Wear Prediction of Ceramic-on-Ceramic Artificial Hip Joints

被引:1
|
作者
Askari, E. [1 ]
Flores, P. [2 ]
Dabirrahmani, D. [1 ]
Appleyard, R. [1 ]
机构
[1] Macquarie Univ, Sydney, NSW 2109, Australia
[2] Univ Minho, Braga, Portugal
关键词
Wear simulation; Ceramic-on-ceramic hip joint; Multibody systems; SIMULATION; IMPLANTS; MODEL;
D O I
10.1007/978-3-319-09411-3_49
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Wear can influence the lifetime and performance of implants and has been found to be a key factor in primary failure of artificial hip joints. The present study aims to present a spatial multibody dynamic model to predict wear in ceramic- on-ceramic hip implants. The problem was formulated by developing a spatial multibody dynamic model of a hip prosthesis taking three-dimensional physiological loading and motion of the human body into account. Then, the Archard wear model was integrated into the dynamic calculation of the hip implant to predict wear. Additionally, geometries of the cup and head were updated throughout the simulation to generate a more realistic wear simulation. The results were validated against current literature. Finally it was illustrated that friction-induced vibration caused excessive wear of hip implant components.
引用
收藏
页码:463 / 470
页数:8
相关论文
共 50 条
  • [31] Wear at the taper-trunnion junction of contemporary ceramic-on-ceramic hips shown in a multistation hip simulator
    Bhalekar, Rohan M.
    Smith, Simon L.
    Joyce, Thomas J.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2019, 107 (04) : 1199 - 1209
  • [32] How much of what is wear of THRs with ceramic-on-ceramic articulation?
    Walter, A
    BIOCERAMICS, 2000, 192-1 : 541 - 544
  • [33] Ceramic-on-Ceramic Total Hip Arthroplasty Incidence of Instability and Noise
    Schroder, David
    Bornstein, Lindsey
    Bostrom, Mathias P. G.
    Nestor, Bryan J.
    Padgett, Douglas E.
    Westrich, Geoffrey H.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) : 437 - 442
  • [34] Arthrosynovial cyst of the hip after total hip arthroplasty with a ceramic-on-ceramic interface
    Legaye, J
    Lenfant, P
    Delos, M
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 2004, 90 (05): : 475 - 478
  • [35] Analysis of a retrieved squeaking ceramic-on-ceramic hip arthroplasty bearing
    Sidebottom, Mark
    Allan, D. Gordon
    Paliwal, Manish
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 2017, 231 (13) : 2373 - 2383
  • [36] Methods for Characterization of Edge Wear in Ceramic-on-Ceramic Acetabular Cups
    Kapadia, Darshil
    Racasan, Radu
    Pagani, Luca
    Al-Hajjar, Mazen
    Bills, Paul
    BEYOND THE IMPLANT: RETRIEVAL ANALYSIS METHODS FOR IMPLANT SURVEILLANCE, 2018, 1606 : 156 - 172
  • [37] Can Squeaking With Ceramic-On-Ceramic Hip Articulations In Total Hip Arthroplasty Be Avoided?
    Kiyama, Takahiko
    Kinsey, Tracy L.
    Mahoney, Ormonde M.
    JOURNAL OF ARTHROPLASTY, 2013, 28 (06): : 1015 - 1020
  • [38] Debris from failed ceramic-on-ceramic and ceramic-on-polyethylene hip prostheses
    Mochida, Y
    Boehler, M
    Salzer, M
    Bauer, TW
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2001, (389) : 113 - 125
  • [39] Numerical study of squeaking suppresses for ceramic-on-ceramic hip endoprosthesis
    Fan, N.
    Chen, G. X.
    TRIBOLOGY INTERNATIONAL, 2012, 48 : 172 - 181
  • [40] A rare complication of ceramic-on-ceramic bearings in total hip arthroplasty
    Rosneck, James
    Klika, Alison
    Barsoum, Wael
    JOURNAL OF ARTHROPLASTY, 2008, 23 (02): : 311 - 313