Sample temperature effect on laser ablation and analytical capabilities of laser induced breakdown spectroscopy

被引:36
|
作者
Lednev, V. N. [1 ,2 ]
Grishin, M. Ya. [1 ,3 ]
Sdvizhenskii, P. A. [1 ,2 ]
Asyutin, R. D. [4 ]
Tretyakov, R. S. [4 ]
Stavertiy, A. Ya. [4 ]
Pershin, S. M. [1 ]
机构
[1] Russian Acad Sci, Prokhorov Gen Phys Inst, Moscow, Russia
[2] Natl Univ Sci & Technol MISiS, Moscow, Russia
[3] State Univ, Moscow Inst Phys & Technol, Dolgoprudnyi, Moscow Region, Russia
[4] Bauman Moscow State Tech Univ, Moscow, Russia
基金
俄罗斯科学基金会;
关键词
QUANTITATIVE-ANALYSIS; OPTICAL-EMISSION; LIQUID STEEL; ALLOY-STEEL; MOLTEN; METAL; LIBS; SPECTROMETRY; PLASMAS;
D O I
10.1039/c8ja00348c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The influence of sample temperature on the laser ablation process and analytical capabilities of laser induced breakdown spectroscopy (LIBS) was studied systematically. We were interested in estimating the analytical capabilities of a LIBS probe which was developed for in situ elemental analysis of molten metals during the production of metal parts by additive manufacturing. The sample surface was heated locally (4 mm diameter spot) in an argon atmosphere using a powerful continuous wave fiber laser while a nanosecond laser ablated only a small spot (0.3 mm diameter) in the high temperature area. This procedure can be utilized for high temperature LIBS studies of any samples including materials with high melting temperatures (tungsten, aluminum oxide etc.). Low alloy steel samples were heated up to their melting points and LIBS experiments were carried out in a wide temperature range (25-1600 degrees C). LIBS plasma imaging revealed that the size and brightness of the expanding plume did not depend on solid sample temperature but these parameters increased manifold for molten steel ablation. Comparison of plasma emissivity, temperature and electron density did not reveal any meaningful differences in the case of LIBS measurements for room temperature (25 degrees C), hot solid (1050 degrees C) and molten (1550 degrees C) targets, and thus enhanced plasma emissivity was attributed to greater ablation rate. The surface temperature impact on LIBS analytical capabilities was studied by comparing calibration curve linearity, limits of detection and the root mean square error of cross-validation for chromium, silicon and manganese. High temperature of the sample surface favored greater atomic/ionic line emission but molten sample LIBS analysis revealed poor limits of detection for all elements, which was explained by strong fluctuations of the laser ablation process at the liquid metal surface.
引用
收藏
页码:607 / 615
页数:9
相关论文
共 50 条
  • [21] Laser-Induced Breakdown Spectroscopy: A Closer Look at the Capabilities of LIBS
    Hark, Richard R.
    SPECTROSCOPY, 2015, : 29 - 30
  • [22] Double-pulse laser ablation coupled to laser-induced breakdown spectroscopy
    Glaus, Reto
    Hahn, David W.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2014, 98 : 48 - 53
  • [23] Laser ablation of powdered samples and analysis by means of laser-induced breakdown spectroscopy
    Ctvrtnickova, T.
    Cabalin, L.
    Laserna, J.
    Kanicky, V.
    Nicolas, G.
    APPLIED SURFACE SCIENCE, 2009, 255 (10) : 5329 - 5333
  • [24] Double pulse laser ablation and plasma: Laser induced breakdown spectroscopy signal enhancement
    Babushok, V. I.
    DeLucia, F. C., Jr.
    Gottfried, J. L.
    Munson, C. A.
    Miziolek, A. W.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2006, 61 (09) : 999 - 1014
  • [26] The effect of ablation angle on intensity of laser-induced breakdown spectroscopy under vacuum
    Jia, Yueying
    Wang, Yiqin
    Gao, Lu
    Chen, Ting
    Xiao, Qingmei
    Wu, Ding
    PHYSICA SCRIPTA, 2021, 96 (12)
  • [27] Effect of sample temperature on laser-induced semiconductor plasma spectroscopy
    刘杨
    佟悦
    李苏宇
    王莹
    陈安民
    金明星
    Chinese Optics Letters, 2016, 14 (12) : 136 - 140
  • [28] Effect of sample temperature on laser-induced semiconductor plasma spectroscopy
    Liu, Yang
    Tong, Yue
    Li, Suyu
    Wang, Ying
    Chen, Anmin
    Jin, Mingxing
    CHINESE OPTICS LETTERS, 2016, 14 (12)
  • [29] The effect of sample surface roughness on the microanalysis of microchip laser-induced breakdown spectroscopy
    Wang, Wei
    Sun, Lanxiang
    Wang, Guodong
    Zhang, Peng
    Qi, Lifeng
    Zheng, Liming
    Dong, Wei
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (02) : 357 - 365
  • [30] Effect of sample temperature on femtosecond laser ablation of copper
    Dang, Wei-Jie
    Chen, Yu-Tong
    Chen, An-Min
    Jin, Ming-Xing
    CHINESE PHYSICS B, 2024, 33 (02)