Portable Light Pen 3D Vision Coordinate Measuring System-Probe Tip Center Calibration

被引:14
|
作者
Liu, Shugui [1 ]
Zhang, Hongling [1 ]
Dong, Yinghua [1 ]
Tang, Shaliang [1 ]
Jiang, Zhenzhu [1 ]
机构
[1] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
来源
MEASUREMENT SCIENCE REVIEW | 2013年 / 13卷 / 04期
关键词
Probe tip; center calibration; light pen; portable 3D vision coordinate measuring system; reference conical hole; LASER; METROLOGY; POSITION;
D O I
10.2478/msr-2013-0030
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
For different tasks, probe tip should be changed in the 3D vision coordinate measuring system and the accurate determination of probe tip center position is critical. A novel and simple approach for calibrating the probe tip center position of the light pen is presented in this paper. Hundreds of images of the light pen with different postures are collected while the probe tip is kept in firm contact with a reference conical hole. The probe tip position is determined by computing the rotation matrix and translation vector from the obtained images by using the least square fitting method. The experimental results demonstrate the effectiveness of the proposed approach. Its repeatability reaches 0.033 mm, 0.030 mm, and 0.043 mm in x, y, and z axes, respectively, and its convergence speed is satisfactory.
引用
收藏
页码:194 / 199
页数:6
相关论文
共 50 条
  • [41] Robot performance measurement and calibration using a 3D computer vision system
    Preising, B.
    Hsia, T.C.
    Robotica, 1995, 13 (pt 4) : 327 - 337
  • [42] 3D Coordinate Identification of Object Using Binocular Vision System for Mobile Robot
    Huang, Guo-Shing
    Cheng, Ci-En
    2013 CACS INTERNATIONAL AUTOMATIC CONTROL CONFERENCE (CACS), 2013, : 91 - 96
  • [43] A long-stroke 3D contact scanning probe for micro/nano coordinate measuring machine
    Li, Rui-Jun
    Fan, Kuang-Chao
    Huang, Qiang-Xian
    Zhou, Hao
    Gong, Er-Min
    Xiang, Meng
    PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2016, 43 : 220 - 229
  • [44] Flexible Projector Calibration in the Structured Light 3D Measurement System
    Wu, Haitao
    Li, Biao
    Zhang, Jiancheng
    Yang, Jie
    Fu, Yanjun
    ADVANCES IN IMAGE AND GRAPHICS TECHNOLOGIES (IGTA 2015), 2015, 525 : 158 - 166
  • [45] Simulation of 3D measurement system based on structured light calibration
    Yu, Xiaoyang
    Ji, Ran
    Wu, Mingyang
    Wu, Halbin
    Wei, Kaifeng
    DYNAMICS OF CONTINUOUS DISCRETE AND IMPULSIVE SYSTEMS-SERIES B-APPLICATIONS & ALGORITHMS, 2006, 13E : 1481 - 1484
  • [46] The study on encoding and calibration of 3D measuring device using projective structured light
    Fan, JY
    Wei, XN
    Wang, Y
    Yu, XY
    Wang, FG
    ISTM/2005: 6th International Symposium on Test and Measurement, Vols 1-9, Conference Proceedings, 2005, : 3254 - 3258
  • [47] A robot cell calibration algorithm and its use with a 3D measuring system
    Rocadas, PS
    McMaster, RS
    ISIE '97 - PROCEEDINGS OF THE IEEE INTERNATIONAL SYMPOSIUM ON INDUSTRIAL ELECTRONICS, VOLS 1-3, 1997, : SS297 - SS302
  • [48] A laser tracking 3D coordinate measuring system and design of its tracking mechanism
    Zhao, SZ
    Zhang, GX
    Li, XH
    PROCEEDINGS OF THE THIRD INTERNATIONAL SYMPOSIUM ON INSTRUMENTATION SCIENCE AND TECHNOLOGY, VOL 2, 2004, : 642 - 647
  • [49] Error Analysis and Compensation of 3D Coordinate Measuring System by Watching an Image of Object
    Peng Kai
    Liu Li-Bing
    Zhou Xing-Lin
    PROCEEDINGS OF THE 2009 2ND INTERNATIONAL CONGRESS ON IMAGE AND SIGNAL PROCESSING, VOLS 1-9, 2009, : 2368 - 2372
  • [50] The Design of the 3D Reconstruction System of Line Structured Light Vision
    He, Tao
    Zhu, Caisheng
    Xie, Yulang
    Lai, Kexue
    Yang, Liangen
    PROCEEDINGS OF THE 2015 2ND INTERNATIONAL CONFERENCE ON ELECTRICAL, COMPUTER ENGINEERING AND ELECTRONICS (ICECEE 2015), 2015, 24 : 1368 - 1373