Optimal fiducial points for pulse rate variability analysis from forehead and finger photoplethysmographic signals

被引:28
|
作者
Peralta, Elena [1 ,2 ]
Lazaro, Jesus [1 ,2 ,5 ]
Bailon, Raquel [1 ,2 ]
Marozas, Vaidotas [3 ,4 ]
Gil, Eduardo [1 ,2 ]
机构
[1] Univ Zaragoza, IIS Aragon, Aragon Inst Engn Res I3A, Biomed Signal Interpretat & Computat Simulat BSIC, Zaragoza, Spain
[2] CIBER BBN, Zaragoza, Spain
[3] Kaunas Univ Technol, Biomed Engn Inst, Kaunas, Lithuania
[4] Kaunas Univ Technol, Elect Engn Dept, Kaunas, Lithuania
[5] Univ Connecticut, Dept Biomed Engn, Storrs, CT USA
基金
欧盟地平线“2020”;
关键词
photoplethysmography (PPG); ECG; heart rate variability; pulse rate variability; autonomic nervous system; fiducial point selection; transmission and reflection modes; HEART-RATE-VARIABILITY; SHORT-TERM ANALYSIS; RELIABILITY; SURROGATE;
D O I
10.1088/1361-6579/ab009b
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Objective: The aim of this work is to evaluate and compare five fiducial points for the temporal location of each pulse wave from forehead and finger photoplethysmographic (PPG) pulse wave signals to perform pulse rate variability (PRV) analysis as a surrogate for heart rate variability (HRV) analysis. Approach: Forehead and finger PPG signals were recorded during a tilt-table test simultaneously with the electrocardiogram (ECG). Artefacts were detected and removed and five fiducial points were computed: apex, middle-amplitude and foot points of the PPG signal, apex point of the first derivative signal and the intersection point of the tangent to the PPG waveform at the apex of the derivative PPG signal and the tangent to the foot of the PPG pulse, defined as the intersecting tangents method. Pulse period (PP) time interval series were obtained from both PPG signals and compared with the RR intervals obtained from the ECG. HRV and PRV signals were estimated and classical time and frequency domain indices were computed. Main results: The middle-amplitude point of the PPG signal (n(M)), the apex point of the first derivative (n(A)*), and the tangent intersection point (n(T)) are the most suitable fiducial points for PRV analysis, resulting in the lowest relative errors estimated between PRV and HRV indices and higher correlation coefficients and reliability indices. Statistically significant differences according to the Wilcoxon test between PRV and HRV signals were found for the apex and foot fiducial points of the PPG, as well as the lowest agreement between RR and PP series according to Bland-Altman analysis. Hence, these signals have been considered less accurate for variability analysis. In addition, the relative errors are significantly lower for n(M) and n(A)(*) using Friedman statistics with a Bonferroni multiple-comparison test, and we propose that n(M) is the most accurate fiducial point. Based on our results, forehead PPG seems to provide more reliable information for a PRV assessment than finger PPG. Significance: The accuracy of the pulse wave detection depends on the morphology of the PPG. There is therefore a need to widely define the most accurate fiducial point for performing a PRV analysis under non-stationary conditions based on different PPG sensor locations and signal acquisition techniques.
引用
下载
收藏
页数:15
相关论文
共 50 条
  • [1] Spectral analysis for pulse rate variability assessment from simulated photoplethysmographic signals
    Mejia-Mejia, Elisa
    Kyriacou, Panicos A.
    FRONTIERS IN PHYSIOLOGY, 2022, 13
  • [2] Robust Pulse Rate Variability Analysis from Reflection and Transmission Photoplethysmographic Signals
    Peralta, Elena
    Lazaro, Jesus
    Gil, Eduardo
    Bailon, Raquel
    Marozas, Vaidotas
    2017 COMPUTING IN CARDIOLOGY (CINC), 2017, 44
  • [3] Duration of photoplethysmographic signals for the extraction of Pulse Rate Variability Indices
    Mejia-Mejia, Elisa
    Kyriacou, Panicos A.
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2023, 80
  • [4] On the Heart rate variability and Pulse rate variability evaluations based on electrocardiographic and photoplethysmographic signals
    Podaru, Alexandru Constantin
    David, Valeriu
    2021 12TH INTERNATIONAL SYMPOSIUM ON ADVANCED TOPICS IN ELECTRICAL ENGINEERING (ATEE), 2021,
  • [5] Photoplethysmographic Waveform and Pulse Rate Variability Analysis in Hyperbaric Environments
    Pelaez-Coca, Maria Dolores
    Hernando, Alberto
    Lozano, Maria Teresa
    Sanchez, Carlos
    Izquierdo, David
    Gil, Eduardo
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2021, 25 (05) : 1550 - 1560
  • [6] Contour analysis of the photoplethysmographic pulse measured at the finger
    Millasseau, Sandrine C.
    Ritter, James M.
    Takazawa, Kenji
    Chowienczyk, Philip J.
    JOURNAL OF HYPERTENSION, 2006, 24 (08) : 1449 - 1456
  • [7] Impact of the PPG Sampling Rate in the Pulse Rate Variability Indices Evaluating Several Fiducial Points in Different Pulse Waveforms
    Dolores Pelaez-Coca, Maria
    Hernando, Alberto
    Lazaro, Jesus
    Gil, Eduardo
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2022, 26 (02) : 539 - 549
  • [8] Finger and Forehead PPG Signal Comparison for Respiratory Rate Estimation Based on Pulse Amplitude Variability
    Hernando, Alberto
    Dolores Pelaez, Maria
    Teresa Lozano, Maria
    Aiger, Montserrat
    Gil, Eduardo
    Lazaro, Jesus
    2017 25TH EUROPEAN SIGNAL PROCESSING CONFERENCE (EUSIPCO), 2017, : 2076 - 2080
  • [9] A NEW CONTOUR ANALYSIS OF PHOTOPLETHYSMOGRAPHIC PULSE MEASURED AT THE FINGER
    Wu, H. -T.
    Lin, B. -S.
    Liu, C. -C.
    Liu, A. -B.
    ATHEROSCLEROSIS SUPPLEMENTS, 2010, 11 (02) : 190 - 190
  • [10] Evaluation of pulse rate variability obtained by the pulse onsets of the photoplethysmographic signal
    Posada-Quintero, H. F.
    Delisle-Rodriguez, D.
    Cuadra-Sanz, M. B.
    Fernandez de la Vara-Prieto, R. R.
    PHYSIOLOGICAL MEASUREMENT, 2013, 34 (02) : 179 - 187