The validity of emergency department triage blood pressure measurements

被引:19
|
作者
Cienki, JJ [1 ]
DeLuca, LA [1 ]
Daniel, N [1 ]
机构
[1] Univ Miami, Jackson Mem Hosp, Sch Med, Miami, FL 33316 USA
关键词
blood pressure; emergency medical services; triage; hypertension; automated blood pressure;
D O I
10.1197/j.aem.2003.10.021
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: Automated blood pressure (ABP) devices are ubiquitous at emergency department (ED) triage. Previous studies failed to evaluate ABP devices against accepted reference standards or demonstrate triage readings as accurate reflections of blood pressure (BP). This study evaluated ED triage measurements made using an ABP device and assessed agreement between triage BP and BP taken under recommended conditions. Methods: A prospective study was conducted at an urban teaching hospital. Patients were enrolled by convenience sampling. Simultaneous automated and manual triage BPs were obtained using one BP cuff with a Y-tube connector. Research assistants were certified in obtaining manual BP as described by the British Hypertension Society (BHS). Patients were placed in a quiet setting, and manual BP was repeated by American Heart Association (AHA) standards. Data analysis was performed using methods described by Bland and Altman. The ABP device was assessed using Association for the Advancement of Medical Instrumentation (AAMI) and BHS criteria. Results: One hundred seventy-one patients were enrolled. Systolic BP (sBP) range was 81 to 218 mm Hg; diastolic BP (dBP) range was 43 to 130 mm Hg. Automated vs. manual sBP difference was 3.8 +/- 11.2 mm Hg (95% confidence interval [CI] = 2.1 to 5.4); dBP difference was 6.6 +/- 9.0 mm Hg (95% CI = -7.9 to -5.2). Manual triage BP vs. AHA standard SBP difference was 11.6 +/- 12.8 mm Hg (95% CI = 9.1 to 14.1); dBP difference was 9.9 +/- 10.4 mm Hg (95% CI = 7.9 to 12.0). The ABP device failed to meet AAMI criteria and received a BHS rating of "D." Poor operator technique and extraneous patient and operator movement appeared to hamper accuracy. Conclusions: ABP triage measurements show significant discrepancies from a reference standard. Repeat measurements following AHA standards demonstrate significant decreases in the measured blood pressures.
引用
收藏
页码:237 / 243
页数:7
相关论文
共 50 条
  • [31] Implementation of triage in a pediatric emergency department
    Uma Priya Narayanan
    Swathi Padankatti
    Kuruvilla Thomas
    International Journal of Emergency Medicine, 7 (Suppl 1)
  • [32] Emergency department triage: An ethical analysis
    Aacharya R.P.
    Gastmans C.
    Denier Y.
    BMC Emergency Medicine, 11 (1)
  • [33] How reliable is emergency department triage?
    Fernandes, CMB
    Wuerz, R
    Clark, S
    Djurdjev, O
    ANNALS OF EMERGENCY MEDICINE, 1999, 34 (02) : 141 - 147
  • [34] Emergency Department Physician Telemedical Triage
    Traub, Stephen J.
    Butler, Rebecca
    Chang, Yu-Hui
    Lipinski, Christopher
    TELEMEDICINE AND E-HEALTH, 2013, 19 (11) : 841 - 845
  • [35] NURSE TRIAGE IN THE ACCIDENT AND EMERGENCY DEPARTMENT
    YATES, DW
    JOURNAL OF THE ROYAL SOCIETY OF HEALTH, 1987, 107 (04): : 153 - 154
  • [36] Interobserver agreement in emergency department triage
    Nakagawa, J
    Ouk, S
    Schwartz, B
    Schriger, DL
    ANNALS OF EMERGENCY MEDICINE, 2003, 41 (02) : 191 - 195
  • [37] RACIAL DISPARITY IN EMERGENCY DEPARTMENT TRIAGE
    Schrader, Chet D.
    Lewis, Lawrence M.
    JOURNAL OF EMERGENCY MEDICINE, 2013, 44 (02): : 511 - 518
  • [38] TRIAGE AND SATISFACTION IN THE PEDIATRIC EMERGENCY DEPARTMENT
    SANTER, L
    CALHOUN, A
    CLINICAL RESEARCH, 1993, 41 (04): : A759 - A759
  • [39] Triage tools in the paediatric emergency department
    不详
    ARCHIVES OF DISEASE IN CHILDHOOD, 2024, 109 (12) : 979 - 979
  • [40] THE EMERGENCY DEPARTMENT AND TRIAGE OF NONURGENT PATIENTS
    DERLET, RW
    KINSER, D
    ANNALS OF EMERGENCY MEDICINE, 1994, 23 (02) : 377 - 379