The Impact of Flash Intensity on Retinal Vessel Oxygen Saturation Measurements Using Dual Wavelength Oximetry

被引:14
|
作者
Heitmar, Rebekka [1 ]
Cubbidge, Robert Peter [1 ]
机构
[1] Aston Univ, Sch Life & Hlth Sci, Birmingham B4 7ET, W Midlands, England
关键词
oximetry; dual wavelength; photography; ethnicity; pigmentation;
D O I
10.1167/iovs.12-10493
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To establish the optimal flash settings for retinal vessel oxygen saturation parameters using dual-wavelength imaging in a multiethnic group. METHODS. Twelve healthy young subjects (mean age 32 years [SD 7]; three Mediterranean, two South Asian, and seven Caucasian individuals) underwent retinal vessel oxygen saturation measurements using dual-wavelength oximetry, noncontact tonometry, and manual sphygmomanometry. In order to evaluate the impact of flash intensity, we obtained three images (fundus camera angle 30 degrees, ONH centered) per flash setting. Flash settings of the fundus camera were increased in steps of 2 (initial setting of 6 and the final of 22), which reflect logarithmic increasing intensities from 13.5 to 214 Watt seconds (Ws). RESULTS. Flash settings below 27 Ws were too low to obtain saturation measurements, whereas flash settings of more than 214 Ws resulted in overexposed images. Retinal arteriolar and venular oxygen saturation was comparable at flash settings of 27 to 76 Ws (arterioles' range: 85%-92%; venules' range: 45%-53%). Higher flash settings lead to increased saturation measurements in both retinal arterioles (up to 110%) and venules (up to 92%), with a more pronounced increase in venules. CONCLUSIONS. Flash intensity has a significant impact on retinal vessel oxygen saturation measurements using dual-wavelength retinal oximetry. High flash intensities lead to supranormal oxygen saturation measurements with a magnified effect in retinal venules compared with arteries. In addition to even retinal illumination, the correct flash setting is of paramount importance for clinical acquisition of images in retinal oximetry. We recommend flash settings between 27 to 76 Ws.
引用
收藏
页码:2807 / 2811
页数:5
相关论文
共 50 条
  • [41] Retinal Vessel Oxygen Saturation Measurement Protocols and Their Agreement
    Heitmar, Rebekka
    Cubbidge, Robert P.
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2020, 9 (06): : 1 - 9
  • [42] Dual-wavelength retinal images denoising algorithm for improving the accuracy of oxygen saturation calculation
    Xian, Yong-Li
    Dai, Yun
    Gao, Chun-Ming
    Du, Rui
    JOURNAL OF BIOMEDICAL OPTICS, 2017, 22 (01)
  • [43] Dual Wavelength Oximetry: Noninvasive Measurements of Oxygenation in Mouse Retina
    Begum, Momotaz
    Tso, Daniel Y.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [44] Retinal vessel oxygen saturation and vessel diameter in retinitis pigmentosa at various ages
    Yao Zong
    Leilei Lin
    Changxian Yi
    Xia Huang
    Yue Fu
    Yanmin Dong
    Xiaobing Qian
    Yujie Li
    Qianying Gao
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2016, 254 : 243 - 252
  • [45] Retinal vessel oxygen saturation and vessel diameter in retinitis pigmentosa at various ages
    Zong, Yao
    Lin, Leilei
    Yi, Changxian
    Huang, Xia
    Fu, Yue
    Dong, Yanmin
    Qian, Xiaobing
    Li, Yujie
    Gao, Qianying
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2016, 254 (02) : 243 - 252
  • [46] Correlation of the Retinal Parapapillary Perfusion and the Retinal Vessel Oxygen Saturation in Glaucoma Patients
    Hasan, Somar M.
    Hammer, Martin
    Meller, Daniel
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (05) : 1309 - 1315
  • [47] Longitudinal study of retinal vessel oxygen saturation in diabetic retinopathy
    Hardarson, Sveinn Hakon
    Karlsson, Robert Arnar
    Bek, Toke
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [48] The influence of cataract light scatters on retinal vessel oxygen saturation
    Waizel, M.
    Turksever, C.
    Todorova, M. G.
    ACTA OPHTHALMOLOGICA, 2018, 96 : 137 - 137
  • [49] Retinal morphology and vessel oxygen saturation in normal and diabetic patients
    Dubis, Adam M.
    Nicholson, Luke
    Crosby-Nwaobi, Roxanne
    Hykin, Phil
    Sivaprasad, Sobha
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [50] Influence of cataract light scatters on retinal vessel oxygen saturation
    della Volpe Waizel, Maria
    Tuerksever, Cengiz
    Todorova, Margarita G.
    ACTA OPHTHALMOLOGICA, 2020, 98 (01) : E56 - E62