AN INVESTIGATION INTO CT RADIATION DOSE VARIATIONS FOR HEAD EXAMINATIONS ON MATCHED EQUIPMENT

被引:4
|
作者
Zarb, Francis [1 ]
Foley, Shane [2 ]
Holm, Susanne [3 ]
Toomey, Rachel [2 ]
Evanoff, Michael G. [4 ]
Rainford, Louise [2 ]
机构
[1] Univ Malta, Dept Radiog, Fac Hlth Sci, Msida, Malta
[2] Univ Coll Dublin, Sch Med, Diagnost Imaging, Dublin 4, Ireland
[3] Univ Coll Lillebaelt, CONRAD Radiog Res Ctr, Blangstedgaardsvej 4, DK-5220 Odense, Denmark
[4] Amer Board Radiol, 5441 East Williams Blvd,Suite 200, Tucson, AZ USA
关键词
DIAGNOSTIC REFERENCE LEVELS; IMAGE QUALITY; CHEST CT; COMPUTED-TOMOGRAPHY; IONIZING-RADIATION; VGC ANALYSIS; EXPOSURE; OPTIMIZATION; PERFORMANCE; MANAGEMENT;
D O I
10.1093/rpd/ncv549
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This study investigated radiation dose and image quality differences for computed tomography (CT) head examinations across centres with matched CT equipment. Radiation dose records and imaging protocols currently employed across three European university teaching hospitals were collated, compared and coded as Centres A, B and C from specification matched CT equipment models. Patient scans (n = 40) obtained from Centres A and C were evaluated for image quality, based on the visualisation of Commission of European Community (CEC) image quality criteria using visual grading characteristic (VGC) analysis, where American Board of Radiology examiners (n 5 11) stated their confidence in identifying anatomical criteria. Mean doses in terms of CT dose index (CTDIvol-mGy) and dose length product (DLP-mGy cm) were as follows: Centre A-33.12 mGy and 461.45 mGy cm; Centre B -101 mGy (base)/32 mGy (cerebrum) and 762 mGy cm and Centre C-71.98 mGy and 1047.26 mGy cm, showing a significant difference (p <= 0.05) in DLP across centres. VGC analysis indicated better visualisation of CEC criteria on Centre C images (VGC(AUC) 0.225). All three imaging protocols are routinely used clinically, and image quality is acceptable in each centre. Clinical centres with identical model CT scanners have variously customised their protocols achieving a range of dose savings and still resulting in clinically acceptable image quality.
引用
收藏
页码:466 / 474
页数:9
相关论文
共 50 条
  • [21] Evaluation of patient radiation dose and risk of cancer from CT examinations
    Saowapark Poosiri
    Anchali Krisanachinda
    Kitiwat Khamwan
    Radiological Physics and Technology, 2024, 17 : 176 - 185
  • [22] Reducing radiation dose associated with pediatric CT by decreasing unnecessary examinations
    Donnelly, LF
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 184 (02) : 655 - 657
  • [23] Pointers for Optimizing Radiation Dose in Head CT Protocols
    Singh, Sarabjeet
    Kalra, Mannudeep K.
    Thrall, James H.
    Mahesh, Mahadevappa
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2011, 8 (08) : 591 - 593
  • [24] RADIATION DOSE AND IMAGE QUALITY IN PEDIATRIC HEAD CT
    Spampinato, Maria Vittoria
    Stalcup, Seth
    Matheus, Maria Gisele
    Byington, Kathleen
    Tyler, Michael
    Bickley, Stetson
    Tipnis, Sameer
    RADIATION PROTECTION DOSIMETRY, 2018, 182 (03) : 310 - 316
  • [25] Development of age-specific Japanese head phantoms for dose evaluation in paediatric head CT examinations
    Yamauchi-Kawaura, C.
    Fujii, K.
    Akahane, K.
    Yamauchi, M.
    Narai, K.
    Aoyama, T.
    Katsu, T.
    Obara, S.
    Imai, K.
    Ikeda, M.
    RADIATION PROTECTION DOSIMETRY, 2015, 163 (02) : 188 - 201
  • [26] Estimation of Radiation Dose to Korean Population by Cardiovascular SPECT-CT Examinations
    Yun, Y. J.
    Nam, H. W.
    Kim, W. J.
    Kim, K. P.
    MEDICAL PHYSICS, 2020, 47 (06) : E611 - E611
  • [27] Radiation dose exposure in patients affected by lymphoma undergoing repeat CT examinations: how to manage the radiation dose variability
    Fabio Paolicchi
    Luca Bastiani
    Davide Guido
    Antonio Dore
    Giacomo Aringhieri
    Davide Caramella
    La radiologia medica, 2018, 123 : 191 - 201
  • [28] PAEDIATRIC CT EXAMINATIONS IN 19 DEVELOPING COUNTRIES: FREQUENCY. AND RADIATION DOSE
    Muhogora, W. E.
    Ahmed, N. A.
    AlSuwaidi, J. S.
    Beganovic, A.
    Ciraj-Bjelac, O.
    Gershan, V.
    Gershkevitsh, E.
    Grupetta, E.
    Kharita, M. H.
    Manatrakul, N.
    Maroufi, B.
    Milakovic, M.
    Ohno, K.
    Ben Omrane, L.
    Ptacek, J.
    Schandorf, C.
    Shaaban, M. S.
    Toutaoui, N.
    Sakkas, D.
    Wambani, J. S.
    Rehani, M. M.
    RADIATION PROTECTION DOSIMETRY, 2010, 140 (01) : 49 - 58
  • [29] Radiation dose exposure in patients affected by lymphoma undergoing repeat CT examinations: how to manage the radiation dose variability
    Paolicchi, Fabio
    Bastiani, Luca
    Guido, Davide
    Dore, Antonio
    Aringhieri, Giacomo
    Caramella, Davide
    RADIOLOGIA MEDICA, 2018, 123 (03): : 191 - 201
  • [30] CT Dose in Pediatric PET/CT Examinations
    Tonkopi, E.
    Ross, A.
    MacDonald, A.
    MEDICAL PHYSICS, 2011, 38 (06)