摘要
The paper describes a precision measurement system developed at NMI, Australia for the calibration of three-phase instruments measuring harmonics of current up to 20 A, voltage fluctuation and flicker. Particular attention is given to establishing a traceable link to the Australian primary standards of electrical quantities. Current and voltage harmonics up to the 99(th) and fluctuating voltages can be generated and measured with relative uncertainties close to 1x10(-5).
引用
收藏
页码:526 / 531
页数:6
相关论文
共 50 条
  • [1] Precision measurement of power harmonics and flicker
    Budovsky, I
    Hammond, G
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2005, 54 (02) : 483 - 487
  • [2] PRECISION CALIBRATION OF PHASE METERS
    TURGEL, RS
    VECCHIA, DF
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1987, 36 (04) : 918 - 922
  • [3] A Practical Approach for Calibration of Harmonics and Flicker Test Systems
    Phillips, Stephen
    Zombolas, Chris
    [J]. 2017 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY & SIGNAL/POWER INTEGRITY (EMCSI), 2017,
  • [4] A measurement system for traceable calibration of power harmonics and flicker analysers
    Hammond, G
    Budovsky, I
    [J]. 2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST, 2002, : 238 - 239
  • [5] Calibration and evaluation of neutron survey meters used at linac facility
    Salgado, A. P.
    Pereira, W. W.
    da Fonseca, E. S.
    Patrao, K. C. S.
    Batista, D. V. S.
    [J]. RADIATION MEASUREMENTS, 2010, 45 (10) : 1502 - 1504
  • [6] HIGH-PRECISION WOLTER OPTIC CALIBRATION FACILITY
    MORALES, RI
    REMINGTON, BA
    SCHWINN, T
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01): : 700 - 702
  • [7] Automation of Testing the Precision of Electro-Optical Distance Meters at the Calibration Baseline
    Barkovic, Duro
    Zrinjski, Mladen
    Baricevic, Sergej
    [J]. GEODETSKI LIST, 2016, 70 (04) : 311 - 336
  • [8] A unique test facility for calibration of domestic flow meters under dynamic flow conditions
    Buker, Oliver
    Stolt, Krister
    Lindstrom, Kent
    Wennergren, Per
    Penttinen, Olle
    Mattiasson, Kerstin
    [J]. FLOW MEASUREMENT AND INSTRUMENTATION, 2021, 79
  • [9] Development and evaluation of the calibration facility for high-pressure hydrogen gas flow meters
    Morioka, Toshihiro
    Ito, Masaru
    Fujikawa, Shizuichi
    Ishibashi, Masahiro
    Nakao, Shin-ichi
    [J]. FLOW MEASUREMENT AND INSTRUMENTATION, 2014, 39 : 19 - 24
  • [10] Development of filter radiometer based high accurate calibration facility for solar UV meters
    Durak, M
    [J]. OPTIK, 2004, 115 (12): : 533 - 537