Study on the response of indigenously developed CaSO4:Dy phosphor-based neutron dosemeter

被引:4
|
作者
Bakshi, A. K. [1 ]
Pradhan, A. S. [1 ]
Kher, R. K. [1 ]
Srivastava, K. [1 ]
Varadharajan, G. [1 ]
Chatterjee, S. [1 ]
Sathian, V. [2 ]
机构
[1] Bhabha Atom Res Ctr, Radiol Phys & Advisory Div, Mumbai 400085, Maharashtra, India
[2] Bhabha Atom Res Ctr, Radiat Safety Syst Div, Mumbai 400085, Maharashtra, India
关键词
MIXED FIELD DOSIMETRY; DY;
D O I
10.1093/rpd/ncp019
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In the present paper, we report the indigenous development of a neutron-sensitive thermoluminescent dosemeter (TLD) based on CaSO4:Dy Teflon TL disc. The study includes indigenous development of a neutron dosemeter and its response in terms of operational quantity to different energies of neutrons under various irradiation conditions. It was found that the thermal neutron sensitivity of the CaSO4:Dy Teflon neutron disc is about one-third of TLD-600. However, the thermal neutron sensitivity with respect to CaSO4:Dy Teflon gamma disc is about 42 times for in-air irradiation and about 84 times for on-phantom irradiation condition. This newly developed neutron dosemeter can be used as a routine TLD with a slight change in the design of the TLD badge system in the mixed fields of gamma and neutrons of energy up to 500 keV for radiation workers engaged in nuclear fuel cycle operation and other applications.
引用
收藏
页码:73 / 80
页数:8
相关论文
共 50 条
  • [1] RESPONSE OF CASO4(DY) PHOSPHOR TO NEUTRONS
    WENG, PS
    CHEN, KM
    NUCLEAR INSTRUMENTS & METHODS, 1974, 117 (01): : 89 - 92
  • [2] Study of Residual TL of CaSO4:Dy-based Thermoluminescence Dosemeter
    Sahai, Manish K.
    Pradhan, S. M.
    Bakshi, A. K.
    Datta, D.
    RADIATION PROTECTION DOSIMETRY, 2017, 177 (03) : 342 - 347
  • [3] Mechano and thermoluminescence of γ-irradiated CaSO4:Dy phosphor
    Kher, RS
    Pandey, RK
    Dhoble, SJ
    Khokhar, MSK
    SOLID STATE DOSIMETRY, PTS 1 AND 2, PROCEEDINGS, 2002, : 281 - 284
  • [4] Mechano and thermoluminescence of γ-irradiated CaSO4:Dy phosphor
    Kher, RS
    Pandey, RK
    Dhoble, SJ
    Khokhar, MSK
    RADIATION PROTECTION DOSIMETRY, 2002, 100 (1-4) : 281 - 284
  • [5] Features of EPR dosimetry with CaSO4:Dy phosphor
    Bortolin, E
    Onori, S
    APPLIED RADIATION AND ISOTOPES, 2005, 62 (02) : 349 - 352
  • [6] Defect centres and thermoluminescence in CaSO4:Dy,Ag phosphor
    Bhatt, B. C.
    Dhabekar, Bhushan
    Kumar, Rajesh
    Rao, T. K. Gundu
    Lakshmanan, A. R.
    RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) : 53 - 56
  • [7] Response of CaSO4:Dy based TL dosemeter system to high-energy photon beams:: Theoretical simulation
    Bakshi, A. K.
    Chatterjee, S.
    Kher, R. K.
    RADIATION PROTECTION DOSIMETRY, 2008, 129 (04) : 456 - 462
  • [8] CaSO4:Dy microphosphor for thermal neutron dosimetry
    Bhadane, Mahesh S.
    Mandlik, Nandkumar
    Patil, B. J.
    Dahiwale, S. S.
    Sature, K. R.
    Bhoraskar, V. N.
    Dhole, S. D.
    JOURNAL OF LUMINESCENCE, 2016, 170 : 226 - 230
  • [9] Operational experience of electronic active personal dosemeter and comparison with CaSo4:Dy TL dosemeter in Indian PHWR
    Singh, Vishwanath P.
    Managanvi, S. S.
    Bihari, R. R.
    Bhat, H. R.
    RADIATION PROTECTION DOSIMETRY, 2013, 156 (01) : 93 - 102
  • [10] STABILITY OF CASO4 - DY THERMOLUMINESCENT PHOSPHOR IN HIGH DOSE REGION
    SRIVASTAVA, JK
    SUPE, SJ
    NUCLEAR INSTRUMENTS & METHODS, 1978, 155 (1-2): : 233 - 235