Performance of Iterative Image Reconstruction in CT of the Paranasal Sinuses: A Phantom Study

被引:53
|
作者
Schulz, B. [1 ]
Beeres, M. [1 ]
Bodelle, B. [1 ]
Bauer, R. [1 ]
Al-Butmeh, F. [1 ]
Thalhammer, A. [1 ]
Vogl, T. J. [1 ]
Kerl, J. M. [1 ]
机构
[1] Clin Goethe Univ, Dept Diagnost & Intervent Radiol, D-60590 Frankfurt, Germany
关键词
DOSE REDUCTION; QUALITY; ANGIOGRAPHY; ALGORITHM;
D O I
10.3174/ajnr.A3339
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: CT in low dose technique is the criterion standard imaging modality for evaluation of the paranasal sinus. Our aim was to evaluate the dose-reduction potential of a recently available sinogram-affirmed iterative reconstruction technique, regarding noise, image quality, and time duration when evaluating this region. MATERIALS AND METHODS: CT was performed on a phantom head at different tube voltages (120 kV, 100 kV) and currents (100 mAs, 50 mAs, 25 mAs). Each protocol was reconstructed (in soft tissue and bony kernel) by using standard filtered back-projection and 5 different SAFIRE strengths, and image noise was evaluated. Subjective image quality was evaluated on noise-aligned image triplets acquired at tube currents of 100% (FBP), 50% (SAFIRE), and 25% (SAFIRE) by using a 5-point scale (1 = worst, 5 = best). The time duration for image reconstruction was noted for calculations with FBP and SAFIRE. RESULTS: SAFIRE reduced image noise by 15%-85%, depending on the iterative strength, rendering kernel, and dose parameters. Noise reduction was stronger at a bone kernel algorithm both in 1- and 3-mm images (P < .05). Subjective quality evaluation of the noise-adapted images showed preference for those acquired at 100% tube current with FBP (4.7-5.0) versus 50% dose with SAFIRE (3.4-4.4) versus 25% dose with SAFIRE (2.0-3.1). The time duration for FBP image sets was 2.9-6.6 images per second versus SAFIRE with 0.9-1.6 images per second. CONCLUSIONS: For CT of the paranasal sinus, SAFIRE algorithms are suitable for image-noise reduction. Because image quality decreases with dosage, careful choice of the appropriate iterative method is necessary to achieve an optimal balance between image noise and quality.
引用
收藏
页码:1072 / 1076
页数:5
相关论文
共 50 条
  • [21] Deep Learning Reconstruction at CT: Phantom Study of the Image Characteristics
    Higaki, Toru
    Nakamura, Yuko
    Zhou, Jian
    Yu, Zhou
    Nemoto, Takuya
    Tatsugami, Fuminari
    Awai, Kazuo
    ACADEMIC RADIOLOGY, 2020, 27 (01) : 82 - 87
  • [22] SPIRAL CT SCANNING OF THE PARANASAL SINUSES
    SUOJANEN, JN
    REGAN, F
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1995, 16 (04) : 787 - 789
  • [23] Paranasal sinuses and nasopharynx CT and MRI
    Sievers, KW
    Greess, H
    Baum, U
    Dobritz, M
    Lenz, M
    EUROPEAN JOURNAL OF RADIOLOGY, 2000, 33 (03) : 185 - 202
  • [24] Role of MR and CT in the paranasal sinuses
    Branstetter, BF
    Weissman, JL
    OTOLARYNGOLOGIC CLINICS OF NORTH AMERICA, 2005, 38 (06) : 1279 - +
  • [25] CT EVALUATION OF DISEASES OF PARANASAL SINUSES
    Ranjan, Pushpa
    Ranjan, Rajeev
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2016, 5 (67): : 4843 - 4850
  • [26] Anatomical variations of paranasal air sinuses - A CT scan study
    Chaitanya, D. Krishna
    Suseelamma, D.
    Singh, Vishram
    JOURNAL OF THE ANATOMICAL SOCIETY OF INDIA, 2015, 64 (01) : 87 - 90
  • [27] Image Quality in Oncologic Chest Computerized Tomography With Iterative Reconstruction: A Phantom Study
    Jensen, Kristin
    Aalokken, Trond Mogens
    Tingberg, Anders
    Fosse, Erik
    Martinsen, Anne Catrine T.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2016, 40 (03) : 351 - 356
  • [28] CT scanning of the paranasal sinuses: The effect of reducing mAs on image quality and radiation
    Sohaib, SA
    Horrock, J
    Keene, M
    Kenyon, G
    Reznek, RH
    RADIOLOGY, 1999, 213P : 444 - 444
  • [29] Usefulness of CT iterative reconstruction method for low-dose CT attenuation correction on the accuracy of PET image in obesity phantom study
    Matsutomo, Norikazu
    Onishi, Hideo
    Nagaki, Akio
    Matsumoto, Keiichi
    Sasaki, Masayuki
    JOURNAL OF NUCLEAR MEDICINE, 2013, 54
  • [30] Optimization of chest CT protocols based on pixel image matrix, kernels and iterative reconstruction levels-A phantom study
    Johnsen, A. -M. Sandbukt
    Fenn, J. M.
    Henning, M. K.
    Hauge, I. H.
    RADIOGRAPHY, 2023, 29 (04) : 752 - 759