EPID-based dosimetry to verify IMRT planar dose distribution for the aS1200 EPID and FFF beams

被引:44
|
作者
Miri, Narges [1 ]
Keller, Peter [2 ]
Zwan, Benjamin J. [1 ,3 ]
Greer, Peter [1 ,4 ]
机构
[1] Univ Newcastle, Sch Math & Phys Sci, Math Bldg V123,Univ Dr, Callaghan, NSW 2308, Australia
[2] Varian Med Syst Imaging Lab, Baden, Switzerland
[3] Gosford Hosp, Cent Coast Canc Ctr, Gosford, NSW, Australia
[4] Calvary Mater Newcastle Hosp, Newcastle, NSW, Australia
来源
关键词
EPID; dosimetry; IMRT; IMRT treatment plan verification; FFF beams; PORTAL IMAGING DEVICE; PRETREATMENT VERIFICATION; RADIATION-THERAPY; PHOTON BEAMS; BACKSCATTER; ACCELERATOR; CALIBRATION; PREDICTION; ALGORITHM; FIELDS;
D O I
10.1120/jacmp.v17i6.6336
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We proposed to perform a basic dosimetry commissioning on a new imager system, the Varian aS1200 electronic portal imaging device (EPID) and TrueBeam 2.0 linear accelerator for flattened (FF) and flattening filter-free (FFF) beams, then to develop an image-based quality assurance (QA) model for verification of the system delivery accuracy for intensity-modulated radiation therapy (IMRT) treatments. For dosimetry testing, linearity of dose response with MU, imager lag, and effectiveness of backscatter shielding were investigated. Then, an image-based model was developed to convert images to planar dose onto a virtual water phantom. The model parameters were identified using energy fluence of the Acuros treatment planning system (TPS) and, reference dose profiles and output factors measured at depths of 5, 10, 15, and 20 cm in water phantom for square fields. To validate the model, its calculated dose was compared to measured dose from MapCHECK 2 diode arrays for 36 IMRT fields at 10 cm depth delivered with 6X, 6XFFF, 10X, and 10XFFF energies. An in-house gamma function was used to compare planar doses pixel-by-pixel. Finally, the method was applied to the same IMRT fields to verify their pretreatment delivery dose compared with Eclipse TPS dose. For the EPID commissioning, dose linearity was within 0.4% above 5 MU and similar to 1% above 2 MU, measured lag was smaller than the previous EPIDs, and profile symmetry was improved. The model was validated with mean gamma pass rates (standard deviation) of 99.0% (0.4%), 99.5% (0.6%), 99.3% (0.4%), and 98.0% (0.8%) at 3%/3 mm for respectively 6X, 6XFFF, 10X, and 10XFFF beams. Using the same comparison criteria, the beam deliveries were verified with mean pass rates of 100% (0.0%), 99.6% (0.3%), 99.9% (0.1%), and 98.7% (1.4%). Improvements were observed in dosimetric response of the aS1200 imager compared to previous EPID models, and the model was successfully developed for the new system and delivery energies of 6 and 10 MV, FF, and FFF modes.
引用
收藏
页码:292 / 304
页数:13
相关论文
共 50 条
  • [1] EPID-BASED DOSIMETRY
    Wendling, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S17 - S17
  • [2] Development of EPID-based dosimetry for FFF-beam verification in radiation therapy
    Thongsawad, S.
    Chanton, T.
    Saiyo, N.
    Udee, N.
    [J]. INTERNATIONAL NUCLEAR SCIENCE AND TECHNOLOGY CONFERENCE, 2019, 1285
  • [3] EPID-BASED DOSE RECONSTRUCTION MODEL FOR WEDGED BEAMS
    Camargo, P.
    Wendling, M.
    Mans, A.
    Sonke, J. J.
    van Herk, M.
    Mijnheer, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S230 - S230
  • [4] 3D EPID-based in vivo dosimetry for IMRT and VMAT
    Mijnheer, B.
    Olaciregui-Ruiz, I.
    Rozendaal, R.
    Sonke, J-J
    Spreeuw, H.
    Tielenburg, R.
    van Herk, M.
    Vijlbrief, R.
    Mans, A.
    [J]. 7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [5] IMRT Dose Reconstruction Using An A-Si EPID-Based System
    Fitzpatrick, M.
    Vazquez-Quino, L.
    Alkhatib, H.
    Chen, X.
    Shi, C.
    Papanikolaou, N.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [6] A simple approach to using an amorphous silicon EPID to verify IMRT planar dose maps
    Lee, Christopher
    Menk, Fred
    Cadman, Patrick
    Greer, Peter B.
    [J]. MEDICAL PHYSICS, 2009, 36 (03) : 984 - 992
  • [7] Free Information - EPID-Based In Vivo Dosimetry
    Greer, P.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3745 - 3745
  • [8] Planar EPID-Based Dosimetry for SRS and SRT Patient-Specific QA
    Thongsawad, Sangutid
    Chanton, Tadchapong
    Saiyo, Nipon
    Udee, Nuntawat
    [J]. LIFE-BASEL, 2021, 11 (11):
  • [9] A Novel EPID-Based 3D Dose Reconstruction for IMRT QA
    Alhazmi, A.
    Gianoli, C.
    Neppl, S.
    Veloza, S.
    Martins, J.
    Podesta, M.
    Verhaegen, F.
    Reiner, M.
    Belka, C.
    Parodi, K.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2887 - 2887
  • [10] Establishing action levels for EPID-based QA for IMRT
    Howell, Rebecca M.
    Smith, Iris P. N.
    Jarrio, Christie S.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (03): : 16 - 25