Intracochlear Pressure Transients During Cochlear Implant Electrode Insertion

被引:47
|
作者
Greene, Nathaniel T. [1 ,2 ]
Mattingly, Jameson K. [1 ]
Hartl, Renee M. Banakis [1 ]
Tollin, Daniel J. [1 ,2 ]
Cass, Stephen P. [1 ]
机构
[1] Univ Colorado, Sch Med, Dept Otolaryngol, Aurora, CO USA
[2] Univ Colorado, Sch Med, Dept Physiol & Biophys, Aurora, CO 80045 USA
关键词
Cochlear implant; Electroacoustic stimulation; Hearing preservation; Intracochlear pressures; MIDDLE-EAR IMPLANT; HEARING PRESERVATION; CONDUCTION; NOISE; STIMULATION; EXPOSURE; SURGEON; DESIGN; SYSTEM; AIR;
D O I
10.1097/MAO.0000000000001232
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Hypothesis: Cochlear implant (CI) electrode insertion into the round window induces pressure transients in the cochlear fluid comparable to high-intensity sound transients. Background: Many patients receiving a CI have some remaining functional hearing at low frequencies; thus, devices and surgical techniques have been developed to use this residual hearing. To maintain functional acoustic hearing, it is important to retain function of any hair cells and auditory nerve fibers innervating the basilar membrane; however, in a subset of patients, residual low-frequency hearing is lost after CI insertion. Here, we test the hypothesis that transient intracochlear pressure spikes are generated during CI electrode insertion, which could cause damage and compromise residual hearing. Methods: Human cadaveric temporal bones were prepared with an extended facial recess. Pressures in the scala vestibuli and tympani were measured with fiber-optic pressure sensors inserted into the cochlea near the oval and round windows, whereas CI electrodes (five styles from two manufacturers) were inserted into the cochlea via a round window approach. Results: Pressures in the scala tympani tended to be larger in magnitude than pressures in the scala vestibuli, consistent with electrode insertion into the scala tympani. CI electrode insertion produced a range of pressure transients in the cochlea that could occur alone or as part of a train of spikes with equivalent peak sound pressure levels in excess of 170 dB sound pressure level. Instances of pressure transients varied with electrode styles. Conclusion: Results suggest electrode design, insertion mechanism, and surgical technique affect the magnitude and rate of intracochlear pressure transients during CI electrode insertion. Pressure transients showed intensities similar to those elicited by high-level sounds and thus could cause damage to the basilar membrane and/or hair cells.
引用
收藏
页码:1541 / 1548
页数:8
相关论文
共 50 条
  • [1] Measurement and Mitigation of Intracochlear Pressure Transients During Cochlear Implant Electrode Insertion
    Hartl, Renee M. Banakis
    Greene, Nathaniel T.
    OTOLOGY & NEUROTOLOGY, 2022, 43 (02) : 174 - 182
  • [2] Dynamic intracochlear pressure measurement during cochlear implant electrode insertion
    Ordonez, F.
    Riemann, C.
    Mueller, S.
    Sudhoff, H.
    Todt, I
    ACTA OTO-LARYNGOLOGICA, 2019, 139 (10) : 860 - 865
  • [3] Intracochlear Pressure Transients During Cochlear Implant Electrode Insertion: Effect of Micro-mechanical Control on Limiting Pressure Trauma
    Hartl, Renee M. Banakis
    Kaufmann, Christopher
    Hansen, Marlan R.
    Tollin, Daniel J.
    OTOLOGY & NEUROTOLOGY, 2019, 40 (06) : 736 - 744
  • [4] Effects of Different Insertion Techniques of a Cochlear Implant Electrode on the Intracochlear Pressure
    Todt, Ingo
    Ernst, Arneborg
    Mittmann, Philipp
    AUDIOLOGY AND NEURO-OTOLOGY, 2016, 21 (01) : 30 - 37
  • [5] Detection of intracochlear damage during cochlear implant electrode insertion using extracochlear measurements in the gerbil
    Ahmad, Faisal I.
    Choudhury, Baishakhi
    De Mason, Christine E.
    Adunka, Oliver F.
    Finley, Charles C.
    Fitzpatrick, Douglas C.
    LARYNGOSCOPE, 2012, 122 (03): : 636 - 644
  • [6] Cochlear implant electrode sealing techniques and related intracochlear pressure changes
    Todt, Ingo
    Utca, Julica
    Karimi, Dania
    Ernst, Arne
    Mittmann, Philipp
    JOURNAL OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2017, 46
  • [7] Patterns Seen During Electrode Insertion Using Intracochlear Electrocochleography Obtained Directly Through a Cochlear Implant
    Harris, Michael S.
    Riggs, William J.
    Giardina, Christopher K.
    O'Connell, Brendan P.
    Holder, Jourdan T.
    Dwyer, Robert T.
    Koka, Kanthaiah
    Labadie, Robert F.
    Fitzpatrick, Douglas C.
    Adunka, Oliver F.
    OTOLOGY & NEUROTOLOGY, 2017, 38 (10) : 1415 - 1420
  • [8] Effects of Round Window Opening Size and Moisturized Electrodes on Intracochlear Pressure Related to the Insertion of a Cochlear Implant Electrode
    Todt, Ingo
    Ernst, Arneborg
    Mittmann, Philipp
    AUDIOLOGY AND NEUROTOLOGY EXTRA, 2016, 6 (01): : 1 - 8
  • [9] Cochlear implant electrode sealing techniques and related intracochlear pressure changes
    Ingo Todt
    Julica Utca
    Dania Karimi
    Arne Ernst
    Philipp Mittmann
    Journal of Otolaryngology - Head & Neck Surgery, 46
  • [10] Direct Feedback during Insertion of the Cochlear Implant Electrode
    Kessing, Richard
    LARYNGO-RHINO-OTOLOGIE, 2024, 103 (09) : 629 - 630