Comparison of breast specific gamma imaging and molecular breast tomosynthesis in breast cancer detection: Evaluation in phantoms

被引:15
|
作者
Gong, Zongyi [1 ]
Williams, Mark B. [1 ,2 ,3 ]
机构
[1] Univ Virginia, Dept Radiol & Med Imaging, Charlottesville, VA 22908 USA
[2] Univ Virginia, Dept Biomed Engn, Charlottesville, VA 22908 USA
[3] Univ Virginia, Dept Phys, Charlottesville, VA 22904 USA
关键词
breast specific gamma imaging; molecular imaging; nuclear medicine; breast; tomosynthesis; PROOF-OF-CONCEPT; HIGH-RISK; DENSITY; MODALITY; SCINTIMAMMOGRAPHY; MAMMOGRAPHY; SYSTEMS; CAMERA; WOMEN; DIAGNOSIS;
D O I
10.1118/1.4922398
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Breast specific gamma imaging or molecular breast imaging (BSGI) obtains 2D images of Tc-99m sestamibi distribution in the breast. Molecular breast tomosynthesis (MBT) maps the tracer distribution in 3D by acquiring multiple projections over a limited angular range. Here, the authors compare the performance of the two technologies in terms of spatial resolution, lesion contrast, and contrast-to-noise ratio (CNR) in phantom studies under conditions of clinically relevant sestamibi dose and imaging time. Methods: The systems tested were a Dilon 6800 and a MBT prototype developed at the University of Virginia. Both systems comprise a pixelated sodium iodide scintillator, an array of position sensitive photomultipliers, and a parallel hole collimator. The active areas and energy resolution of the systems are similar. System sensitivity, spatial resolution, lesion contrast, and CNR were measured using a Petri dish, a point source phantom, and a breast phantom containing simulated lesions at two depths, respectively. A single BSGI projection was acquired. Five MBT projections were acquired over +/- 20 degrees. For both modalities, the total scan count density was comparable to that observed for each in typical 10 min human scans following injection of 22 mCi (814 MBq) of Tc-99m-sestamibi. To assess the impact of reducing the tracer dose, the pixel counts of projection images were later binomially subsampled by a factor of 2 to give images corresponding to an injected activity of approximately 11 mCi (407 MBq). Both unprocessed (pixelated) BSGI projections and interpolated (smoothed) BSGI images displayed by default on the Dilon 6800 workstation were analyzed. Volumetric images were reconstructed from the MBT projections using a maximum likelihood expectation maximization algorithm and extracted slices were analyzed. Results: Over a depth range of 1.5-7.5 cm, BSGI spatial resolution was 5.6-11.5 mm in unprocessed projections and 5.7-12.0 mm in interpolated images. Over the same range, the in-slice MBT spatial resolution was 6.7-9.4 mm. Lesion contrast was significantly improved with MBT relative to BSGI for five out of eight lesions imaged at either the 22 mCi or the 11 mCi dose level (p < 0.05). At both dose levels, significant improvements in CNR with MBT were also found for five out of eight lesions (9.8, 7.8, 6.2 mm lesions at water depth of 1.7 cm and 9.8, 7.8 mm lesions at water depth of 4.5 cm, p < 0.05). The 6.2 and 4.9 mm lesions located at 4.5 cm below the water surface were not visible in either modality at either activity level. Conclusions: Under conditions of equal dose, imaging time and similar detectors, compared to BSGI, MBT provided higher lesion contrast, higher CNR, and spatial resolution that was less depth dependent. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:4250 / 4259
页数:10
相关论文
共 50 条
  • [31] Cancer Detection with Digital Breast Tomosynthesis
    Lisa Katherine Harris
    Vivian Mai-Tran
    Melissa Durand
    [J]. Current Radiology Reports, 8
  • [32] Development of digital breast tomosynthesis and diffuse optical tomography fusion imaging for breast cancer detection
    Chae, Eun Young
    Kim, Hak Hee
    Sabir, Sohail
    Kim, Yejin
    Kim, Hyeongseok
    Yoon, Sungho
    Ye, Jong Chul
    Cho, Seungryong
    Heo, Duchang
    Kim, Kee Hyun
    Bae, Young Min
    Choi, Young-Wook
    [J]. SCIENTIFIC REPORTS, 2020, 10 (01)
  • [33] Development of digital breast tomosynthesis and diffuse optical tomography fusion imaging for breast cancer detection
    Eun Young Chae
    Hak Hee Kim
    Sohail Sabir
    Yejin Kim
    Hyeongseok Kim
    Sungho Yoon
    Jong Chul Ye
    Seungryong Cho
    Duchang Heo
    Kee Hyun Kim
    Young Min Bae
    Young-Wook Choi
    [J]. Scientific Reports, 10
  • [34] Tomosynthesis imaging of the breast
    Niklason, L
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1369 - 1370
  • [35] Breast Specific Gamma Imaging
    Stern, L.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [36] Molecular Breast Imaging Using Emission Tomosynthesis
    Gopan, O.
    Gilland, D.
    Weisenberger, A.
    Kross, B.
    Welch, B.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [37] Microwave Breast Imaging: experimental tumour phantoms for the evaluation of new breast cancer diagnosis systems
    Oliveira, Barbara L.
    O'Loughlin, Declan
    O'Halloran, Martin
    Porter, Emily
    Glavin, Martin
    Jones, Edward
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (02):
  • [38] Tumor Characteristics and Molecular Subtypes in Breast Cancer Screening with Digital Breast Tomosynthesis: The Malmo Breast Tomosynthesis Screening Trial
    Johnson, Kristin
    Zackrisson, Sophia
    Rosso, Aldana
    Sartor, Hanna
    Saal, Lao H.
    Andersson, Ingvar
    Lang, Kristina
    [J]. RADIOLOGY, 2019, 293 (02) : 273 - 281
  • [39] Effect of Dose Level on Radiologists' Detection of Microcalcifications in Digital Breast Tomosynthesis: An Observer Study with Breast Phantoms
    Chan, Heang-Ping
    Helvie, Mark A.
    Klein, Katherine A.
    McLaughlin, Carol
    Neal, Colleen H.
    Oudsema, Rebecca
    Rahman, W. Tania
    Roubidoux, Marilyn A.
    Hadjiiski, Lubomir M.
    Zhou, Chuan
    Samala, Ravi K.
    [J]. ACADEMIC RADIOLOGY, 2022, 29 : S42 - S49
  • [40] Comparison Study of Reconstruction Algorithms for Prototype Digital Breast Tomosynthesis Using Various Type of Breast Phantoms
    Kim, Y.
    Kim, H.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3252 - 3252