Mapping and quantifying hyperpolarized 3He magnetic resonance imaging apparent diffusion coefficient gradients

被引:21
|
作者
Evans, Andrea [1 ,2 ]
McCormack, David G. [3 ,4 ]
Santyr, Giles [1 ,2 ,5 ]
Parraga, Grace [1 ,2 ,4 ,5 ]
机构
[1] Univ Western Ontario, Robarts Res Inst, Imaging Res Labs, London, ON N6A 5K8, Canada
[2] Univ Western Ontario, Dept Med Biophys, London, ON N6A 5K8, Canada
[3] Univ Western Ontario, Dept Med, Div Respirol, London, ON N6A 5K8, Canada
[4] Univ Western Ontario, Lawson Hlth Res Inst, London, ON N6A 5K8, Canada
[5] Univ Western Ontario, Dept Radiol & Nucl Med, London, ON N6A 5K8, Canada
关键词
emphysema; chronic obstructive pulmonary disease; thoracic imaging; noble gas lung imaging;
D O I
10.1152/japplphysiol.00178.2008
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
We measured hyperpolarized He-3 magnetic resonance imaging (MRI) apparent diffusion coefficients (ADC) and quantified ADC gradients in each three-by-three voxel region of interest (ROI). Such local ADC gradients can be represented in vector maps showing the magnitude (vertical bar G(3x3)vertical bar) and direction of ADC gradients, providing a qualitative visualization tool and quantitative measurement of airway and air space heterogeneity. Twenty-four subjects (15 male, mean age = 67 +/- 7 yr) with global initiative for chronic obstructive lung disease (GOLD) stage II (n = 9, mean age 68 +/- 6 yr), GOLD stage III chronic obstructive pulmonary disease (COPD; n = 7, mean age 67 +/- 8 yr), and age- matched healthy volunteers (n = 8, mean age 67 +/- 6 yr) were enrolled based on their age and spirometry results. Hyperpolarized 3He MRI was performed on a whole body 3.0 Tesla system. Mean He-3 ADC and ADC standard deviation were calculated for the center coronal slice, and the mean magnitude and direction of the ADC gradient vectors were calculated for each three- by- three voxel matrix (vertical bar G(3x3)vertical bar). While the He-3 ADC standard deviation was not significantly different, mean vertical bar G(3x3)vertical bar was significantly different between subjects with stage II (0.14 +/- 0.03 cm/s) and stage III COPD (0.19 +/- 0.03 cm/s; P < 0.005) and between healthy subjects (0.12 +/- 0.03 cm/ s) and those with stage II COPD (P < 0.02). The second order statistic vertical bar G(3x3)vertical bar may provide a sensitive measure of ADC heterogeneity for ROI representing 9.4 x 9.4 x 30 mm or 2.6 cm(3) of lung tissue.
引用
收藏
页码:693 / 699
页数:7
相关论文
共 50 条
  • [31] Dependence of apparent diffusion coefficient on slice position in magnetic resonance diffusion imaging
    Takatsu, Yasuo
    Nakamura, Masafumi
    Suzuki, Yuichi
    Miyati, Tosiaki
    MAGNETIC RESONANCE IMAGING, 2023, 99 : 41 - 47
  • [32] Imaging parameter effects in apparent diffusion coefficient determination of magnetic resonance imaging
    Ogura, Akio
    Hayakawa, Katsumi
    Miyati, Tosiaki
    Maeda, Fumie
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 77 (01) : 185 - 188
  • [33] Hyperpolarized 3He magnetic resonance imaging: Preliminary evaluation of phenotyping potential in chronic obstructive pulmonary disease
    Mathew, Lindsay
    Kirby, Miranda
    Etemad-Rezai, Roya
    Wheatley, Andrew
    McCormack, David G.
    Parraga, Grace
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 79 (01) : 140 - 146
  • [34] Two and Three-Dimensional Segmentation of Hyperpolarized 3He Magnetic Resonance Imaging of Pulmonary Gas Distribution
    Heydarian, Mohammadreza
    Kirby, Miranda
    Wheatley, Andrew
    Fenster, Aaron
    Parraga, Grace
    MEDICAL IMAGING 2012: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2012, 8317
  • [35] Ventilation Defects Observed with Hyperpolarized 3He Magnetic Resonance Imaging in a Mouse Model of Acute Lung Injury
    Thomas, Abe C.
    Nouls, John C.
    Driehuys, Bastiaan
    Voltz, James W.
    Fubara, Boma
    Foley, Julie
    Bradbury, J. Alyce
    Zeldin, Darryl C.
    AMERICAN JOURNAL OF RESPIRATORY CELL AND MOLECULAR BIOLOGY, 2011, 44 (05) : 648 - 654
  • [36] Hyperpolarized 3He functional magnetic resonance imaging of bronchoscopic airway bypass in chronic obstructive pulmonary disease
    Mathew, Lindsay
    Kirby, Miranda
    Farquhar, Donald
    Licskai, Christopher
    Santyr, Giles
    Etemad-Rezai, Roya
    Parraga, Grace
    McCormack, David G.
    CANADIAN RESPIRATORY JOURNAL, 2012, 19 (01) : 41 - 43
  • [37] Magnetic resonance imaging using hyperpolarized 3He-gas
    Gast, KK
    Eberle, B
    Schmiedeskamp, J
    Kauczor, HU
    ACADEMIC RADIOLOGY, 2003, 10 (10) : 1119 - 1131
  • [38] Diffusion lung imaging of hyperpolarized 3He in patients with emphysema: Changes in the structure of alveoli
    Yablonskiy, DA
    Saam, BT
    Gierada, DS
    Lefrak, SS
    Cooper, JD
    Conradi, MS
    RADIOLOGY, 1999, 213P : 344 - 344
  • [39] Diffusion-Weighted Magnetic Resonance Imaging and Apparent Diffusion Coefficient Mapping for Diagnosing Infectious Spondylodiscitis: A Preliminary Study
    Chen, Tai-Yuan
    Wu, Te-Chang
    Tsui, Yu-Kun
    Chen, Hou-Hsun
    Lin, Chien-Jen
    Lee, Huey-Jen
    Wu, Tai-Ching
    JOURNAL OF NEUROIMAGING, 2015, 25 (03) : 482 - 487
  • [40] The influence of field strength on the apparent diffusion coefficient of 3He gas in human lungs
    Parra-Robles, J.
    Ajraoui, S.
    Marshall, H.
    Deppe, M. H.
    Xu, X.
    Wild, J. M.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (02) : 322 - 325