Effect of cementless acetabular cup geometry on strain distribution and press-fit stability

被引:38
|
作者
Ries, MD
Harbaugh, M
Shea, J
Lambert, R
机构
[1] Mary Imogene Bassett Hospital, Cooperstown, NY
[2] State University of New York, Stony Brook, NY
[3] Smith and Nephew Orthopaedics, Memphis, TN
[4] Mary Imogene Bassett Hospital, Cooperstown, NY 13326, One Atwell Road
来源
JOURNAL OF ARTHROPLASTY | 1997年 / 12卷 / 02期
关键词
acetabular cup; strain; stability; press-fit; cementless; implant;
D O I
10.1016/S0883-5403(97)90068-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Use of cementless acetabular cups, which are slightly larger than the reamed acetabulum, can provide press-fit stability without screws; however, the ideal cup geometry to maximize stability is not clear. Acetabular strain distribution, deformation, and implant stability were studied using an axisymmetric finite-element model, and mechanical stability was assessed by testing lever-out and extraction forces required to displace different cup geometries from foam bones. The implants tested included four nonhemispheric cup geometries and 1- and 2-mm oversized hemispheric geometries. A nonhemispheric cup that provides a gradual transition from a hemisphere at the dome to a larger peripheral dimension appears to maximize peripheral strains and implant stability without increasing overall acetabular deformation as much as a larger oversized hemispheric cup.
引用
收藏
页码:207 / 212
页数:6
相关论文
共 50 条
  • [21] Monitoring the press-fit insertion of an acetabular cup by impact measurements: Influence of bone abrasion
    Michel, Adrien
    Bosc, Romain
    Mathieu, Vincent
    Hernigou, Philippe
    Haiat, Guillaume
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2014, 228 (10) : 1027 - 1034
  • [22] A novel approach to determine primary stability of acetabular press-fit cups
    Weissmann, Volker
    Boss, Christian
    Bader, Rainer
    Hansmann, Harald
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2018, 80 : 1 - 10
  • [23] A novel method to assess primary stability of press-fit acetabular cups
    Crosnier, Emilie A.
    Keogh, Patrick S.
    Miles, Anthony W.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2014, 228 (11) : 1126 - 1134
  • [25] Late loosening of press-fit cementless acetabular components (vol 418, pg 174, 2004)
    Jiranek, WA
    Whiddon, DR
    Johnstone, WT
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (420) : 328 - 328
  • [26] MICROMOTION OF CEMENTLESS HEMISPHERICAL ACETABULAR COMPONENTS - DOES PRESS-FIT NEED ADJUNCTIVE SCREW FIXATION
    WON, CH
    HEARN, TC
    TILE, M
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (03): : 484 - 489
  • [27] Acetabular component deformation with press-fit fixation
    Squire, Matthew
    Griffin, William L.
    Mason, Bohannon
    Peindl, Richard D.
    Odum, Susan
    [J]. JOURNAL OF ARTHROPLASTY, 2006, 21 (06): : 72 - 77
  • [28] Periprosthetic occult acetabular fracture: an unknown side effect of press-fit techniques in primary cementless total hip arthroplasty
    Ho Hyun Yun
    Seung Hee Cheon
    Jung Taek Im
    Yeong Yoon Koh
    [J]. European Journal of Orthopaedic Surgery & Traumatology, 2021, 31 : 1411 - 1419
  • [29] Periprosthetic occult acetabular fracture: an unknown side effect of press-fit techniques in primary cementless total hip arthroplasty
    Yun, Ho Hyun
    Cheon, Seung Hee
    Im, Jung Taek
    Koh, Yeong Yoon
    [J]. EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2021, 31 (07): : 1411 - 1419
  • [30] Analysis of early loosening after revison cup arthroplasty using structural allografts and a cementless press-fit cup.
    Krüger, T
    Reicher, H
    Grubitzsch, U
    Hein, W
    [J]. ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2000, 138 (03): : 209 - 214