Development of filter radiometer based high accurate calibration facility for solar UV meters

被引:3
|
作者
Durak, M [1 ]
机构
[1] UME, Natl Metrol Inst Turkey, TR-41470 Gebze, Kocaeli, Turkey
来源
OPTIK | 2004年 / 115卷 / 12期
关键词
filter radiometer; trap detector; aperture; solar UV meter;
D O I
10.1078/0030-4026-00533
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A reference filter radiometer based measurement system was established in the laboratory to calibrate solar UV radiometers. Filter radiometer was constructed from three-element silicon photodiode based reflection type trap detector, band-pass filters and precision aperture in a temperature-controlled enclosure. Spectral responsivity of trap detector was determined as traceable to the cryogenic radiometer and electrically calibrated pyroelectric detector. Relative uncertainties in the filter radiometer's responsivities at six wavelengths between 291 nm and 434 nm were calculated. In order to engage this radiometer in various types of implementations, filter(s) having previously known spectral transmittance was used in the radiometer housing. Calibration of solar UV meters can be achived accurately by using the developed system.
引用
收藏
页码:533 / 537
页数:5
相关论文
共 50 条
  • [1] Development and evaluation of the calibration facility for high-pressure hydrogen gas flow meters
    Morioka, Toshihiro
    Ito, Masaru
    Fujikawa, Shizuichi
    Ishibashi, Masahiro
    Nakao, Shin-ichi
    [J]. FLOW MEASUREMENT AND INSTRUMENTATION, 2014, 39 : 19 - 24
  • [2] On the calibration of a solar UV radiometer to measure broadband UV radiation from blacklight lamps
    Vijayaraghavan, S
    Goswami, DY
    [J]. JOURNAL OF SOLAR ENERGY ENGINEERING-TRANSACTIONS OF THE ASME, 2002, 124 (03): : 317 - 319
  • [3] Accurate calibration of filter radiometers against a cryogenic radiometer using a trap detector
    Friedrich, R
    Fischer, J
    Stock, M
    [J]. METROLOGIA, 1996, 32 (06) : 509 - 513
  • [4] CALIBRATION OF THE ABSOLUTE SPECTRAL SENSITIVITY OF A SOLAR UV RADIOMETER WITH THE USE OF SYNCHROTRON RADIATION
    ANEVSKY, SI
    VERNYI, AE
    GONYUKH, DA
    KAZACHEVSKAYA, TV
    KONEV, IV
    SAPRITSKY, VI
    KHROMCHENKO, VB
    TSIGELNITSKII, YN
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 308 (1-2): : 165 - 168
  • [5] Evaluation of a calibration constant of a narrow-band solar UV-B radiometer
    Takeshita, S
    Sasaki, M
    [J]. ULTRAVIOLET GROUND- AND SPACE-BASED MEASUREMENTS, MODELS, AND EFFECTS IV, 2004, 5545 : 49 - 55
  • [6] Solar radiation based calibration of an airborne radiometer for vicarious calibration of earth observing sensors
    Kuester, MA
    Thome, KJ
    Biggar, SF
    Krause, KS
    [J]. EARTH OBSERVING SYSTEMS VI, 2002, 4483 : 85 - 92
  • [7] Development of a heliostat facility for solar-radiation-based calibration of earth observing sensors
    Kuester, Michele A.
    Czapla-Myers, Jeffrey
    Kaptchen, Paul
    Good, William
    Lin, Tony
    To, Raymund
    Biggar, Stuart
    Thome, Kurtis
    [J]. EARTH OBSERVING SYSTEMS XIII, 2008, 7081
  • [8] Detector based calibration of solar UV radiometers in laboratory environments with high accuracy
    Durak, M
    San, SE
    Aslan, MH
    [J]. MEASUREMENT, 2005, 37 (03) : 213 - 217
  • [9] Highly accurate optical radiation calibration of InGaAs trap based on cryogenic radiometer
    Remote Sensing Department, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China
    [J]. Guangxue Jishu, 2006, SUPPL. (64-66):
  • [10] Solar contamination effects on the radiometric calibration of the advanced very high resolution radiometer
    Cao, CY
    Weinreb, M
    [J]. EARTH OBSERVING SYSTEMS V, 2000, 4135 : 364 - 373