A comparative study on the initial stability of different implants placed above the bone level using resonance frequency analysis

被引:7
|
作者
Kang, In-Ho [1 ]
Kim, Chang-Whe [1 ]
Lim, Young-Jun [1 ]
Kim, Myung-Joo [1 ]
机构
[1] Seoul Natl Univ, Sch Dent, Dept Prosthodont, Grad Sch, Seoul 110768, South Korea
来源
JOURNAL OF ADVANCED PROSTHODONTICS | 2011年 / 3卷 / 04期
关键词
Effective implant length; Initial stability; Implant design; Resonance frequency analysis; TITANIUM IMPLANTS; IDENTIFICATION;
D O I
10.4047/jap.2011.3.4.190
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
PURPOSE. This study evaluated the initial stability of different implants placed above the bone level in different types of bone. MATERIALS AND METHODS. As described by Lekholm and Zarb, cortical layers of bovine bone specimens were trimmed to a thickness of 2 mm, 1 mm or totally removed to reproduce bone types II, III, and IV respectively. Three Implant system (Branemark System (R) Mk III TiUnite (TM), Straumann Standard Implant SLA (R), and Astra Tech Microthread (TM)-OsseoSpeed (TM)) were tested. Control group implants were placed in level with the bone, while test group implants were placed 1, 2, 3, and 4 mm above the bone level. Initial stability was evaluated by resonance frequency analysis. Data was statistically analyzed by one-way analysis of variance in confidence level of 95%. The effective implant length and the Implant Stability Quotient (ISQ) were compared using simple linear regression analysis. RESULTS. In the control group, there was a significant difference in the ISQ values of the 3 implants in bone types III and IV (P<.05). The ISQ values of each implant decreased with increased effective implant length in all types of bone. In type II bone, the decrease in ISQ value per 1-mm increase in effective implant length of the Branemark and Astra implants was less than that of the Straumann implant. In bone types III and IV, this value in the Astra implant was less than that in the other 2 implants. CONCLUSION. The initial stability was much affected by the implant design in bone types III, IV and the implant design such as the short pitch interval was beneficial to the initial stability of implants placed above the bone level. [J Adv Prosthodont 2011;3: 190-5]
引用
收藏
页码:190 / 195
页数:6
相关论文
共 50 条
  • [31] Bone Loss in Implants Placed at Subcrestal and Crestal Level: A Systematic Review and Meta-Analysis
    Palacios-Garzon, Natalia
    Velasco-Ortega, Eugenio
    Lopez-Lopez, Jose
    MATERIALS, 2019, 12 (01)
  • [32] Insertion and removal torque of dental implants placed using different drilling protocols: An experimental study on artificial bone substitutes
    Romeo, Davide
    Chochlidakis, Konstantinos
    Barmak, Abdul Basir
    Agliardi, Enrico
    Lo Russo, Lucio
    Ercoli, Carlo
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2023, 32 (07): : 633 - 638
  • [33] Comparison of Crestal Bone Loss and Implant Stability Among the Implants Placed With Conventional Procedure and Using Osteotome Technique: A Clinical Study
    Padmanabhan, Thallam Veer
    Gupta, Rajiv Kumar
    JOURNAL OF ORAL IMPLANTOLOGY, 2010, 36 (06) : 475 - 483
  • [34] Laser resonance frequency analysis of pedicle screw stability: A cadaveric model bone study
    Nakashima, Daisuke
    Mikami, Katsuhiro
    Kikuchi, Shunsuke
    Nishikino, Masaharu
    Kitamura, Toshiyuki
    Hasegawa, Noboru
    Matsumoto, Morio
    Nakamura, Masaya
    Nagura, Takeo
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2021, 39 (11) : 2474 - 2484
  • [35] Comparative study of wide-diameter implants placed after dental extraction and implants positioned in mature bone for molar replacement
    Penarrocha-Diago, Miguel
    Carrillo-Garcia, Celia
    Boronat-Lopez, Araceli
    Garcia-Mira, Berta
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2008, 23 (03) : 497 - 501
  • [36] Stability measurements of 1-stage implants in the maxilla by means of resonance frequency analysis: A pilot study
    Zix, J
    Kessler-Liechti, G
    Mericske-Stern, D
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2005, 20 (05) : 747 - 752
  • [37] Comparison of Initial Implant Stability of Implants Placed Using Bicortical Fixation, Indirect Sinus Elevation, and Unicortical Fixation
    Hsu, Andrea
    Seong, Wook-Jin
    Wolff, Ryan
    Zhang, Lei
    Hodges, James
    Olin, Paul S.
    Hinrichs, James E.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2016, 31 (02) : 459 - 468
  • [38] Resonance Frequency Analysis to Assess the Stability of Immediate Implants: A Retrospective Clinical Trial
    Marchio, Vincenzo
    Cinquini, Chiara
    Derchi, Giacomo
    Brevi, Bruno Carlo
    Miccoli, Mario
    Barone, Antonio
    INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2023, 43 (04) : E165 - E172
  • [39] Evaluation of primary stability in modified implants: Analysis by resonance frequency and insertion torque
    Sciasci, Plinio
    Casalle, Nicole
    Vaz, Luis Geraldo
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2018, 20 (03) : 274 - 279
  • [40] Stability changes of miniscrew implants over time A pilot resonance frequency analysis
    Ure, Derid S.
    Oliver, Donald R.
    Kim, Ki Beom
    Melo, Ana Claudia
    Buschang, Peter H.
    ANGLE ORTHODONTIST, 2011, 81 (06) : 994 - 1000