Impact of inline magnetic fields on dose distributions for VMAT in lung tumor

被引:3
|
作者
Kubota, Takahiro [1 ]
Araki, Fujio [2 ]
Ohno, Takeshi [2 ]
机构
[1] Kumamoto Univ, Grad Sch Hlth Sci, 4-24-1 Kuhonji, Kumamoto, Japan
[2] Kumamoto Univ, Dept Hlth Sci, Fac Life Sci, 4-24-1 Kuhonji, Kumamoto, Japan
关键词
MRI-guided radiation therapy; Lung tumor; Dose distribution; Monte Carlo method; MV RADIOTHERAPY ACCELERATOR; MONTE-CARLO-SIMULATION; PHOTON BEAMS; MRI SCANNER; TRANSVERSE; ELECTRON; DOSIMETRY; THERAPY; DEPOSITION; INCREASE;
D O I
10.1016/j.ejmp.2019.03.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to investigate the impact of inline magnetic field on dose distribution for volumetric modulated arc therapy (VMAT) in lung tumors located at the chest wall and mediastinum. Two VMAT plans for a thorax phantom with lung tumors of 1 cm and 2 cm in diameter located at the chest wall were created by a treatment planning system. Next, five clinical VMAT plans for a non-small cell lung cancer (NSCLC) at early stages I and II of 5 cm or less in diameter were also used. The planning target volume (PTV) sizes were in the range from 11.1 to 82.7 cm(3). The prescription dose was 60 Gy for D-95 in the PTV. The VMAT dose distributions without and with uniform inline magnetic field of 0.5 T and 1.0 T were calculated using the Monte Carlo method. The dose distributions were analyzed by dose volume histograms, dose differences, and dose indices. In all VMAT plans, the PTV dose was enhanced by inline magnetic field. The dose enhancement was larger with 1.0 T than with 0.5 T. In phantom plans, D-98 in the PTV with 0.5 T and 1.0 T increased by 2.9-6.6 Gy and 3.9-9.8 Gy, respectively, in comparison with that at 0 T. Similarly, in clinical plans, it increased by 2.2-6.0 Gy and 3.9-10.7 Gy, respectively. Thus, the VMAT with the inline magnetic field was proved useful for the dose enhancement in the lung tumor located at the chest wall and mediastinum.
引用
收藏
页码:100 / 106
页数:7
相关论文
共 50 条
  • [31] Dose impact of rectal gas on prostatic IMRT and VMAT
    Sasaki, Motoharu
    Ikushima, Hitoshi
    Tominaga, Masahide
    Kamomae, Takeshi
    Kishi, Taro
    Oita, Masataka
    Harada, Masafumi
    JAPANESE JOURNAL OF RADIOLOGY, 2015, 33 (12) : 723 - 733
  • [32] A dosimetric study of dose algorithm impact and number of VMAT arcs in single fraction lung SBRT.
    Koh, W. Y. C.
    Tan, H. Q.
    Ng, Y. Y.
    Ang, K. W.
    Lew, K. S.
    Lin, Y. H.
    Chua, G. A. C.
    Yap, H. H. J.
    Phua, J. H.
    Wong, Y. M.
    Park, S. Y.
    Lee, J. C. L.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1542 - S1543
  • [33] Dose impact of rectal gas on prostatic IMRT and VMAT
    Motoharu Sasaki
    Hitoshi Ikushima
    Masahide Tominaga
    Takeshi Kamomae
    Taro Kishi
    Masataka Oita
    Masafumi Harada
    Japanese Journal of Radiology, 2015, 33 : 723 - 733
  • [34] A Study on the Dose Scalability of Volumetric Modulated Arc Therapy (VMAT) Fields
    Xiong, L.
    Iftimia, I. N.
    Lo, T. C.
    Kazi, M. H.
    Mower, H. W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S757 - S757
  • [35] Characterization of RTOG Dose Spillage Constraints in VMAT Lung SBRT
    Desai, D.
    Srinivasan, S.
    ElAsmar, H.
    MEDICAL PHYSICS, 2018, 45 (06) : E501 - E501
  • [36] Impacts of mechanical errors on dose distributions and control quality results for VMAT plans
    Legrand, C.
    Posnic, M.
    Autret, D.
    Fontbonne, J.
    Fontbonne, C.
    Dufreneix, S.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1419 - S1421
  • [37] Cadaveric Verification of VMAT Dose Distributions for Spine SBRT Treatments With Titanium Implants
    Grams, M.
    Clarke, M.
    Antolak, J. A.
    Brinkmann, D. H.
    de los Santos, L. Fong
    Mayo, C.
    Park, S. S.
    Olivier, K.
    Whitaker, T. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E547 - E548
  • [38] Comparison of Two VMAT Planning Techniques Using Planned and Delivered Dose Distributions
    Mathew, D.
    Edlund, S.
    Watanabe, Y.
    MEDICAL PHYSICS, 2017, 44 (06) : 2913 - 2913
  • [39] Hydrogel and scalp/skin conductivities impact dose from tumor treating fields
    Lok, Edwin
    Liang, Olivia
    Haack, Monika
    Wong, Eric T.
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2025, 13
  • [40] The Impact of Tumor Volume Decrease during Radiotherapy on Radiation Dose to the Lung
    Baluna, R. G.
    Ampil, F. L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S718 - S718