Comparison of breast specific gamma imaging and molecular breast tomosynthesis in breast cancer detection: Evaluation in phantoms

被引:15
|
作者
Gong, Zongyi [1 ]
Williams, Mark B. [1 ,2 ,3 ]
机构
[1] Univ Virginia, Dept Radiol & Med Imaging, Charlottesville, VA 22908 USA
[2] Univ Virginia, Dept Biomed Engn, Charlottesville, VA 22908 USA
[3] Univ Virginia, Dept Phys, Charlottesville, VA 22904 USA
关键词
breast specific gamma imaging; molecular imaging; nuclear medicine; breast; tomosynthesis; PROOF-OF-CONCEPT; HIGH-RISK; DENSITY; MODALITY; SCINTIMAMMOGRAPHY; MAMMOGRAPHY; SYSTEMS; CAMERA; WOMEN; DIAGNOSIS;
D O I
10.1118/1.4922398
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Breast specific gamma imaging or molecular breast imaging (BSGI) obtains 2D images of Tc-99m sestamibi distribution in the breast. Molecular breast tomosynthesis (MBT) maps the tracer distribution in 3D by acquiring multiple projections over a limited angular range. Here, the authors compare the performance of the two technologies in terms of spatial resolution, lesion contrast, and contrast-to-noise ratio (CNR) in phantom studies under conditions of clinically relevant sestamibi dose and imaging time. Methods: The systems tested were a Dilon 6800 and a MBT prototype developed at the University of Virginia. Both systems comprise a pixelated sodium iodide scintillator, an array of position sensitive photomultipliers, and a parallel hole collimator. The active areas and energy resolution of the systems are similar. System sensitivity, spatial resolution, lesion contrast, and CNR were measured using a Petri dish, a point source phantom, and a breast phantom containing simulated lesions at two depths, respectively. A single BSGI projection was acquired. Five MBT projections were acquired over +/- 20 degrees. For both modalities, the total scan count density was comparable to that observed for each in typical 10 min human scans following injection of 22 mCi (814 MBq) of Tc-99m-sestamibi. To assess the impact of reducing the tracer dose, the pixel counts of projection images were later binomially subsampled by a factor of 2 to give images corresponding to an injected activity of approximately 11 mCi (407 MBq). Both unprocessed (pixelated) BSGI projections and interpolated (smoothed) BSGI images displayed by default on the Dilon 6800 workstation were analyzed. Volumetric images were reconstructed from the MBT projections using a maximum likelihood expectation maximization algorithm and extracted slices were analyzed. Results: Over a depth range of 1.5-7.5 cm, BSGI spatial resolution was 5.6-11.5 mm in unprocessed projections and 5.7-12.0 mm in interpolated images. Over the same range, the in-slice MBT spatial resolution was 6.7-9.4 mm. Lesion contrast was significantly improved with MBT relative to BSGI for five out of eight lesions imaged at either the 22 mCi or the 11 mCi dose level (p < 0.05). At both dose levels, significant improvements in CNR with MBT were also found for five out of eight lesions (9.8, 7.8, 6.2 mm lesions at water depth of 1.7 cm and 9.8, 7.8 mm lesions at water depth of 4.5 cm, p < 0.05). The 6.2 and 4.9 mm lesions located at 4.5 cm below the water surface were not visible in either modality at either activity level. Conclusions: Under conditions of equal dose, imaging time and similar detectors, compared to BSGI, MBT provided higher lesion contrast, higher CNR, and spatial resolution that was less depth dependent. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:4250 / 4259
页数:10
相关论文
共 50 条
  • [41] Comparison Study of Reconstruction Algorithms for Prototype Digital Breast Tomosynthesis Using Various Type of Breast Phantoms
    Kim, Y.
    Kim, H.
    MEDICAL PHYSICS, 2015, 42 (06) : 3252 - 3252
  • [42] Diagnosis of breast cancer using breast-specific gamma imaging in patients with microcalcifications on mammography: comparison with ultrasonography
    Chung, Hyun Woo
    Lee, Eun Jeong
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [43] Effect of Dose Level on Radiologists' Detection of Microcalcifications in Digital Breast Tomosynthesis: An Observer Study with Breast Phantoms
    Chan, Heang-Ping
    Helvie, Mark A.
    Klein, Katherine A.
    McLaughlin, Carol
    Neal, Colleen H.
    Oudsema, Rebecca
    Rahman, W. Tania
    Roubidoux, Marilyn A.
    Hadjiiski, Lubomir M.
    Zhou, Chuan
    Samala, Ravi K.
    ACADEMIC RADIOLOGY, 2022, 29 : S42 - S49
  • [44] Breast Cancer: Computer-aided Detection with Digital Breast Tomosynthesis
    Morra, Lia
    Sacchetto, Daniela
    Durando, Manuela
    Agliozzo, Silvano
    Carbonaro, Luca Alessandro
    Delsanto, Silvia
    Pesce, Barbara
    Persano, Diego
    Mariscotti, Giovanna
    Marra, Vincenzo
    Fonio, Paolo
    Bert, Alberto
    RADIOLOGY, 2015, 277 (01) : 56 - 63
  • [45] Silicone breast phantoms for elastographic imaging evaluation
    Kashif, Amer S.
    Lotz, Thomas F.
    McGarry, Matthew D.
    Pattison, Adam J.
    Chase, James G.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [46] Evaluation of a novel collimator for molecular breast tomosynthesis
    Gilland, David R.
    Welch, Benjamin L.
    Lee, Seungjoon
    Kross, Brian
    Weisenberger, Andrew G.
    MEDICAL PHYSICS, 2017, 44 (11) : 5740 - 5748
  • [47] Digital breast tomosynthesis may improve breast cancer detection rates
    Barton, Mary Kay
    CA-A CANCER JOURNAL FOR CLINICIANS, 2013, 63 (05) : 291 - 292
  • [48] Experimental evaluation of seven quality control phantoms for digital breast tomosynthesis
    Sage, Julie
    Fezzani, Karen L.
    Fitton, Isabelle
    Hadid, Lama
    Moussier, Aurelie
    Pierrat, Noelle
    Martineau, Antoine
    Dreuil, Serge
    Heulers, Loic
    Etard, Cecile
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 57 : 137 - 144
  • [49] Alternative Screening for Women With Dense Breasts: Breast-Specific Gamma Imaging (Molecular Breast Imaging)
    Holbrook, Anna
    Newell, Mary S.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (02) : 252 - 256
  • [50] Breast-Specific Gamma Imaging: Correlations With Mammographic and Clinicopathologic Characteristics of Breast Cancer
    Park, Ji Yeon
    Yi, Seong Yoon
    Park, Hee Jin
    Kim, Mi Sung
    Kwon, Heon Ju
    Park, Noh Hyuck
    Moon, Soon Young
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (01) : 223 - 228