Accuracy combining different brands of implants and abutments

被引:12
|
作者
Sola-Ruiz, Maria-Fernanda [1 ]
Selva-Otaolaurruchi, Eduardo [1 ]
Senent-Vicente, Gisela [1 ]
Gonzalez-de-Cossio, Ines [1 ]
Amigo-Borras, Vicente [2 ]
机构
[1] Univ Valencia, Fac Med & Dent, Dept Buccofacial Prosthet, E-46003 Valencia, Spain
[2] Univ Politecn Valencia, Dept Mech & Mat Engn, Valencia, Spain
来源
关键词
Vertical fit; implant; prosthetic abutment; combination; MARGINAL FIT; INTERFACE; MISFIT; PROSTHESES; STABILITY;
D O I
10.4317/medoral.18137
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To evaluate the vertical misfit between different brands of dental implants and prosthetic abutments, with or without mechanical torque, and to study their possible combination. Study design: Five different brands of implant were used in the study: Biofit (Castemaggiore, Italy), Bioner S. A. (Barcelona, Spain), 3i Biomet (Palm Beach, U. S. A.), BTI (Alava, Spain) and Nobel Biocare (Goteborg, Sweden), with standard 4.1 mm heads and external hexagons, and their respective machined prosthetic abutments. The implant-to-abutment fit/misfit was evaluated at four points (vestibular, lingual/palatine, mesial and distal) between implants and abutments of the same brand and different brands, with or without mechanical torque, using SEM micrographs at 5000X. Image analysis was performed using NIS-Elements software (Nikon Instruments Europe B. V.). Results: Before applying torque, vertical misfit (microgaps) of the different combinations tested varied between 1.6 and 5.4 microns and after applying torque, between 0.9 and 5.9 microns, an overall average of 3.46 +/- 2.96 microns. For manual assembly without the use of mechanical torque, the best results were obtained with the combination of the 3i implant and the BTI abutment. The Nobel implant and Nobel abutment, 3i-3i and BTI-BTI and the combination of 3i implant with BTI or Nobel abutment provided the best vertical fit when mechanical torque was applied. Conclusions: The vertical fits obtained were within the limits considered clinically acceptable. The application of mechanical torque improved outcomes. There is compatibility between implants and abutments of different brand and so their combination is a clinical possibility.
引用
收藏
页码:E332 / E336
页数:5
相关论文
共 50 条
  • [31] Biomechanical evaluation of anterior implants associated with titanium and zirconia abutments and monotype zirconia implants
    Brandao de Holanda, Kelynne Alcantara
    Caldas, Ricardo Armini
    Amaral, Marina
    da Silva Concilio, Lais Regiane
    Vitti, Rafael Pino
    JOURNAL OF PROSTHODONTIC RESEARCH, 2021, 65 (01) : 73 - 77
  • [32] Evaluation of Torsional Strength, Design, and Stress Distribution of Different Brands of Orthodontic Mini-Implants
    Lopes-Neto, Dario F.
    Rossi, Leandro B.
    Lunardi, Nadia
    Boeck, Eloisa M.
    Boeck-Neto, Rodolfo J.
    Luvizuto, Eloa R.
    JOURNAL OF CRANIOFACIAL SURGERY, 2015, 26 (05) : 1717 - 1718
  • [33] A positioning jig to verify the accuracy of implant abutments
    Al-Abbas, H
    Al-Ajmi, M
    Pipko, DJ
    JOURNAL OF PROSTHETIC DENTISTRY, 2002, 87 (01): : 115 - 116
  • [34] Machining accuracy of selected implant abutments.
    Kaus, T
    Benzing, U
    JOURNAL OF DENTAL RESEARCH, 1996, 75 : 1333 - 1333
  • [35] Retrospective comparison of posterior fixed dental prostheses supported by two different titanium abutments on tissue level implants
    Hsiao, Chih-Chien
    Liang, Chao-Hua
    Shen, Yu-Fu
    Hsu, Kuang-Wei
    JOURNAL OF PROSTHETIC DENTISTRY, 2021, 125 (06): : 877 - 882
  • [36] Compatibility of Nonoriginal Abutments With Implants: Evaluation of Microgap at the Implant-Abutment Interface, With Original and Nonoriginal Abutments
    Duraisamy, Revathi
    Krishnan, Chitra Shankar
    Ramasubramanian, Hariharan
    Sampathkumar, Jayakrishnakumar
    Mariappan, Saravanakumar
    Sivaprakasam, Azhagarasan Navarasampatti
    IMPLANT DENTISTRY, 2019, 28 (03) : 289 - 295
  • [37] Mechanical Complications Related to the Retention Screws of Prefabricated Metal Abutments With Different Angulations: A Retrospective Study With 916 Implants
    Araujo, Patricia Medeiros
    Schuldt Filho, Guenther
    Ferreira, Cimara Fortes
    Magalhaes Benfatti, Cesar Augusto
    Cagna, David Richard
    Bianchini, Marco Aurelio
    IMPLANT DENTISTRY, 2018, 27 (02) : 209 - 212
  • [38] Accuracy of Impression Techniques for Dental Implants Placed in Five Different Orientations
    Kim, Hyo-Seon
    Lee, Jae-Hyun
    Lee, Su Young
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2022, 37 (05) : 997 - 1002
  • [39] Microbiological features of stable osseointegrated implants used as abutments for overdentures
    Mombelli, A.
    Mericske-Stern, R.
    CLINICAL ORAL IMPLANTS RESEARCH, 1990, 1 (01) : 1 - 7
  • [40] A comparative analysis of the accuracy of different direct impression techniques for multiple implants
    Ongul, D.
    Gokcen-Rohlig, B.
    Sermet, B.
    Keskin, H.
    AUSTRALIAN DENTAL JOURNAL, 2012, 57 (02) : 184 - 189