摘要
The purpose of this study was to evaluate the achievable organ dose savings in low-dose multislice computed tomography (MSCT) of the heart using different tube voltages (80 kVp, 100 kVp, 120 kVp) and compare it with calculated values. A female Alderson-Rando phantom was equipped with thermoluminescent dosimeters (TLDs) in five different positions to assess the mean doses within representative organs (thyroid gland, thymus, oesophagus, pancreas, liver). Radiation exposure was performed on a 16-row MSCT scanner with six different routine scan protocols: a 120-kV and a 100-kV CT angiography (CTA) protocol with the same collimation, two 120-kV Ca-scoring (CS) protocols with different collimations and two 80-kV CS protocols with the same collimation as the 120-kV CS protocols. Each scan protocol was repeated five times. The measured dose values for the organs were compared with the values calculated by a commercially available computer program. Directly irradiated organs, such as the esophagus, received doses of 34.7 mSv (CTA 16x0.75 120 kVp), 21.9 mSv (CTA 16x0.75 100 kVp) and 4.96 mSv (CS score 12x1.5 80 kVp), the thyroid as an organ receiving only scattered radiation collected organ doses of 2.98 mSv (CTA 16x0.75 120 kVp), 1.97 mSv (CTA 16x0.75 100 kVp) and 0.58 mSv (CS score 12x1.5 80 kVp). The measured relative organ dose reductions from standard to low-kV protocols ranged from 30.9% to 55.9% and were statistically significant (P < 0.05). The comparison with the calculated organ doses showed that the calculation program can predict the relative dose reduction of cardiac low photon-energy protocols precisely.
引用
收藏
页码:1841 / 1846
页数:6
相关论文
共 50 条
  • [31] Low-dose Radiation Exposure and Carcinogenesis
    Suzuki, Keiji
    Yamashita, Shunichi
    JAPANESE JOURNAL OF CLINICAL ONCOLOGY, 2012, 42 (07) : 563 - 568
  • [32] Dose Optimization for Computed Tomography Localizer Radiographs for Low-Dose Lung Computed Tomography Examinations
    Schmidt, Bernhard T.
    Hupfer, Martin
    Saltybaeva, Natalia
    Kolditz, Daniel
    Kalender, Willi A.
    INVESTIGATIVE RADIOLOGY, 2017, 52 (02) : 81 - 86
  • [33] Impact of Low-Dose Computed Tomography on Computed Tomography Orders and Scan Length
    Simmons, Curtis
    Milburn, James
    OCHSNER JOURNAL, 2019, 19 (04): : 303 - 308
  • [34] Daily computed tomography image guidance: Dosimetric advantages outweigh low-dose radiation exposure for treatment of mediastinal lymphoma
    Zhu, Lin L.
    Martin, Rachael M.
    Gunther, Jillian R.
    Wong, Pei-Fong
    Hancock, Donald
    Moreno, Amy C.
    Dabaja, Bouthaina S.
    Pinnix, Chelsea C.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : 14 - 18
  • [35] Dose quantities for characterization of patients' exposure to radiation in computed tomography and their significance for risk estimation
    N. Hidajat
    R.-J. Schröder
    M. Wolf
    T. Vogl
    R. Felix
    Der Radiologe, 1997, 37 : 464 - 469
  • [36] Dose quantities for characterization of patients' exposure to radiation in computed tomography and their significance for risk estimation
    Hidajat, N
    Schroder, RJ
    Wolf, M
    Vogl, T
    Felix, R
    RADIOLOGE, 1997, 37 (06): : 464 - 469
  • [37] Evaluation of a Low-Dose Computed Tomography Lung Cancer Screening Program in Henan, China
    Guo, Lan-Wei
    Chen, Qiong
    Shen, Yin-Chen
    Meng, Qing-Cheng
    Zheng, Li-Yang
    Wu, Yue
    Cao, Xiao-Qin
    Xu, Hui-Fang
    Liu, Shu-Zheng
    Sun, Xi-Bin
    Qiao, You-Lin
    Zhang, Shao-Kai
    JAMA NETWORK OPEN, 2020, 3 (11) : E2019039
  • [38] Radiation dose of the new dual source CT in multislice computed tomography of the coronaries
    Conradi, G.
    Rixe, J.
    Rolf, A.
    Deetjen, A.
    Hamm, C. H. R.
    Dill, T.
    EUROPEAN HEART JOURNAL, 2007, 28 : 22 - 22
  • [39] Dose reduction strategies for electrocardiography gated multislice spiral computed tomography of the heart
    Flohr, T
    Ohnesorge, B
    Becker, C
    Süss, C
    Langer, C
    Kopp, A
    Knez, A
    Reiser, MF
    AMERICAN JOURNAL OF CARDIOLOGY, 2001, 88 (2A): : 76E - 77E
  • [40] Benefit-to-radiation-risk of low-dose computed tomography lung cancer screening
    Hendrick, R. Edward
    Smith, Robert A.
    CANCER, 2024, 130 (02) : 216 - 223