EPID-based dosimetry to verify IMRT planar dose distribution for the aS1200 EPID and FFF beams

被引:44
|
作者
Miri, Narges [1 ]
Keller, Peter [2 ]
Zwan, Benjamin J. [1 ,3 ]
Greer, Peter [1 ,4 ]
机构
[1] Univ Newcastle, Sch Math & Phys Sci, Math Bldg V123,Univ Dr, Callaghan, NSW 2308, Australia
[2] Varian Med Syst Imaging Lab, Baden, Switzerland
[3] Gosford Hosp, Cent Coast Canc Ctr, Gosford, NSW, Australia
[4] Calvary Mater Newcastle Hosp, Newcastle, NSW, Australia
来源
关键词
EPID; dosimetry; IMRT; IMRT treatment plan verification; FFF beams; PORTAL IMAGING DEVICE; PRETREATMENT VERIFICATION; RADIATION-THERAPY; PHOTON BEAMS; BACKSCATTER; ACCELERATOR; CALIBRATION; PREDICTION; ALGORITHM; FIELDS;
D O I
10.1120/jacmp.v17i6.6336
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We proposed to perform a basic dosimetry commissioning on a new imager system, the Varian aS1200 electronic portal imaging device (EPID) and TrueBeam 2.0 linear accelerator for flattened (FF) and flattening filter-free (FFF) beams, then to develop an image-based quality assurance (QA) model for verification of the system delivery accuracy for intensity-modulated radiation therapy (IMRT) treatments. For dosimetry testing, linearity of dose response with MU, imager lag, and effectiveness of backscatter shielding were investigated. Then, an image-based model was developed to convert images to planar dose onto a virtual water phantom. The model parameters were identified using energy fluence of the Acuros treatment planning system (TPS) and, reference dose profiles and output factors measured at depths of 5, 10, 15, and 20 cm in water phantom for square fields. To validate the model, its calculated dose was compared to measured dose from MapCHECK 2 diode arrays for 36 IMRT fields at 10 cm depth delivered with 6X, 6XFFF, 10X, and 10XFFF energies. An in-house gamma function was used to compare planar doses pixel-by-pixel. Finally, the method was applied to the same IMRT fields to verify their pretreatment delivery dose compared with Eclipse TPS dose. For the EPID commissioning, dose linearity was within 0.4% above 5 MU and similar to 1% above 2 MU, measured lag was smaller than the previous EPIDs, and profile symmetry was improved. The model was validated with mean gamma pass rates (standard deviation) of 99.0% (0.4%), 99.5% (0.6%), 99.3% (0.4%), and 98.0% (0.8%) at 3%/3 mm for respectively 6X, 6XFFF, 10X, and 10XFFF beams. Using the same comparison criteria, the beam deliveries were verified with mean pass rates of 100% (0.0%), 99.6% (0.3%), 99.9% (0.1%), and 98.7% (1.4%). Improvements were observed in dosimetric response of the aS1200 imager compared to previous EPID models, and the model was successfully developed for the new system and delivery energies of 6 and 10 MV, FF, and FFF modes.
引用
收藏
页码:292 / 304
页数:13
相关论文
共 50 条
  • [31] Determining Threshold-Of-Failure for Dead Pixel Rows in EPID-Based Dosimetry
    Gersh, J.
    Wiant, D.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 331 - 331
  • [32] Direct dose to water dosimetry for pretreatment IMRT verification using a modified EPID
    Gustafsson, Helen
    Vial, Philip
    Kuncic, Zdenka
    Baldock, Clive
    Denham, James W.
    Greer, Peter B.
    [J]. MEDICAL PHYSICS, 2011, 38 (11) : 6257 - 6264
  • [33] Characterization of a new EPID-based system for in-vivo dosimetry in VMAT treatments
    Bresciani, S.
    Poli, M.
    Miranti, A.
    Maggio, A.
    Di Dia, A.
    Bracco, C.
    Stasi, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S736 - S737
  • [34] GLAaS absolute dose calibration for iViewGT EPID with flat and FFF beams: multicenter experience
    de Palma, E. Vanetti
    Perez, J. M.
    Kaiser, S. Ren
    Nicolini, G.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S941 - S942
  • [35] EPID-Based in Vivo Dosimetry: The Next Step in Patient-Specific QA
    Mijnheer, B.
    Olaciregui-Ruiz, I.
    Rozendaal, R.
    Sonke, J-J
    Spreeuw, H.
    Tielenburg, R.
    van Herk, M.
    Mans, A.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [36] Step-and-Shoot IMRT by Siemens Beams: An EPID Dosimetry Verification During Treatment
    Russo, M.
    Piermattei, A.
    Greco, F.
    Azario, L.
    Orlandini, L.
    Zucca, S.
    Cilla, S.
    Menna, S.
    Grusio, M.
    Chiatti, L.
    Fidanzio, A.
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2016, 15 (04) : 535 - 545
  • [37] Absolute dosimetry of 6 and 18 MV IMRT beams using a dedicated fluoroscopic EPID
    Monti, A. F.
    Gelosa, S.
    Frigerio, M.
    Gentile, C. C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S183 - S183
  • [38] EPID-based 3D dosimetry for pre-treatment FFF VMAT stereotactic body radiotherapy plan verification using dosimetry Check™
    Chendi, Agnese
    Botti, Andrea
    Orlandi, Matteo
    Sghedoni, Roberto
    Iori, Mauro
    Cagni, Elisabetta
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 81 : 227 - 236
  • [39] EPID-based 3D in vivo dosimetry in head and neck radiotherapy
    Biltekin, F.
    Yazici, G.
    Ozyigit, G.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1354 - S1355
  • [40] A convolutional neural network model for EPID-based non-transit dosimetry
    Bosco, Lucas Dal
    Franceries, Xavier
    Romain, Blandine
    Smekens, Francois
    Husson, Francois
    Le Lann, Marie-Veronique
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (06):