Impact of inline magnetic fields on dose distributions for VMAT in lung tumor

被引:3
|
作者
Kubota, Takahiro [1 ]
Araki, Fujio [2 ]
Ohno, Takeshi [2 ]
机构
[1] Kumamoto Univ, Grad Sch Hlth Sci, 4-24-1 Kuhonji, Kumamoto, Japan
[2] Kumamoto Univ, Dept Hlth Sci, Fac Life Sci, 4-24-1 Kuhonji, Kumamoto, Japan
关键词
MRI-guided radiation therapy; Lung tumor; Dose distribution; Monte Carlo method; MV RADIOTHERAPY ACCELERATOR; MONTE-CARLO-SIMULATION; PHOTON BEAMS; MRI SCANNER; TRANSVERSE; ELECTRON; DOSIMETRY; THERAPY; DEPOSITION; INCREASE;
D O I
10.1016/j.ejmp.2019.03.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to investigate the impact of inline magnetic field on dose distribution for volumetric modulated arc therapy (VMAT) in lung tumors located at the chest wall and mediastinum. Two VMAT plans for a thorax phantom with lung tumors of 1 cm and 2 cm in diameter located at the chest wall were created by a treatment planning system. Next, five clinical VMAT plans for a non-small cell lung cancer (NSCLC) at early stages I and II of 5 cm or less in diameter were also used. The planning target volume (PTV) sizes were in the range from 11.1 to 82.7 cm(3). The prescription dose was 60 Gy for D-95 in the PTV. The VMAT dose distributions without and with uniform inline magnetic field of 0.5 T and 1.0 T were calculated using the Monte Carlo method. The dose distributions were analyzed by dose volume histograms, dose differences, and dose indices. In all VMAT plans, the PTV dose was enhanced by inline magnetic field. The dose enhancement was larger with 1.0 T than with 0.5 T. In phantom plans, D-98 in the PTV with 0.5 T and 1.0 T increased by 2.9-6.6 Gy and 3.9-9.8 Gy, respectively, in comparison with that at 0 T. Similarly, in clinical plans, it increased by 2.2-6.0 Gy and 3.9-10.7 Gy, respectively. Thus, the VMAT with the inline magnetic field was proved useful for the dose enhancement in the lung tumor located at the chest wall and mediastinum.
引用
收藏
页码:100 / 106
页数:7
相关论文
共 50 条
  • [41] Impact of dose-volume distributions on lung function of breast cancer and lymphoma patients
    Sanchez-Nieto, B.
    Cordova, A.
    Goset, K.
    Delgado, I.
    Caviedes, I.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S158 - S158
  • [42] Investigation of Dose Calculation Accuracy in VMAT Planning for SBRT Lung Treatment
    Rao, M.
    Wu, J.
    Cao, D.
    Ye, J.
    Chen, F.
    Wong, T.
    Mehta, V.
    Shepard, D.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S774 - S774
  • [43] Biological consequences of dynamic dose interplay in VMAT SBRT lung treatments
    Sjolin, M.
    Elezaj, D.
    Ottosson, W.
    Edmund, J. M.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S830 - S830
  • [44] The impact of organ motion on dose distributions
    Rietzel, E
    Jackson, J
    Adams, J
    Harris, T
    Chen, G
    MEDICAL PHYSICS, 2003, 30 (06) : 1428 - 1428
  • [45] IMRT/VMAT dose distributions generated for HD® and Millennium® collimators TrueBeam® and Clinac® accelerators
    Slosarek, Krzysztof
    Braclik, Iwona
    Leszczynski, Wojciech
    Kopczynska, Joanna
    Osewski, Wojciech
    Wendykier, Jacek
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2019, 24 (01) : 20 - 27
  • [46] "Is There a Threshold Limit for Dosimetric Impact Due to Breathing Motion During Lung VMAT Delivery"? - Experimental Measurement of Dose Measured in Moving Lung Tumors
    Varadhan, R.
    Way, S.
    Bedoya, M.
    Dickson, N.
    Anderson, N.
    MEDICAL PHYSICS, 2019, 46 (06) : E530 - E530
  • [47] EFFECTS OF LONGITUDINAL MAGNETIC-FIELDS ON THE DOSE DISTRIBUTIONS OF HIGH-ENERGY ELECTRON-BEAMS
    WEINHOUS, MS
    NATH, R
    SCHULZ, RJ
    MEDICAL PHYSICS, 1981, 8 (04) : 582 - 582
  • [48] Integrability of magnetic fields created by current distributions
    Aguirre, J.
    Gine, J.
    Peralta-Salas, D.
    NONLINEARITY, 2008, 21 (01) : 51 - 69
  • [49] DOSE DISTRIBUTIONS OF PENDULUM FIELDS IN THE FIELD BORDER PLANE
    SCHRADER, R
    STRAHLENTHERAPIE UND ONKOLOGIE, 1986, 162 (12): : 764 - 767
  • [50] Impact of target reproducibility on tumor dose in stereotactic radiotherapy of targets in the lung and liver
    Wulf, J
    Hädinger, U
    Oppitz, U
    Thiele, W
    Flentje, M
    RADIOTHERAPY AND ONCOLOGY, 2003, 66 (02) : 141 - 150