Synchrotron applications of an amorphous silicon flat-panel detector

被引:49
|
作者
Lee, John H. [1 ]
Aydiner, C. Can [2 ]
Almer, Jonathan [1 ]
Bernier, Joel [1 ]
Chapman, Karena W. [1 ]
Chupas, Peter J. [1 ]
Haeffner, Dean [1 ]
Kump, Ken
Lee, Peter L. [1 ]
Lienert, Ulrich [1 ]
Miceli, Antonino [1 ]
Vera, German
机构
[1] Argonne Natl Lab, XSD Adv Photon Source, Argonne, IL 60439 USA
[2] Los Alamos Natl Lab, Los Alamos, NM 87545 USA
关键词
X-ray detector; strain measurement; pair distribution function measurement;
D O I
10.1107/S090904950801755X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A GE Revolution 41RT flat-panel detector (GE 41RT) from GE Healthcare (GE) has been in operation at the Advanced Photon Source for over two years. The detector has an active area of 41. cm x 41 cm with 200 mu m x 200 mu m pixel size. The nominal working photon energy is around 80 keV The physical set-up and utility software of the detector system are discussed in this article. The linearity of the detector response was measured at 80.7 keV The memory effect of the detector element, called lag, was also measured at different exposure times and gain settings. The modulation transfer function was measured in terms of the line-spread function using a 25 mu m x 1 cm tungsten slit. The background (dark) signal, the signal that the detector will carry without exposure to X-rays, was measured at three different gain settings and with exposure times of 1 ms to 15 s. The radial geometric flatness of the sensor panel was measured using the diffraction pattern from a CeO2 powder standard. The large active area and fast data-capturing rate, i.e. 8 frames s(-1) in radiography mode, 30 frames s(-1) in fluoroscopy mode, make the GE 41RT one of a kind and very versatile in synchrotron diffraction. The loading behavior of a Cu/Nb multilayer material is used to demonstrate the use of the detector in a strain-stress experiment. Data from the measurement of various samples, amorphous SiO2 in particular, are presented to show the detector effectiveness in pair distribution function measurements.
引用
收藏
页码:477 / 488
页数:12
相关论文
共 50 条
  • [1] Fast imaging of a 41cm amorphous silicon flat-panel detector for radiographic applications
    Kump, KS
    [J]. MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 87 - 93
  • [2] Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography
    Floyd, CE
    Warp, RJ
    Dobbins, JT
    Chotas, HG
    Baydush, AH
    Vargas-Voracek, R
    Ravin, CE
    [J]. RADIOLOGY, 2001, 218 (03) : 683 - 688
  • [3] Non-uniformity correction technique for amorphous silicon flat-panel detector
    Liang, Li-Hong
    Wang, Xiao
    Lu, Hong-Nian
    [J]. Guangdian Gongcheng/Opto-Electronic Engineering, 2005, 32 (08): : 36 - 39
  • [4] Characterization of a flat-panel amorphous silicon detector for X-ray computed tomography
    Lee, H.-R.
    Ellingson, W.
    [J]. Journal of X-Ray Science and Technology, 2001, 9 (01) : 43 - 53
  • [5] Radiation dose reduction in chest radiography using a flat-panel amorphous silicon detector
    Hosch, WP
    Fink, C
    Radeleff, B
    Kampschulte, A
    Kauffmann, GW
    Hansmann, J
    [J]. CLINICAL RADIOLOGY, 2002, 57 (10) : 902 - 907
  • [6] Characterization and application of a GE amorphous silicon flat panel detector in a synchrotron light source
    Lee, J. H.
    Almer, J.
    Aydiner, C. C.
    Bernier, J.
    Chapman, K.
    Chupas, P.
    Haeffner, D.
    Kump, K.
    Lee, P. L.
    Lienert, U.
    Miceli, A.
    Vera, G.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 582 (01): : 182 - 184
  • [7] Image quality vs radiation dose for a flat-panel amorphous silicon detector: a phantom study
    Geijer, H
    Beckman, KW
    Andersson, T
    Persliden, J
    [J]. EUROPEAN RADIOLOGY, 2001, 11 (09) : 1704 - 1709
  • [8] Image quality vs radiation dose for a flat-panel amorphous silicon detector: a phantom study
    Håkan Geijer
    Karl-Wilhelm Beckman
    Torbjörn Andersson
    Jan Persliden
    [J]. European Radiology, 2001, 11 : 1704 - 1709
  • [9] Simulation of the response function of an amorphous silicon flat-panel array
    Cremers, F
    Kausch, C
    Albers, D
    Todorovic, M
    Schmidt, R
    [J]. ADVANCED MONTE CARLO FOR RADIATION PHYSICS, PARTICLE TRANSPORT SIMULATION AND APPLICATIONS, 2001, : 395 - 399
  • [10] Imaging performance of amorphous selenium flat-panel detector for digital fluoroscopy
    Hunt, DC
    Tousignanth, O
    Demers, Y
    Laperrière, L
    Rowlands, JA
    [J]. MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 226 - 234