Effect of Different Implant Designs on Strain and Stress Distribution under Non-Axial Loading: A Three-Dimensional Finite Element Analysis

被引:27
|
作者
Oliveira, Helder [1 ]
Velasco, Aritza Brizuela [2 ]
Rios-Santos, Jose-Vicente [3 ]
Lasheras, Fernando Sanchez [4 ]
Lemos, Bernardo Ferreira [1 ]
Gil, Francisco Javier [5 ]
Carvalho, Alexandrine [1 ]
Herrero-Climent, Mariano [6 ]
机构
[1] Univ Fernando Pessoa, Fac Hlth Sci, P-4200150 Porto, Portugal
[2] Univ Oviedo, Dept Surg & Med Surg Specialties, Oviedo 33006, Spain
[3] Univ Seville, Sch Dent, Dept Periodontol, Seville 41009, Spain
[4] Univ Oviedo, Fac Sci, Math Dept, Oviedo 33007, Spain
[5] Univ Int Catalunya, Bioengn Inst Technol, Barcelona 08017, Spain
[6] Porto Dent Inst, P-4150518 Oporto, Portugal
关键词
dental implants; design; finite element analysis; strain distribution; stress distribution; bone quality; CONTROLLED CLINICAL-TRIAL; DENTAL IMPLANTS; OSSEOINTEGRATED IMPLANTS; EDENTULOUS PATIENTS; PRIMARY STABILITY; BONE-RESORPTION; EXTRACTION; IMMEDIATE; MICROMOTION; INTERFACE;
D O I
10.3390/ijerph17134738
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Implant design evolved alongside the development of implant therapy. The purpose of this finite element analysis (FEA) study was to analyze the influence of different implant designs on the stress and strain distribution to the implants and surrounding bone. Three implant designs with the same length and diameter were used. The three-dimensional geometry of the bone was simulated with a cortical bone of three different thicknesses and two medullar bone densities: low density (150 Hounsfield units) and high density (850 Hounsfield units). A 30 degrees oblique load of 150 N was applied to the implant restoration. Displacement and stress (von Mises) results were obtained for bone and dental implants. The strain and stress distributions to the bone were higher for the tissue-level implant for all types of bone. The maximum principal strain and stress decreased with an increase in cortical bone thickness for both cancellous bone densities. The distribution of the load was concentrated at the coronal portion of the bone and implants. All implants showed a good distribution of forces for non-axial loads, with higher forces concentrated at the crestal region of the bone-implant interface. Decrease in medullar bone density negatively affects the strain and stress produced by the implants.
引用
收藏
页码:1 / 17
页数:17
相关论文
共 50 条
  • [31] Influence of prosthetic index structures and implant materials on stress distribution in implant restorations: a three-dimensional finite element analysis
    Anniwaer, Annikaer
    Muhetaer, Aihemaiti
    Yin, Zhengrong
    Zhu, Jiakang
    Jin, Chunxiao
    Huang, Cui
    BMC ORAL HEALTH, 2024, 24 (01):
  • [32] Implant-Bone Interface Stress Distribution in Immediately Loaded Implants of Different Diameters: A Three-Dimensional Finite Element Analysis
    Ding, Xi
    Zhu, Xing-Hao
    Liao, Sheng-Hui
    Zhang, Xiu-Hua
    Chen, Hong
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2009, 18 (05): : 393 - 402
  • [33] Comparison of the effects of different implant apical designs on the magnitude and distribution of stress and strain in bone: A finite element analysis study
    Kadkhodazadeh, Mahdi
    Lafzi, Ardeshir
    Raoofi, Saeed
    Khademi, Maryam
    Amid, Reza
    Movahhedy, Mohammad Reza
    Torabi, Hasti
    Journal of Long-Term Effects of Medical Implants, 2014, 24 (2-3) : 109 - 120
  • [34] Three-dimensional finite element analysis of extra short implants focusing on implant designs and materials
    Araki, Haruka
    Nakano, Tamaki
    Ono, Shinji
    Yatani, Hirofumi
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2020, 6 (01)
  • [35] Three-dimensional finite element analysis of extra short implants focusing on implant designs and materials
    Haruka Araki
    Tamaki Nakano
    Shinji Ono
    Hirofumi Yatani
    International Journal of Implant Dentistry, 6
  • [36] Finite Element Analysis of Compressive Stress and Strain of Different Implant Forms During Vertical Loading
    Tanasic, I.
    Tihacek-Sojic, L.
    Milic-Lemic, A.
    INTERNATIONAL JOURNAL OF COMPUTERIZED DENTISTRY, 2014, 17 (02) : 125 - 133
  • [37] The effect of implant number and position on the stress behavior of mandibular implant retained overdentures: A three-dimensional finite element analysis
    Topkaya, Tolga
    Solmaz, Murat Yavuz
    JOURNAL OF BIOMECHANICS, 2015, 48 (10) : 2102 - 2109
  • [38] Effect of margin designs and loading conditions on the stress distribution of endocrowns: a finite element analysis
    Zeng B.
    Luo B.
    Hu J.
    Meng D.
    Zhang J.
    Cao X.
    Jiang Q.
    BMC Oral Health, 24 (1)
  • [39] Three-dimensional finite element analysis of stress distribution in peri-implant bone with relined dentures and different heights of healing caps
    dos Santos, M. B. F.
    da Silva Neto, J. P.
    Consani, R. L. X.
    Mesquita, M. F.
    JOURNAL OF ORAL REHABILITATION, 2011, 38 (09) : 691 - 696
  • [40] Three-dimensional finite element stress analysis: The technique and methodology of non-linear property simulation and soft tissue loading behavior for different partial denture designs
    Kanbara, Ryo
    Nakamura, Yoshinori
    Ochiai, Kent T.
    Kawai, Tatsushi
    Tanaka, Yoshinobu
    DENTAL MATERIALS JOURNAL, 2012, 31 (02) : 297 - 308