Continuous glucose monitoring: quality of hypoglycaemia detection

被引:35
|
作者
Zijlstra, E. [1 ]
Heise, T. [1 ]
Nosek, L. [1 ]
Heinemann, L. [1 ]
Heckermann, S. [1 ]
机构
[1] Profil, Neuss, Germany
来源
DIABETES OBESITY & METABOLISM | 2013年 / 15卷 / 02期
关键词
clinical trial; glycaemic control; type; 1; diabetes; CLINICAL ACCURACY;
D O I
10.1111/dom.12001
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims To evaluate the accuracy of a (widely used) continuous glucose monitoring (CGM)-system and its ability to detect hypoglycaemic events. Methods A total of 18 patients with type 1 diabetes mellitus used continuous glucose monitoring (Guardian REAL-Time CGMS) during two 9-day in-house periods. A hypoglycaemic threshold alarm alerted patients to sensor readings <70?mg/dl. Continuous glucose monitoring sensor readings were compared to laboratory reference measurements taken every 4?h and in case of a hypoglycaemic alarm. Results A total of 2317 paired data points were evaluated. Overall, the mean absolute relative difference (MARD) was 16.7%. The percentage of data points in the clinically accurate or acceptable Clarke Error Grid zones A?+?B was 94.6%. In the hypoglycaemic range, accuracy worsened (MARD 38.8%) leading to a failure to detect more than half of the true hypoglycaemic events (sensitivity 37.5%). Furthermore, more than half of the alarms that warn patients for hypoglycaemia were false (false alert rate 53.3%). Above the low alert threshold, the sensor confirmed 2077 of 2182 reference values (specificity 95.2%). Conclusions Patients using continuous glucose monitoring should be aware of its limitation to accurately detect hypoglycaemia.
引用
收藏
页码:130 / 135
页数:6
相关论文
共 50 条
  • [31] Impact on hypoglycaemia awareness of real time continuous glucose monitoring and intermittent continuous glucose data (I HART CGM)
    Reddy, M.
    Jugnee, N.
    Spanudakis, E.
    Oliver, N.
    DIABETIC MEDICINE, 2017, 34 : 29 - 29
  • [32] Sustained reductions in severe hypoglycaemia, but no recovery of hypoglycaemia awareness, with continuous glucose monitoring in adults with Type 1 diabetes
    Slattery, D.
    Iftikhar, M. I.
    Brackenridge, A.
    Hopkins, D. H.
    Amiel, S.
    Choudhary, P.
    DIABETIC MEDICINE, 2018, 35 : 25 - 25
  • [33] Detection of patient-specific intra-day patterns among hypoglycaemia episodes using continuous glucose monitoring
    Cheng, R.
    Dunn, T. C.
    DIABETOLOGIA, 2010, 53
  • [34] Continuous Glucose Monitoring in Resource-Constrained Settings for Hypoglycaemia Detection: Looking at the Problem from the Other Side of the Coin
    Bila, Rubao
    Varo, Rosauro
    Madrid, Lola
    Sitoe, Antonio
    Bassat, Quique
    BIOSENSORS-BASEL, 2018, 8 (02):
  • [35] HYPOGLYCAEMIA, ENDURANCE EXERCISE, PEER SUPPORT AND CONTINUOUS GLUCOSE MONITORING: TEAM BLOOD GLUCOSE CYCLING RESEARCH
    Hill, N. E.
    Buchanan, P.
    Knight, M.
    Oliver, N. S.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 : A159 - A160
  • [36] Hypoglycemia detection with continuous tissue glucose monitoring
    Jendrike, N
    Patte, C
    Buck, H
    Haug, C
    Freckmann, G
    DIABETES, 2005, 54 : A99 - A99
  • [37] Detection of Hypoglycemia with the continuous glucose monitoring system
    Sibenaler, A
    Boullu-Sanchis, S
    Jean-Didier, N
    Pingét, M
    Meyer, L
    DIABETES, 2003, 52 : A99 - A99
  • [38] Risk factors of severe hypoglycaemia and occurrence of silent hypoglycaemia in patients with type 1 diabetes assessed by continuous glucose monitoring
    Pedersen-Bjergaard, U
    Hoi-Hansen, T
    Pramming, S
    Thorsteinsson, B
    DIABETOLOGIA, 2005, 48 : A293 - A293
  • [39] Severe pancreatic allograft associated hypoglycaemia captured on an ambulatory continuous glucose monitoring system
    George, Elizabeth
    Harrison, Natalie
    CLINICAL ENDOCRINOLOGY, 2018, 89 : 87 - 87
  • [40] REVISITING THE RELATIONSHIPS BETWEEN MEASURES OF GLYCAEMIC CONTROL AND HYPOGLYCAEMIA IN CONTINUOUS GLUCOSE MONITORING DATASETS
    Gimenez, M.
    Tannen, A. J.
    Reddy, M.
    Moscardo, V.
    Conget, I.
    Oliver, N.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2018, 20 : A24 - A24