CONSTRUCTION AND TEST OF A PORTABLE TISSUE-EQUIVALENT PROPORTIONAL COUNTER SYSTEM

被引:3
|
作者
Zhang, Weihua [1 ]
Ni, Ning [1 ]
Wang, Zhiqiang [1 ]
Liu, Yina [1 ]
Li, Chunjuan [1 ]
Luo, Hailong [1 ]
机构
[1] China Inst Atom Energy, Radiat Metrol Dept, Beijing 102413, Peoples R China
关键词
DIFFERENT DOSIMETERS; COSMIC-RADIATION; ALTITUDES; H-ASTERISK(10);
D O I
10.1093/rpd/ncx214
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Tissue-equivalent proportional counter (TEPC) is a very promising dosemeter for radiation protection in mixed radiation fields. In this article, a portable TEPC system including a spherical homemade TEPC detector and the compact low-power integrated circuit prototype was constructed and tested in gamma and neutron radiation fields. The microdosimetric spectra results displayed that the portable TEPC system could deal with four orders of magnitude voltage pulse signals. And the measured dose equivalent values agreed with reference values, which means the portable TEPC system had fulfilled the test and satisfied the requirement for radiation monitor.
引用
收藏
页码:95 / 100
页数:6
相关论文
共 50 条
  • [41] PERFORMANCE OF A NEUTRON DOSIMETER BASED ON A TISSUE-EQUIVALENT PROPORTIONAL CHAMBER
    EISEN, Y
    BRAKE, R
    VASILIK, D
    ERKKILA, B
    LITTLEJOHN, G
    HEALTH PHYSICS, 1984, 47 (01): : 171 - 171
  • [42] MCNP modelling of the wall effects observed in tissue-equivalent proportional counters
    Hoff, JL
    Townsend, LW
    RADIATION PROTECTION DOSIMETRY, 2002, 99 (1-4) : 369 - 370
  • [43] Microdosimetry with tissue-equivalent proportional counters at an ion beam therapy facility
    Barna, S.
    Meouchi, C.
    Magrin, G.
    Conte, V.
    Stock, M.
    Resch, A.
    Georg, D.
    Palmans, H.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S735 - S736
  • [44] Characterization of miniature tissue-equivalent proportional counters for neutron radiotherapy applications
    Burmeister, J
    Kota, C
    Maughan, RL
    Waker, AJ
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (10): : 1633 - 1645
  • [45] Measurement of lineal-energy distributions for neutrons of 8 keV to 65 MeV by using a tissue-equivalent proportional counter
    Nunomiya, T
    Kim, E
    Kurosawa, T
    Taniguchi, S
    Nakamura, T
    Nakane, Y
    Sakamoto, Y
    Tanaka, S
    RADIATION PROTECTION DOSIMETRY, 2002, 102 (01) : 49 - 59
  • [46] Calibration and background measurements with a tissue equivalent proportional counter
    Autischer, M.
    Beck, P.
    Kindl, P.
    Latocha, M.
    Rollet, S.
    RADIATION PROTECTION DOSIMETRY, 2007, 125 (1-4) : 429 - 432
  • [47] A TISSUE-EQUIVALENT MODIFIED FBX DOSIMETRIC SYSTEM
    UPADHYAY, SN
    REDDY, AR
    GUPTA, MM
    NAGARATNAM, A
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (01): : 47 - 49
  • [48] THE DOSE EQUIVALENT RESPONSE OF TISSUE-EQUIVALENT PROPORTIONAL-COUNTERS TO LOW-ENERGY NEUTRONS
    SCHUHMACHER, H
    KUNZ, A
    MENZEL, HG
    COYNE, JJ
    SCHWARTZ, RB
    RADIATION PROTECTION DOSIMETRY, 1990, 31 (1-4) : 383 - 387
  • [49] Dosimetric response to γ rays and neutrons of a tissue-equivalent microstrip gas counter
    Dubeau, J
    Waker, AJ
    Biggar, M
    Rayner, MD
    Sommerville, EW
    Welch, SE
    RADIATION PROTECTION DOSIMETRY, 2000, 91 (04) : 391 - 401
  • [50] MICRODOSIMETRIC MEASUREMENT OF SECONDARY RADIATION IN THE PASSIVE SCATTERED PROTON THERAPY ROOM OF iTHEMBA LABS USING A TISSUE-EQUIVALENT PROPORTIONAL COUNTER
    Chiriotti, S.
    Parisi, A.
    Vanhavere, F.
    De Saint-Hubert, M.
    Vandevoorde, C.
    Slabbert, J.
    Beukes, P.
    de Kock, E.
    Symons, J.
    RADIATION PROTECTION DOSIMETRY, 2018, 182 (02) : 252 - 257