Evaluation of optical radiation detectors in the range from 0.8 μm to 20 μm at the NIST infrared spectral calibration facility

被引:3
|
作者
Podobedov, Vyacheslav B. [1 ]
Eppeldauer, George P. [1 ]
Larason, Thomas C. [1 ]
机构
[1] NIST, Gaithersburg, MD 20899 USA
来源
OPTICAL SYSTEMS DESIGN 2012 | 2012年 / 8550卷
关键词
Detector; IR; irradiance; calibration; spectral responsivity; radiant power; radiometry;
D O I
10.1117/12.980937
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Analysis of multiple factors affecting the uncertainty of the absolute spectral responsivity of optical radiation detectors is presented. This includes both the validation of the radiometric scale of the infrared reference detectors and the scale transfer process to the unit under test. Reference detectors include a low NEP pyroelectric detector, an InSb detector, and a sphere-input extended InGaAs detector. While all three types of reference detectors were calibrated independently, less than 0.5 % mismatch of spectral responsivities was observed in the spectrally overlapping regions. We provide a performance evaluation of the NIST IR Detector Calibration Facility, which was designed for testing optical radiation detectors in both radiant power and irradiance measurement modes. This facility utilizes a high throughput monochromator with interchangeable diffraction gratings. Depending on the spectral range, a blackbody at 1100 degrees C, or a quartz halogen lamp with about a 10(-4) long-term relative output variation was used as a radiation source. In order to minimize the uncertainty budget for calibration data, specific attention was given to the profile of the incident beam, the precise positioning of detectors, and the influence of atmospheric absorption. In addition to the spectral responsivity calibration of detectors, the facility allows precise mapping of detector active area for spatial non-uniformity of response. Typical calibration uncertainties that can be achieved are about 1 % and 2.5 % (k=2) in the radiant power and irradiance modes, respectively. Examples for responsivity calibrations of different detectors are presented.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] A GaN/AlN quantum cascade detector with a broad response from the mid-infrared (4.1μm) to the visible (550nm) spectral range
    Quach, P.
    Liu, S. F.
    Jollivet, A.
    Wang, D.
    Cheng, J. Y.
    Isac, N.
    Pirotta, S.
    Bouville, D.
    Sheng, S. S.
    Imran, A.
    Chen, L.
    Li, D.
    Zheng, X. T.
    Wang, Y. X.
    Qin, Z. X.
    Tchernycheva, M.
    Julien, F. H.
    Shen, B.
    Wang, X. Q.
    APPLIED PHYSICS LETTERS, 2020, 116 (17)
  • [32] Mid-infrared optical frequency comb in the 2.7–4.0 μm range via difference frequency generation from a compact laser system
    Lian Zhou
    Yang Liu
    Gehui Xie
    Chenglin Gu
    Zejiang Deng
    Zhiwei Zhu
    Cheng Ouyang
    Zhong Zuo
    Daping Luo
    Bin Wu
    Kunfeng Chen
    Wenxue Li
    HighPowerLaserScienceandEngineering, 2020, 8 (04) : 21 - 26
  • [33] Light emitting diodes for the spectral range of λ=3.3–4.3 μm fabricated from the InGaAs-and InAsSbP-based solid solutions: Electroluminescence in the temperature range of 20–180°C
    M. Aidaraliev
    N. V. Zotova
    S. A. Karandashev
    B. A. Matveev
    M. A. Remennyi
    N. M. Stus’
    G. N. Talalakin
    Semiconductors, 2000, 34 : 104 - 107
  • [34] Effect of an aerosol on the spectral and energy characteristics of radiation from the earth's surface, tropospheric clouds, and the atmosphere observed from space in the range 1-3 μm
    Veselov, DP
    Lobanova, GI
    Mirzoeva, LA
    Popov, OI
    Semenova, VI
    JOURNAL OF OPTICAL TECHNOLOGY, 1999, 66 (04) : 308 - 312
  • [35] Effect of an aerosol on the spectral and energy characteristics of radiation from the earth's surface, tropospheric clouds, and the atmosphere observed from space in the range 1-3 μm
    Veselov, D.P.
    Lobanova, G.I.
    Mirzoeva, L.A.
    Popov, O.I.
    Semenova, V.I.
    Journal of Optical Technology (A Translation of Opticheskii Zhurnal), 1999, 66 (04): : 308 - 312
  • [36] Evaluation of optical linear and non-linear parameters of thermally deposited GeTeSeGa thin films in NIR (1 μm-2.6 μm) wavelength range from their transmission spectra
    Sharma, Ekta
    Barman, P. B.
    Sharma, Pankaj
    OPTIK, 2020, 219
  • [37] A new method of determining the optical constants of a thin film from its reflectance and transmittance interferograms in a wide spectral range:: 0.2-3 μm
    Khashan, MA
    El-Naggar, AM
    Shaddad, E
    OPTICS COMMUNICATIONS, 2000, 178 (1-3) : 123 - 132
  • [38] Mid-infrared optical frequency comb in the 2.7-4.0 μm range via difference frequency generation from a compact laser system
    Zhou, Lian
    Liu, Yang
    Xie, Gehui
    Gu, Chenglin
    Deng, Zejiang
    Zhu, Zhiwei
    Ouyang, Cheng
    Zuo, Zhong
    Luo, Daping
    Wu, Bin
    Chen, Kunfeng
    Li, Wenxue
    HIGH POWER LASER SCIENCE AND ENGINEERING, 2020, 8
  • [39] Selective doping of Ga20Ge20Se60 glass-ceramic matrix with lanthanum(III) selenide to increase its optical transparency in the 2-10 μ m spectral range
    Sukhanov, M. V.
    Blagin, R. D.
    Velmuzhov, A. P.
    Fukina, D. G.
    Sidorenko, K. V.
    Stepanov, B. S.
    Shiryaev, V. S.
    JOURNAL OF NON-CRYSTALLINE SOLIDS, 2025, 647
  • [40] Light emitting diodes for the spectral range λ=3.3-4.3 μm fabricated from InGaAs and InAsSbP solid solutions:: Electroluminescence in the temperature range of 20-180°C (Part 2)
    Aidaraliev, M
    Zotova, NV
    Karandashev, SA
    Matveev, BA
    Remennyi, MA
    Stus', NM
    Talalakin, GN
    SEMICONDUCTORS, 2001, 35 (05) : 598 - 604