Long-term hermeticity and biological performance of anodically bonded glass-silicon implantable packages

被引:16
|
作者
Harpster, TJ [1 ]
Nikles, SA [1 ]
Dokmeci, MR [1 ]
Najafi, K [1 ]
机构
[1] Univ Michigan, Ctr Wireless Integrated Microsyst, Ann Arbor, MI 48109 USA
基金
美国国家卫生研究院;
关键词
biocompatible package; hermetic package; in vivo testing; microsystems packages;
D O I
10.1109/TDMR.2005.854374
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper reviews long-term test results obtained from a series of tests on glass-silicon (Si) hermetically sealed packages. Results are presented from 1) a 9.9-year ongoing room temperature phosphate-buffered saline (PBS) soak test of four packages; 2) accelerated soak tests in high temperature saline of 28 samples resulting in an extrapolated mean-time-to-failure (MTTF) at 37 degrees C of 177 years; 3) a 2.7-year in vitro 97 degrees C PBS soak test of a single package; and 4) in situ hermeticity and biocompatibility tests from 12 packages implanted in four guinea pigs-three packages implanted in two guinea pigs (each) for 1 month and another two guinea pigs for 20 and 22 months. All of the packages remained hermetically sealed over the lifetime of the implant. A detailed histological report of the implants is provided suggesting that they elicit no profound adverse reaction from the body.
引用
收藏
页码:458 / 466
页数:9
相关论文
共 50 条
  • [21] The tensile properties and creep performance of a long-term thermally aged plastic bonded explosive
    Zhou H.-P.
    He Q.
    Li M.
    Pang H.-Y.
    Wei X.-W.
    Wen M.-P.
    Li, Ming (li_ming@caep.cn), 1600, Institute of Chemical Materials, China Academy of Engineering Physics (24): : 826 - 831
  • [22] ENCAPSULATED, IMPLANTABLE METAL-OXYGEN CELL AS A LONG-TERM POWER SOURCE FOR MEDICAL AND BIOLOGICAL APPLICATIONS
    TSEUNG, ACC
    KING, WJ
    WAN, BYC
    MEDICAL & BIOLOGICAL ENGINEERING, 1971, 9 (03): : 175 - &
  • [23] Improvements in long-term output energy performance of Nd:glass regenerative amplifiers
    Peng Zhang
    Youen Jiang
    Jiangfeng Wang
    Wei Fan
    Xuechun Li
    Jianqiang Zhu
    High Power Laser Science and Engineering, 2017, 5 (04) : 5 - 11
  • [24] Improvements in long-term output energy performance of Nd:glass regenerative amplifiers
    Zhang, Peng
    Jiang, Youen
    Wang, Jiangfeng
    Fan, Wei
    Li, Xuechun
    Zhu, Jianqiang
    HIGH POWER LASER SCIENCE AND ENGINEERING, 2017, 5
  • [25] Long-term performance of a glass fiber-reinforced polymer truss bridge
    Keller, Thomas
    Bai, Yu
    Vallee, Till
    JOURNAL OF COMPOSITES FOR CONSTRUCTION, 2007, 11 (01) : 99 - 108
  • [26] A changing scenario in the clinical use of implantable defibrillators: the need for long-term data on lead performance
    Boriani, Giuseppe
    Biffi, Mauro
    Marziali, Alessandro
    Diemberger, Igor
    Martignani, Cristian
    EUROPACE, 2009, 11 (01): : 1 - 3
  • [27] CALIBRATION POINT ACCURACY CHARACTERIZATION ON THE PERFORMANCE OF A LONG-TERM, IMPLANTABLE CGM: A SUBANALYSIS OF THE PRECISE STUDY
    Kropff, J.
    DeVries, J. H.
    Rastogi, R.
    Dehennis, A.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2016, 18 : A75 - A75
  • [28] Long-term performance of single-connector (DF4) implantable defibrillator leads
    Ibrahim, Rand
    Al-Gibbawi, Mounir
    Mekary, Wissam
    Bhatia, Neal Kumar
    Kiani, Soroosh
    Westerman, Stacy B.
    Shah, Anand D.
    Lloyd, Michael S.
    Leal, Miguel
    De Lurgio, David B.
    Patel, Anshul M.
    Tompkins, Christine
    Leon, Angel R.
    Merchant, Faisal M.
    El-Chami, Mikhael F.
    EUROPACE, 2023, 25 (12):
  • [29] Experimental study on long-term mechanical performance of retard-bonded prestressed concrete beams
    Cao, Guo-Hui
    Hu, Jia-Xing
    Zhang, Kai
    Zhang, Xian
    Cao, G.-H. (cgfcivil@163.com), 1600, Science Press (35): : 107 - 111
  • [30] Effects of moisture on the long-term performance of adhesively bonded FRP/steel joints used in bridges
    Heshmati, Mohsen
    Haghani, Reza
    Al-Emrani, Mohammad
    COMPOSITES PART B-ENGINEERING, 2016, 92 : 447 - 462