Low-Temperature Plasma Ionization Differential Ion Mobility Spectrometry

被引:20
|
作者
Kuklya, Andriy [1 ]
Engelhard, Carsten [2 ]
Uteschil, Florian [1 ]
Kerpen, Klaus [1 ]
Marks, Robert [1 ]
Telgheder, Ursula [1 ,3 ]
机构
[1] UDE, Dept Instrumental Analyt Chem, D-45141 Essen, Germany
[2] Univ Siegen, Dept Biol & Chem, D-57068 Siegen, Germany
[3] IWW Water Ctr, D-45476 Mulheim, Germany
关键词
DESORPTION/IONIZATION MASS-SPECTROMETRY; ATMOSPHERIC-PRESSURE; PROBE; OPTIMIZATION; FIELD;
D O I
10.1021/acs.analchem.5b02077
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A low-temperature plasma (LTP) was used as an ionization source for differential ion mobility spectrometry (DMS) for the first time. This ionization source enhances the potential of DMS as a miniaturized System for on-site rapid monitoring. The effects of experimental parameters (e.g., discharge/carrier gas composition and flow rate, applied voltage) on the analysis of model aromatic compounds were investigated and discussed. It was found that the nature of reactant ion positive (RIP) is dependent on the discharge/carrier gas composition. The best response to the analyte was achieved when pure nitrogen was used as the discharge/carrier gas. The ability to perform analysis with zero helium consumption is especially attractive in view of the potential application of LTP-DMS for online (and on-site) monitoring. Analytical performance was determined with six environmentally relevant model compounds (benzene, toluene, ethylbenzene, p-xylene, 1,2,4-trimethylbenzene, and naphthalene) using LTP and directly compared to APPI and APCI (Ni-63) ionization sources. When LTP was coupled to DMS, calculated LOD values were found to be in the range of 35-257 ng L-1 (concentration in the carrier gas). These values are competitive with those calculated for two DMS equipped with traditional ionization sources (APPI, Ni-63). The obtained results are promising enough to ensure the potential of LTP as ionization source for DMS.
引用
收藏
页码:8932 / 8940
页数:9
相关论文
共 50 条
  • [41] IONIZATION KINETICS IN LOW-TEMPERATURE PLASMA OF NOBLE-GASES
    KUDRJAVTSEV, A
    MUSIENKO, T
    SKREBOV, V
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1987, 20 (06) : 718 - 727
  • [42] Electrospray Ionization Ion Mobility Mass Spectrometry
    Kohoutek, Katie M.
    Harrington, Peter de B.
    CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 2023, 53 (03) : 483 - 497
  • [43] ELECTROSPRAY-IONIZATION ION MOBILITY SPECTROMETRY
    WITTMER, D
    LUCKENBILL, BK
    HILL, HH
    CHEN, YH
    ANALYTICAL CHEMISTRY, 1994, 66 (14) : 2348 - 2355
  • [44] LOW-TEMPERATURE MOBILITY OF ION-IMPLANTED HELIUM IN NICKEL
    POKER, DB
    WILLIAMS, JM
    JOURNAL OF METALS, 1982, 34 (08): : 54 - 54
  • [45] SECONDARY ION MASS-SPECTROMETRY OF LOW-TEMPERATURE SOLIDS
    MICHL, J
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1983, 53 (SEP): : 255 - 272
  • [46] 'Plug and Play' assembly of a low-temperature plasma ionization mass spectrometry imaging (LTP-MSI) system
    Maldonado-Torres, Mauricio
    Fabricio Lopez-Hernandez, Jose
    Jimenez-Sandoval, Pedro
    Winkler, Robert
    JOURNAL OF PROTEOMICS, 2014, 102 : 60 - 65
  • [47] Characterization of a Low-Temperature Plasma (LTP) Ambient Ionization Source Using Temporally Resolved Monochromatic Imaging Spectrometry
    Chan, George C. -Y.
    Engelhard, Carsten
    Wiley, Joshua S.
    Shoulds, Ayanna U.
    Cooks, R. Graham
    Hieftje, Gary M.
    Shelley, Jacob T.
    APPLIED SPECTROSCOPY, 2023, 77 (08) : 940 - 956
  • [48] LOW-TEMPERATURE IONIZATION WAVE
    KATIN, VV
    MARTYNENKO, YV
    YAVLINSKII, YN
    PISMA V ZHURNAL TEKHNICHESKOI FIZIKI, 1987, 13 (11): : 665 - 668
  • [49] Gas Chromatography Interfaced with Atmospheric Pressure Ionization-Quadrupole Time-of-Flight-Mass Spectrometry by Low-Temperature Plasma Ionization
    Norgaard, Asger W.
    Kofoed-Sorensen, Vivi
    Svensmark, Bo
    Wolkoff, Peder
    Clausen, Per Axel
    ANALYTICAL CHEMISTRY, 2013, 85 (01) : 28 - 32
  • [50] Research of Influence Factors on Performance of Microwave Induced Plasma Ionization Ion Mobility Spectrometry
    Dai Jian-Xiong
    Duan Yi-Xiang
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2016, 44 (11) : 1686 - 1691