A novel deformable lung phantom with programably variable external and internal correlation

被引:7
|
作者
Ranjbar, Maida [1 ]
Sabouri, Pouya [1 ]
Repetto, Carlo [1 ]
Sawant, Amit [1 ]
机构
[1] Univ Maryland, Dept Radiat Oncol, Sch Med, Baltimore, MD 21201 USA
基金
美国国家卫生研究院;
关键词
deformable lung phantom; programmable; respiratory motion management; surrogate-based lung motion model; variable correlation; REAL-TIME TUMOR; RESPIRATORY MOTION; GATED RADIOTHERAPY; IMAGE REGISTRATION; QUALITY-ASSURANCE; LOCALIZATION; VERIFICATION; MOVEMENT; SYSTEM; CANCER;
D O I
10.1002/mp.13507
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Lung motion phantoms used to validate radiotherapy motion management strategies have fairly simplistic designs that do not adequately capture complex phenomena observed in human respiration such as external and internal deformation, variable hysteresis and variable correlation between different parts of the thoracic anatomy. These limitations make reliable evaluation of sophisticated motion management techniques quite challenging. In this work, we present the design and implementation of a programmable, externally and internally deformable lung motion phantom that allows for a reproducible change in external-internal and internal-internal correlation of embedded markers. Methods An in-house-designed lung module, made from natural latex foam was inserted inside the outer shell of a commercially available lung phantom (RSD, Long Beach, CA, USA). Radiopaque markers were placed on the external surface and embedded into the lung module. Two independently programmable high-precision linear motion actuators were used to generate primarily anterior-posterior (AP) and primarily superior-inferior (SI) motion in a reproducible fashion in order to enable (a) variable correlation between the displacement of interior volume and the exterior surface, (b) independent changes in the amplitude of the AP and SI motions, and (c) variable hysteresis. The ability of the phantom to produce complex and variable motion accurately and reproducibly was evaluated by programming the two actuators with mathematical and patient-recorded lung tumor motion traces, and recording the trajectories of various markers using kV fluoroscopy. As an example application, the phantom was used to evaluate the performance of lung motion models constructed from kV fluoroscopy and 4DCT images. Results The phantom exhibited a high degree of reproducibility and marker motion ranges were reproducible to within 0.5 mm. Variable correlation was observed between the displacements of internal-internal and internal-external markers. The SI and AP components of motion of a specific marker had a correlation parameter that varied from -11 to 17. Monitoring a region of interest on the phantom's surface to estimate internal marker motion led to considerably lower uncertainties than when a single point was monitored. Conclusions We successfully designed and implemented a programmable, externally and internally deformable lung motion phantom that allows for a reproducible change in external-internal and internal-internal correlation of embedded markers.
引用
收藏
页码:1995 / 2005
页数:11
相关论文
共 50 条
  • [21] Development of a deformable lung phantom with 3D-printed flexible airways
    Shin, Dong-Seok
    Kang, Seong-Hee
    Kim, Kyeong-Hyeon
    Kim, Tae-Ho
    Kim, Dong-Su
    Chung, Jin-Beom
    Lucero, Steven Andrew
    Suh, Tae Suk
    Yamamoto, Tokihiro
    MEDICAL PHYSICS, 2020, 47 (03) : 898 - 908
  • [22] Design and Geometric Validation of An Externally and Internally Deformable, Programmable Lung Motion Phantom
    Cheung, Y.
    Sawant, A.
    MEDICAL PHYSICS, 2014, 41 (06) : 98 - 98
  • [23] Machine Learning Radiomics Model for External and Internal Respiratory Motion Correlation Prediction in Lung Tumor
    Zhang, Xiangyu
    Song, Xinyu
    Li, Guangjun
    Duan, Lian
    Wang, Guangyu
    Dai, Guyu
    Song, Ying
    Li, Jing
    Bai, Sen
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2022, 21
  • [24] Audiovisual biofeedback improves the correlation between internal/external surrogate motion and lung tumor motion
    Lee, Danny
    Greer, Peter B.
    Paganelli, Chiara
    Ludbrook, Joanna Jane
    Kim, Taeho
    Keall, Paul
    MEDICAL PHYSICS, 2018, 45 (03) : 1009 - 1017
  • [25] Evaluation of Internal/External Correlation with Missed Volume
    Wu, H.
    He, C.
    Zhao, Q.
    Berbeco, R.
    Shirato, H.
    Nishioka, S.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [26] Realistic reference adult and paediatric phantom series for internal and external dosimetry
    Stabin, M. G.
    Emmons, M. A.
    Segars, W. P.
    Fernald, M. J.
    RADIATION PROTECTION DOSIMETRY, 2012, 149 (01) : 56 - 59
  • [27] Characterization of a novel anthropomorphic plastinated lung phantom
    Yoon, S.
    Henry, R.
    Bouley, D.
    Fahrig, R.
    MEDICAL PHYSICS, 2006, 33 (06) : 1998 - 1998
  • [28] Variable Composition of the Internal and External Branches of the Accessory Nerve
    Liu, Hong-Fu
    Won, Hyung-Sun
    Chung, In-Hyuk
    Oh, Chang-Seok
    Kim, In-Beom
    CLINICAL ANATOMY, 2014, 27 (01) : 97 - 101
  • [29] INTERNAL-EXTERNAL CONTROL AS AN ENVIRONMENTAL VARIABLE IN LISTENING
    MATHIS, RW
    JAMES, WH
    JOURNAL OF EXPERIMENTAL EDUCATION, 1972, 40 (03): : 60 - &
  • [30] Characterization of a novel anthropomorphic plastinated lung phantom
    Yoon, Sungwon
    Henry, Robert W.
    Bouley, Donna M.
    Bennett, N. Robert
    Fahrig, Rebecca
    MEDICAL PHYSICS, 2008, 35 (12) : 5934 - 5943