Estrogens determination in wastewater samples by automatic in-syringe dispersive liquid-liquid microextraction prior silylation and gas chromatography

被引:36
|
作者
Gonzalez, Alba [1 ]
Avivar, Jessica [2 ]
Cerda, Victor [1 ]
机构
[1] Univ Balearic Isl, Dept Chem, Grp Analyt Chem Automat & Environm, E-07122 Palma De Mallorca, Spain
[2] Univ Balearic Isl, SL Spin, Sciware Syst, E-07193 Bunyola, Spain
关键词
In-syringe magnetic stirring-assisted dispersive liquid-liquid microextraction (in-syringe-MSA-DLLME); Gas chromatography (GC); Estrogens; Wastewater samples; Automation; MASS-SPECTROMETRY; IONIC LIQUID; 17-ALPHA-ETHINYLESTRADIOL; DERIVATIZATION; OPTIMIZATION; HORMONES; ESTRONE;
D O I
10.1016/j.chroma.2015.08.031
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new procedure for the extraction, preconcentration and simultaneous determination of the estrogens most used in contraception pharmaceuticals (estrone, 17 beta-estradiol, estriol, and 17 alpha-ethynylestradiol), cataloged as Contaminants of Emergent Concern by the Environmental Protection Agency of the United States (US-EPA), is proposed. The developed system performs an in-syringe magnetic stirring-assisted dispersive liquid-liquid microextraction (in-syringe-MSA-DLLME) prior derivatization and gas chromatography (GC-MS). Different extraction (carbon tetrachloride, ethyl acetate, chloroform and trichloroethylene) and disperser solvents (acetone, acetonitrile and methanol) were tested. Chloroform and acetone were chosen as extraction and disperser solvent, respectively, as they provided the best extraction efficiency. Then, a multivariate optimization of the extraction conditions was carried out. Derivatization conditions were also studied to ensure the conversion of the estrogens to their respective trimethylsilyl derivatives. Low LODs and LOQs were achieved, i.e. between 11 and 82 ng L-1, and 37 and 272 ng L-1, respectively. Good values for intra and inter-day precision were obtained (RSDs <= 7.06% and RSD <= 7.11%, respectively). The method was successfully applied to wastewater samples. (C) 2015 Elsevier B.V. All rights reserved.
引用
下载
收藏
页码:1 / 8
页数:8
相关论文
共 50 条
  • [41] Determination of ppb-level phenol index using in-syringe dispersive liquid-liquid microextraction and liquid waveguide capillary cell spectrophotometry
    Burkhard Horstkotte
    Fernando Maya
    Carlos M. Duarte
    Víctor Cerdà
    Microchimica Acta, 2012, 179 : 91 - 98
  • [42] Automatic determination of copper by in-syringe dispersive liquid-liquid microextraction of its bathocuproine-complex using long path-length spectrophotometric detection
    Horstkotte, Burkhard
    Alexovic, Michal
    Maya, Fernando
    Duarte, Carlos M.
    Andruch, Vasil
    Cerda, Victor
    TALANTA, 2012, 99 : 349 - 356
  • [43] Determination of Phorate in Water Using Dispersive Liquid-Liquid Microextraction Coupled with Gas Chromatography
    Xie Hong-Xue
    He Li-Jun
    Wu Xiu-Ling
    Fan Lu
    Lu Kui
    Wang Mei-Mei
    Meng Da-Wei
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2008, 36 (11) : 1543 - 1546
  • [44] DETERMINATION OF PHTHALATE ESTERS IN WINE USING DISPERSIVE LIQUID-LIQUID MICROEXTRACTION AND GAS CHROMATOGRAPHY
    Zhu, Hualing
    Cui, Shumin
    Wang, Weiping
    Miao, Jigen
    Feng, Jiuju
    Chen, Jianrong
    ANALYTICAL LETTERS, 2014, 47 (11) : 1874 - 1887
  • [45] Determination of Three Estrogens in Environmental Water Samples Using Dispersive Liquid-Liquid Microextraction by High-Performance Liquid Chromatography and Fluorescence Detector
    Érika M. L. Sousa
    Reyla A. S. Dias
    Eliane R. Sousa
    Natilene M. Brito
    Arlan S. Freitas
    Gilberto S. Silva
    Lanna K. Silva
    Diana L. D Lima
    Valdemar I. Esteves
    Gilmar S. Silva
    Water, Air, & Soil Pollution, 2020, 231
  • [46] Determination of Psychotropic Drugs by Gas Chromatography/Mass Spectrometry with Dispersive Liquid-Liquid Microextraction
    Kohiyama, Ayako
    Sato, Masako
    Kaneko, Tsuyoshi
    BUNSEKI KAGAKU, 2009, 58 (08) : 661 - 665
  • [47] Determination of Three Estrogens in Environmental Water Samples Using Dispersive Liquid-Liquid Microextraction by High-Performance Liquid Chromatography and Fluorescence Detector
    Sousa, Erika M. L.
    Dias, Reyla A. S.
    Sousa, Eliane R.
    Brito, Natilene M.
    Freitas, Arlan S.
    Silva, Gilberto S.
    Silva, Lanna K.
    Lima, Diana L. D.
    Esteves, Valdemar, I
    Silva, Gilmar S.
    WATER AIR AND SOIL POLLUTION, 2020, 231 (04):
  • [48] Automated method for volatile fatty acids determination in anaerobic processes using in-syringe magnetic stirring assisted dispersive liquid-liquid microextraction and gas chromatography with flame ionization detector
    Vargas-Munoz, M. A.
    Cerda, Victor
    Cadavid-Rodriguez, L. S.
    Palacio, Edwin
    JOURNAL OF CHROMATOGRAPHY A, 2021, 1643
  • [49] Combination of dispersive solid phase extraction and dispersive liquid-liquid microextraction for extraction of some aryloxy pesticides prior to their determination by gas chromatography
    Farajzadeh, Mir Ali
    Yadeghari, Adeleh
    Khoshmaram, Leila
    MICROCHEMICAL JOURNAL, 2017, 131 : 182 - 191
  • [50] Analysis of pyrethroid pesticides in water samples by dispersive liquid-liquid microextraction coupled with gas chromatography
    Zang Xiao-Huan
    Wang Chun
    Gao Shu-Tao
    Zhou Xin
    Wang Zhi
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2008, 36 (06) : 765 - 769