMan® HIV-1 Test Ver.2.0 for quantification of HIV-1 viral load in Indonesia
    Kosasih, Agus Susanto
    Sugiarto, Christine
    Hayuanta, Hubertus Hosti
    Juhaendi, Runingsih
    Setiawan, Lyana
    [J]. ASIAN PACIFIC JOURNAL OF ALLERGY AND IMMUNOLOGY, 2018, 36 (02): : 120 - 125
  • [49] Field evaluation of Abbott Real Time HIV-1 Qualitative test for early infant diagnosis using dried blood spots samples in comparison to Roche COBAS Ampliprep/COBAS TaqMan HIV-1 Qual Test in Kenya
    Chang, Joy
    Omuomo, Kenneth
    Anyango, Emily
    Kingwara, Leonard
    Basiye, Frank
    Morwabe, Alex
    Shanmugam, Vedapuri
    Shon Nguyen
    Sabatier, Jennifer
    Zeh, Clement
    Ellenberger, Dennis
    [J]. JOURNAL OF VIROLOGICAL METHODS, 2014, 204 : 25 - 30
  • [50] HBV DNA QUANTIFICATION IN PLASMA AND SERUM WITH VERSION 2.0 OF THE COBAS AMPLIPREP/COBAS TAQMAN (CAP/CTM) REAL-TIME PCR ASSAY: INTRINSIC PERFORMANCE AND CLINICAL UTILITY
    Chevaliez, Stephane
    Bouvier-Alias, Magali
    Laperche, Syria
    Hezode, Christophe
    Pawlotsky, Jean-Michel
    [J]. HEPATOLOGY, 2010, 52 (04) : 525A - 525A