Comparison of breast specific gamma imaging and molecular breast tomosynthesis in breast cancer detection: Evaluation in phantoms

被引:15
|
作者
Gong, Zongyi [1 ]
Williams, Mark B. [1 ,2 ,3 ]
机构
[1] Univ Virginia, Dept Radiol & Med Imaging, Charlottesville, VA 22908 USA
[2] Univ Virginia, Dept Biomed Engn, Charlottesville, VA 22908 USA
[3] Univ Virginia, Dept Phys, Charlottesville, VA 22904 USA
关键词
breast specific gamma imaging; molecular imaging; nuclear medicine; breast; tomosynthesis; PROOF-OF-CONCEPT; HIGH-RISK; DENSITY; MODALITY; SCINTIMAMMOGRAPHY; MAMMOGRAPHY; SYSTEMS; CAMERA; WOMEN; DIAGNOSIS;
D O I
10.1118/1.4922398
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Breast specific gamma imaging or molecular breast imaging (BSGI) obtains 2D images of Tc-99m sestamibi distribution in the breast. Molecular breast tomosynthesis (MBT) maps the tracer distribution in 3D by acquiring multiple projections over a limited angular range. Here, the authors compare the performance of the two technologies in terms of spatial resolution, lesion contrast, and contrast-to-noise ratio (CNR) in phantom studies under conditions of clinically relevant sestamibi dose and imaging time. Methods: The systems tested were a Dilon 6800 and a MBT prototype developed at the University of Virginia. Both systems comprise a pixelated sodium iodide scintillator, an array of position sensitive photomultipliers, and a parallel hole collimator. The active areas and energy resolution of the systems are similar. System sensitivity, spatial resolution, lesion contrast, and CNR were measured using a Petri dish, a point source phantom, and a breast phantom containing simulated lesions at two depths, respectively. A single BSGI projection was acquired. Five MBT projections were acquired over +/- 20 degrees. For both modalities, the total scan count density was comparable to that observed for each in typical 10 min human scans following injection of 22 mCi (814 MBq) of Tc-99m-sestamibi. To assess the impact of reducing the tracer dose, the pixel counts of projection images were later binomially subsampled by a factor of 2 to give images corresponding to an injected activity of approximately 11 mCi (407 MBq). Both unprocessed (pixelated) BSGI projections and interpolated (smoothed) BSGI images displayed by default on the Dilon 6800 workstation were analyzed. Volumetric images were reconstructed from the MBT projections using a maximum likelihood expectation maximization algorithm and extracted slices were analyzed. Results: Over a depth range of 1.5-7.5 cm, BSGI spatial resolution was 5.6-11.5 mm in unprocessed projections and 5.7-12.0 mm in interpolated images. Over the same range, the in-slice MBT spatial resolution was 6.7-9.4 mm. Lesion contrast was significantly improved with MBT relative to BSGI for five out of eight lesions imaged at either the 22 mCi or the 11 mCi dose level (p < 0.05). At both dose levels, significant improvements in CNR with MBT were also found for five out of eight lesions (9.8, 7.8, 6.2 mm lesions at water depth of 1.7 cm and 9.8, 7.8 mm lesions at water depth of 4.5 cm, p < 0.05). The 6.2 and 4.9 mm lesions located at 4.5 cm below the water surface were not visible in either modality at either activity level. Conclusions: Under conditions of equal dose, imaging time and similar detectors, compared to BSGI, MBT provided higher lesion contrast, higher CNR, and spatial resolution that was less depth dependent. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:4250 / 4259
页数:10
相关论文
共 50 条
  • [1] Phantoms for Digital Breast Tomosynthesis Imaging System Evaluation
    Jacobson, D.
    Liu, Y.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3329 - 3329
  • [2] Breast-specific Gamma Imaging for Detection of Breast Cancer Response
    Brem, Rachel F.
    Floerke, Angelique C.
    Rapelyea, Jocelyn M.
    [J]. RADIOLOGY, 2009, 251 (01) : 306 - 306
  • [3] Imaging features of breast cancers on digital breast tomosynthesis according to molecular subtype: association with breast cancer detection
    Lee, Su Hyun
    Chang, Jung Min
    Shin, Sung Ui
    Chu, A. Jung
    Yi, Ann
    Cho, Nariya
    Moon, Woo Kyung
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1080):
  • [4] A Prototype Gamma Tomosynthesis System for Molecular Breast Imaging
    Gilland, David
    Welch, Benjamin L.
    Lee, Seung Joon
    Kross, B.
    Weisenberger, Andrew G.
    [J]. 2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [5] Semiquantitative evaluation of breast-specific gamma imaging for the diagnosis of breast cancer: Comparison with mammography and sonography
    Chung, Hyun Woo
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [6] Improved breast cancer detection rates using breast-specific gamma imaging
    Rose, Sam
    [J]. FUTURE ONCOLOGY, 2012, 8 (04) : 347 - 348
  • [7] Diagnostic value of breast specific gamma imaging for breast cancer
    Mallik, D. K.
    Zhang, H.
    Shi, H.
    Tan, H.
    Shen, L.
    Hu, G.
    Hu, H.
    Sejdinaj, F.
    Pang, Y.
    Zhu, W.
    Yang, W.
    [J]. CANCER RESEARCH, 2017, 77
  • [8] UWB MICROWAVE IMAGING FOR BREAST CANCER DETECTION - EXPERIMENTS WITH HETEROGENEOUS BREAST PHANTOMS
    Lai, J. C. Y.
    Soh, C. B.
    Gunawan, E.
    Low, K. S.
    [J]. PROGRESS IN ELECTROMAGNETICS RESEARCH M, 2011, 16 : 19 - 29
  • [9] Molecular Breast Imaging: A multicenter clinical registry to compare breast-specific gamma imaging (BSGI) and breast MRI in the detection of breast carcinoma
    Weigert, J. M.
    Kieper, D. A.
    Stern, L. H.
    Bohm-Velez, M.
    [J]. CANCER RESEARCH, 2012, 72
  • [10] Comparison of MRI and digital breast tomosynthesis in the preoperative evaluation of multifocal breast cancer
    Batohi, Bhavna
    Vinci, Valeria
    Peacock, Clare
    Michell, Michael
    Iqbal, Asif
    Evans, David
    Morel, Juliet
    Satchithananda, Keshthra
    Wasan, Reema
    Rahim, Rumana
    [J]. BREAST CANCER RESEARCH, 2015, 17