Evaluation of intrastromal corneal ring segments for treatment of keratoconus with a mechanical implantation technique

被引:9
|
作者
Tunc, Zeki [1 ]
Helvacioglu, Firat [1 ]
Sencan, Sadik [1 ]
机构
[1] Maltepe Univ, Sch Med, Dept Ophthalmol, Istanbul, Turkey
关键词
Intrastromal corneal ring segments; keraring; keratoconus; FEMTOSECOND LASER; TREATING KERATOCONUS; INTACS INSERTS; FOLLOW-UP; MANAGEMENT; ECTASIA; EYE;
D O I
10.4103/0301-4738.109519
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the clinical outcomes of intrastromal corneal ring segment (ICRS) implantation in patients with keratoconus using a mechanical implantation technique. Materials and Methods: Thirty eyes of 17 patients with keratoconus were enrolled. ICRSs (Keraring) were implanted after dissection of the tunnel using Tunc's specially designed dissector under suction. A complete ophthalmic examination was performed, including uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), spherical equivalent, keratometric readings, inferosuperior asymmetry index (ISAI), and ultrasound pachymetry. All 3-, 6-, and 12-month follow-ups were completed, and statistical analysis was performed. Results: The mean preoperative UDVA for all eyes was 1.36 +/- 0.64 logMAR. At 12 months, the mean UDVA was 0.51 +/- 0.28 logMAR (P = 0.001), and the mean preoperative CDVA was 0.57 +/- 0.29 logMAR, which improved to 0.23 +/- 0.18 (P = 0.001) at 1 year. There was a significant reduction in spherical equivalent refractive error from -6.42 +/- 4.69 diopters (D) preoperatively to -1.26 +/- 1.45 D (P = 0.001) at 1 year. In the same period, the mean K-readings improved from 49.38 +/- 3.72 D to 44.43 +/- 3.13 D (P = 0.001), and the mean ISAI improved from 7.92 +/- 3.12 to 4.21 +/- 1.96 (P = 0.003). No significant changes in mean central corneal thickness were observed postoperatively. There were no major complications during and or after surgery. Conclusion: ICRS implantation using a unique mechanical dissection technique is a safe and effective treatment for keratoconus. All parameters improved by the 1-year follow-up.
引用
收藏
页码:218 / 225
页数:8
相关论文
共 50 条
  • [21] Intrastromal corneal ring segments and posterior chamber phakic intraocular lens implantation for keratoconus correction
    Alfonso, Jose F.
    Lisa, Carlos
    Fernandez-Vega, Luis
    Madrid-Costa, David
    Poo-Lopez, Arancha
    Montes-Mico, Robert
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (04): : 706 - 713
  • [22] Clinical Outcomes After Implantation of 320°-Arc Length Intrastromal Corneal Ring Segments in Keratoconus
    Torquetti, Leonardo
    Cunha, Priscila
    Luz, Allan
    Kwitko, Sergio
    Carrion, Marcio
    Rocha, Guilherme
    Signorelli, Armando
    Coscarelli, Sandro
    Ferrara, Guilherme
    Bicalho, Frederico
    Neves, Renato
    Ferrara, Paulo
    CORNEA, 2018, 37 (10) : 1299 - 1305
  • [23] A Novel Mobile Application for Preoperative and Intraoperative Assessments of Intrastromal Corneal Ring Segments Implantation for Keratoconus
    Fernandes Dias, Francisco Aecio
    Fernandes Dias, Vinicius Jose
    Carneiro, Carolina Lyra Barreira
    Bernardes, Bianca Gomes
    Dutra, Barbara de Araujo Lima
    Crispim Ribeiro, Joao
    CORNEA, 2020, 39 (09) : 1132 - 1138
  • [24] Asymmetric intrastromal corneal ring for the treatment of keratoconus
    Vilar de Siqueira, Marcelo Alves
    de Polo, Jhonny
    Gomes, Manuela Benites
    Moreira, Hamilton
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2010, 73 (05) : 454 - 458
  • [25] Evaluation of a Tangential Map-Based Nomogram for Intrastromal Corneal Ring Segments' Implantation in Keratoconus: One Year Results
    Zaky, Adel Galal
    KhalafAllah, Mahmoud Tawfik
    Sarhan, Abdelrahman Elsebaey
    Elsawy, Moataz Faiz
    JOURNAL OF OPHTHALMOLOGY, 2020, 2020
  • [26] Keratitis after implantation of intrastromal corneal ring segments
    Shehadeh-Masha'our, R
    Modi, N
    Barbara, A
    Garzozi, HJ
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (08): : 1802 - 1804
  • [27] Intrastromal corneal ring segments for keratoconus: results and correlation with preoperative corneal biomechanics
    Ambrosio, Renato, Jr.
    Borges, Jose Salgado
    Costa-Ferreira, Claudia
    Coelho, Vinicius
    da Silva, Renata Siqueira
    Valbon, Bruno
    Siqueira, Jorge Augusto
    Velarde, Guillermo Coca
    REVISTA BRASILEIRA DE OFTALMOLOGIA, 2012, 71 (02) : 89 - 99
  • [28] Biomechanical and morphological corneal response to placement of intrastromal corneal ring segments for keratoconus
    Dauwe, Caroline
    Touboul, David
    Roberts, Cynthia J.
    Mahmoud, Ashraf M.
    Kerautret, Julien
    Fournier, Pierre
    Malecaze, Francois
    Colin, Joseph
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (10): : 1761 - 1767
  • [29] Efficacy of progressive thickness intrastromal corneal ring segments in the treatment of duck phenotype keratoconus
    Arbelaez, Juan G.
    Arbelaez, Maria C.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2021, 31 (05) : 2191 - 2199
  • [30] Keratoconus correction using intrastromal corneal ring segments and posterior chamber phakic intraocular lens implantation
    Ramirez-Miranda, Arturo
    Gomez-Bastar, Arturo
    Navas, Alejandro
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (07): : 1373 - 1374