Dose Optimization for Computed Tomography Localizer Radiographs for Low-Dose Lung Computed Tomography Examinations

被引:14
|
作者
Schmidt, Bernhard T. [1 ,2 ]
Hupfer, Martin [2 ,3 ]
Saltybaeva, Natalia [1 ]
Kolditz, Daniel [1 ,3 ]
Kalender, Willi A. [1 ]
机构
[1] Univ Erlangen Nurnberg, Inst Med Phys, Henkestr 91, D-91052 Erlangen, Germany
[2] Siemens Healthcare GmbH, Forchheim, Germany
[3] CT Imaging GmbH, Erlangen, Germany
关键词
dose; localizer radiograph; effective dose; optimization; CT; FEASIBILITY; ANGIOGRAPHY; REDUCTION;
D O I
10.1097/RLI.0000000000000311
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Recent studies have shown a substantial reduction of radiation dose from computed tomography (CT) scans down to 0.1 mSv for lung cancer screening and cardiac examinations, when applying optimization techniques. Hence, CT localizer radiographs (LRs) might now be considered a significant contributor to the total dose of the CT examination. We investigated in our study the potential for reducing dose of the LRs by adapting the patient-specific acquisition parameters of the LR. Materials and Methods: Localizer radiographs covering the lungs were acquired on 2 clinical scanners (64 slices, conventional detector [CD]; 96 slices, fully integrated detector [ID]) for 3 semianthropomorphic phantoms, representing a slim, a normal, and an obese adult. Starting at 120-kV tube voltage and 250-mA current were reduced until the image quality of the LR, and thereby the accuracy of the automatic exposure control was compromised; this was defined as a deviation of measured attenuation values in the center of the LR of more than 5% from the reference values measured at the highest tube voltage and current. Subsequent Monte Carlo calculations on anthropomorphic phantoms were performed to calculate organ and effective dose values for the respective optimal settings. In addition, effective dose values normalized to CTDIvol for tube voltages ranging from 60 to 160 kV were determined for the different combinations of phantom sizes, sexes, and LR views to evaluate dose efficiency. Results: For the CD scanner, the optimal LR settings depended strongly on phantom size. Higher tube voltage and current were necessary for the larger phantoms. The ID scanner showed uncompromised LR quality for all phantoms using the lowest possible tube voltage-tube current combination of 80 kV and 20 mA. Depending on patient size and LR direction, effective dose values for the optimal settings ranged from 6 to 53 mu Sv and 3 to 11 mu Sv for the CD and ID scanner, respectively. For the example of an anterior-posterior LR on a normal patient, using the optimal settings instead of the standard settings on the ID scanner reduced LR dose from 53 mu Sv (120 kV, 30 mA) to 10 mu Sv (80 kV, 20mA). The simulations for the different tube voltages show that effective dose and CTDIvol behave similarly for different views and patient sizes. However, the tube voltage level itself impacts the relationship between CTDIvol and effective dose, by up to a factor of 2. Discussion: Dose from LRs may contribute significantly to the total effective dose of low-dose CT examinations such as lung cancer screening. Optimal LR settings can reduce LR dose substantially, but adaptations have to consider scanner characteristics, detector technology, and patient size. Thus, for low-dose CT examinations, such as cardiac examinations and lung cancer screening, LR optimization may result in a significant dose reduction and thereby in a substantial reduction of total dose.
引用
收藏
页码:81 / 86
页数:6
相关论文
共 50 条
  • [1] Screening for lung cancer with low-dose computed tomography
    Diederich, S
    Wormanns, D
    Heindel, W
    [J]. RADIOLOGE, 2001, 41 (03): : 256 - 260
  • [2] Lung Cancer Screening with Low-Dose Computed Tomography
    Chiles, Caroline
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2014, 52 (01) : 27 - +
  • [3] Lung Cancer Screening With Low-Dose Computed Tomography
    Loebig, Stephanie
    Haegele, Patricia
    Seizer, Peter
    [J]. DEUTSCHES ARZTEBLATT INTERNATIONAL, 2023, 120 (41):
  • [4] Lung cancer screening with low-dose computed tomography
    Hartman, TE
    Swensen, SJ
    [J]. SEMINARS IN ROENTGENOLOGY, 2003, 38 (01) : 34 - 38
  • [5] Lung cancer screening with low-dose computed tomography
    Seijo Maceiras, Luis M.
    [J]. MEMO-MAGAZINE OF EUROPEAN MEDICAL ONCOLOGY, 2014, 7 (01) : 6 - 9
  • [6] Screening for Lung Cancer With Low-Dose Computed Tomography
    Miller, Anthony B.
    [J]. ONCOLOGIST, 2013, 18 (08): : 897 - 899
  • [7] Lung cancer screening with low-dose computed tomography
    Luis M. Seijo Maceiras
    [J]. memo - Magazine of European Medical Oncology, 2014, 7 (1) : 6 - 9
  • [8] Lung Cancer Screening With Low-Dose Computed Tomography
    Reck, Martin
    Dettmer, Sabine
    Kauczor, Hans-Ulrich
    Kaaks, Rudolf
    Reinmuth, Niels
    Vogel-Claussen, Jens
    [J]. DEUTSCHES ARZTEBLATT INTERNATIONAL, 2023, 120 (23): : 387 - +
  • [9] Low-Dose Computed Tomography Screening for Lung Cancer
    Welch, Laura S.
    Madtes, David
    Ringen, Knut
    [J]. ANNALS OF INTERNAL MEDICINE, 2015, 162 (06) : 459 - 460
  • [10] Screening for Lung Cancer With Low-Dose Computed Tomography
    Brodersen, John
    Dirksen, Asger
    Pedersen, Jesper Holst
    [J]. ANNALS OF INTERNAL MEDICINE, 2014, 160 (03) : 211 - 212