Calibration and evaluation of neutron survey meters used at linac facility

被引:2
|
作者
Salgado, A. P. [1 ]
Pereira, W. W. [1 ]
da Fonseca, E. S. [1 ]
Patrao, K. C. S. [1 ]
Batista, D. V. S. [2 ]
机构
[1] IRD, BR-22780160 Rio De Janeiro, Brazil
[2] INCa, BR-20230130 Rio De Janeiro, Brazil
关键词
Clinical linear accelerator; Neutron metrology; Neutron survey-meter calibration;
D O I
10.1016/j.radmeas.2010.06.033
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Calibrated survey meters from the Neutron Laboratory of the Instituto de Radioprotecao e Dosimetria (IRD) were used to determine the ambient dose-equivalent rate in a 15 MV linear accelerator treatment room at the Instituto Nacional do Cancer (INCa). Three different models of neutron survey meters were calibrated using four neutron radionuclide neutron sources: (AmBe)-Am-241(alpha,n), Cf-252(f,n), heavy-water moderated Cf-252(f,n), and (PuBe)-Pu-238(alpha,n). All neutron sources were standardized in a Manganese Sulphate Bath (MSB) absolute primary system. The response of each of these instruments was compared with reference values of ambient dose-equivalent rate. The results demonstrate the complexity of making measurements in the mixed neutron/photon field produced in electron linear accelerator radiotherapy treatment rooms. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1502 / 1504
页数:3
相关论文
共 50 条
  • [1] FACILITY AT NATIONAL-NUCLEAR-RESEARCH-CENTRE, PELINDABA, FOR CALIBRATION OF GAMMA SURVEY METERS USED IN URANIUM PROSPECTING OPERATIONS
    TOENS, PD
    VANAS, D
    VLEGGAAR, CM
    [J]. JOURNAL OF THE SOUTH AFRICAN INSTITUTE OF MINING AND METALLURGY, 1973, 73 (12): : 428 - 429
  • [2] Analysis of calibration results of radiation survey meters used for area monitoring
    Adjei, D.
    Darko, E. O.
    Annkah, J. K.
    Amoako, J. K.
    Ofori, K.
    Emi-Reynolds, G.
    Obeng, M. K.
    Akomaning-Adofo, E.
    Owusu-Manteaw, P.
    [J]. RADIATION PROTECTION DOSIMETRY, 2013, 156 (04) : 506 - 513
  • [3] A REVIEW OF NEUTRON SCATTERING CORRECTION FOR THE CALIBRATION OF NEUTRON SURVEY METERS USING THE SHADOW CONE METHOD
    Kim, Sang In
    Kim, Bong Hwan
    Kim, Jang Lyul
    Lee, Jung Il
    [J]. NUCLEAR ENGINEERING AND TECHNOLOGY, 2015, 47 (07) : 939 - 944
  • [4] Evaluation of the calibration factors of neutron dose rate meters in a 241Am–Be neutron field
    Thiem Ngoc Le
    Sy Minh Tuan Hoang
    Quynh Ngoc Nguyen
    Hoai-Nam Tran
    [J]. Nuclear Science and Techniques, 2019, 30
  • [5] Evaluation of the calibration factors of neutron dose rate meters in a 241Am-Be neutron field
    Thiem Ngoc Le
    Sy Minh Tuan Hoang
    Quynh Ngoc Nguyen
    Hoai-Nam Tran
    [J]. NUCLEAR SCIENCE AND TECHNIQUES, 2019, 30 (09)
  • [6] Evaluation of the calibration factors of neutron dose rate meters in a 241Am–Be neutron field
    Thiem Ngoc Le
    Sy Minh Tuan Hoang
    Quynh Ngoc Nguyen
    Hoai-Nam Tran
    [J]. Nuclear Science and Techniques, 2019, 30 (09) : 11 - 16
  • [7] Development and evaluation of the calibration facility for high-pressure hydrogen gas flow meters
    Morioka, Toshihiro
    Ito, Masaru
    Fujikawa, Shizuichi
    Ishibashi, Masahiro
    Nakao, Shin-ichi
    [J]. FLOW MEASUREMENT AND INSTRUMENTATION, 2014, 39 : 19 - 24
  • [8] CALIBRATION AND EVALUATION OF HAZE METERS
    BUCKEE, GK
    MORRIS, TM
    BAILEY, TP
    [J]. JOURNAL OF THE INSTITUTE OF BREWING, 1986, 92 (05) : 475 - 482
  • [9] Australian Facility for the Calibration of Precision Harmonics and Flicker Meters
    Budovsky, A.
    Hammond, G. M.
    Georgakopoulos, D.
    [J]. 2008 13TH INTERNATIONAL CONFERENCE ON HARMONICS AND QUALITY OF POWER, VOLS 1 AND 2, 2008, : 526 - 531
  • [10] Evaluation of calibration factors of digital survey meters in a secondary standard dosimetry laboratory
    Azah, Collins K.
    Deatanyah, Philip
    Appiah, Prince M.
    Osei, Samuel
    Amoako, Joseph K.
    Sam, Frederick
    [J]. RADIATION PROTECTION DOSIMETRY, 2024, 200 (02) : 113 - 119