A fiberoptic sensor for tissue carbon dioxide monitoring

被引:0
|
作者
Davenport, John J. [1 ]
Hickey, Michelle [1 ]
Phillips, Justin P. [1 ]
Kyriacou, Panicos A. [1 ]
机构
[1] City Univ London, Sch Comp Sci Math & Engn, London EC1V 0HB, England
基金
美国国家卫生研究院;
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present a new fiberoptic carbon dioxide sensor for transcutaneous and mucosa (indwelling) blood gas monitoring. The sensor is based on optical fluorescence of molecules sensitive to pH changes associated with dissolved CO2. A three layer chemical coating was dip-coated onto the distal tip of an optical fiber (600 mu m core radius). It contained the 50mg/ml 'polym H7', a coating polymer bonded to a fluorescence indicator dye, along with 125mg/ml of the transfer agent tetraoctylammonium hydroxide (TONOH). Light from a blue (460 nm) LED was launched into the fiber to excite the sensing film. The sensing film fluoresced green (530 nm), the intensity of which decreased in the presence of CO2. The sensor was tested in vitro, finding a correlation between change in fluorescence (in AU) and aqueous CO2 concentration with a minimum detection threshold of 40%. The sensor is being developed for medical applications where its small size and ability to continuously monitor the partial pressure of CO2 (PCO2) will make it an extremely useful diagnostic tool.
引用
收藏
页码:7942 / 7945
页数:4
相关论文
共 50 条
  • [31] A FIBEROPTIC SENSOR SYSTEM FOR MONITORING CHLORINATED-HYDROCARBON POLLUTANTS
    MILANOVICH, FP
    BROWN, SB
    COLSTON, BW
    DALEY, PF
    LANGRY, KC
    [J]. TALANTA, 1994, 41 (12) : 2189 - 2194
  • [32] CARBON-DIOXIDE MONITORING
    LILLIE, PE
    ROBERTS, JG
    [J]. ANAESTHESIA AND INTENSIVE CARE, 1988, 16 (01) : 41 - 44
  • [33] Transcutaneous carbon dioxide monitoring
    McCormack, J. G.
    [J]. ANAESTHESIA, 2007, 62 (08) : 850 - 851
  • [34] Monitoring Exhaled Carbon Dioxide
    Siobal, Mark S.
    [J]. RESPIRATORY CARE, 2016, 61 (10) : 1397 - 1416
  • [35] Unattended carbon dioxide monitoring
    不详
    [J]. BRITISH CORROSION JOURNAL, 1997, 32 (01): : 17 - 17
  • [36] Miniaturized Photoacoustic Carbon Dioxide Sensor with Integrated Temperature Compensation for Room Climate Monitoring
    Huber, J.
    Ambs, A.
    Woellenstein, J.
    [J]. EUROSENSORS 2015, 2015, 120 : 283 - 288
  • [37] A Low-Cost Carbon Dioxide Monitoring System for Coastal and Estuarine Sensor Networks
    Bresnahan, Philip J.
    Farquhar, Elizabeth
    Portelli, Daniel
    Tydings, Michael
    Wirth, Taylor
    Martz, Todd
    [J]. OCEANOGRAPHY, 2023, 36 (01) : 8 - 8
  • [38] A novel planar optical sensor for simultaneous monitoring of oxygen, carbon dioxide, pH and temperature
    Sergey M. Borisov
    Roman Seifner
    Ingo Klimant
    [J]. Analytical and Bioanalytical Chemistry, 2011, 400 : 2463 - 2474
  • [39] A novel planar optical sensor for simultaneous monitoring of oxygen, carbon dioxide, pH and temperature
    Borisov, Sergey M.
    Seifner, Roman
    Klimant, Ingo
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (08) : 2463 - 2474
  • [40] Carbon dioxide/methane gas sensor based on the permselectivity of polymeric membranes for biogas monitoring
    Rego, R
    Mendes, A
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2004, 103 (1-2) : 2 - 6