Accuracy Evaluation of NIM5 Cesium Fountain Clock

被引:6
|
作者
Liu Nian-Feng [1 ]
Fang Fang [1 ]
Chen Wei-Liang [1 ]
Lin Ping-Wei [1 ]
Wang Ping [1 ]
Liu Kun [1 ]
Suo Rui [1 ]
Li Tian-Chu [1 ]
机构
[1] Natl Inst Metrol, Beijing 100013, Peoples R China
基金
中国国家自然科学基金;
关键词
PRIMARY FREQUENCY STANDARD; SHIFT;
D O I
10.1088/0256-307X/30/1/010601
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The NIM5 fountain clock is the second fountain clock built at NIM (National Institute of Metrology, China), and has been operating stably and sub-continually since 2008. The fountain operates with a simple one-stage optical molasses to collect cold atoms, which reduces the collisional frequency shift dramatically. The fractional frequency uncertainty is estimated to be 2 x 10(-15). The typical frequency instability of 2.5 x 10(-14) is obtained at 10s. Comparisons with other fountain frequency standards worldwide demonstrate agreement within the stated uncertainties.
引用
收藏
页数:4
相关论文
共 50 条
  • [31] The Research of control system for Cesium atomic fountain clock
    Liu, Dandan
    Wang, Xinliang
    Lin, Rui
    Chen, Jiang
    Zhang, Hui
    Ruan, Jun
    Guan, Yong
    Yu, Fengxiang
    Shi, Junru
    Zhang, Shougang
    2014 IEEE INTERNATIONAL FREQUENCY CONTROL SYMPOSIUM (FCS), 2014, : 391 - 394
  • [32] Progress in building NRC's cesium fountain clock
    Marmet, Louis
    Dube, Pierre
    Gigault, Christian
    2005 IEEE INTERNATIONAL FREQUENCY CONTROL SYMPOSIUM AND EXHIBITION, 2005, : 266 - 269
  • [33] From microwave frequency to optical wavelength at NIM: Cooled atomic Cesium fountain clock-comb optical frequencies - Stabilized lasers
    Li, TC
    Fang, ZJ
    HIGH-POWER LASERS AND APPLICATIONS II, 2002, 4914 : 1 - 5
  • [34] Compact Laser System for Small Cesium Atomic Fountain Clock
    Jin, Shuanghao
    Lin, Pingwei
    Ru, Ning
    Ma, Yanning
    Jiang, Wensong
    Luo, Zai
    Guangxue Xuebao/Acta Optica Sinica, 2024, 44 (19):
  • [35] Improvement of the frequency stability of the cesium fountain clock by data processing
    Li, Zhiqi
    Yang, Ning
    Liu, Suyan
    Zhou, Wei
    AIP ADVANCES, 2021, 11 (12)
  • [36] Priliminary results from the USNO cesium fountain atomic clock
    Burt, E
    Swanson, T
    Ekstrom, C
    2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 10 - 11
  • [37] Initial Research of Frequency Prediction for Hydrogen Maser in New TA(NIM) Based on NIM4 Fountain Clock
    Liu Xiaojin
    Zhu Jiangmiao
    Gao Yuan
    Liu Mingliang
    PROCEEDINGS OF THE SECOND INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOL. 3, 2008, : 1496 - 1499
  • [38] Stability and accuracy results with the ON/METAS continuous cesium fountain
    Joyet, A
    Mileti, G
    Thomann, P
    Dudle, G
    Mandache, C
    Ascente, T
    2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST, 2002, : 442 - 443
  • [39] An ion clock reaches the accuracy of the best atomic fountain
    Levi, BG
    PHYSICS TODAY, 1998, 51 (02) : 21 - 23
  • [40] Development of cesium fountain clock as local precision time and frequency standard
    Li H.
    Du Y.
    Liu H.
    Lu Z.
    Huazhong Keji Daxue Xuebao (Ziran Kexue Ban)/Journal of Huazhong University of Science and Technology (Natural Science Edition), 2022, 50 (09): : 23 - 29