PHOTONEUTRON CONTAMINATION FROM AN 18 MV SATURNE MEDICAL LINEAR ACCELERATOR IN THE TREATMENT ROOM

被引:9
|
作者
Khosravi, Mostafa [1 ,2 ]
Shahbazi-Gahrouei, Daryoush [1 ]
Jabbari, Keyvan [1 ]
Nasri-Nasrabadi, Mehdi [3 ]
Baradaran-Ghahfarokhi, Milad [1 ]
Siavashpour, Zahra [4 ]
Gheisari, Ruhollah [2 ,5 ]
Amiri, Behnam [6 ]
机构
[1] Isfahan Univ Med Sci, Sch Med, Dept Med Phys & Med Engn, Esfahan, Iran
[2] Radiat & Pollut Protect Ctr Mamasani, Fars, Iran
[3] Univ Isfahan, Sch Adv Sci & Technol, Dept Nucl Engn, Esfahan, Iran
[4] Shahid Beheshti Univ, Dept Med Radiat Engn, Tehran, Iran
[5] Persian Gulf Univ, Sch Sci, Dept Phys, Booshehr, Iran
[6] Sharif Tecnol Univ, Dept Mech Engn, Tehran, Iran
关键词
MONTE-CARLO; RADIATION-THERAPY; SIEMENS; VARIAN; ENTRANCE; SPECTRA; LINACS; BEAMS; MODEL; CODE;
D O I
10.1093/rpd/nct078
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Dose escalation with high-energy X rays of medical linear accelerators (linacs) in radiotherapy offers several distinct advantages over the lower energy photons. However, owing to photoneutron reactions, interaction of high-energy photons (>8 MV) with various high-Z nuclei of the materials in the linac head components produces unavoidable neutrons. The aim of this study was to evaluate the photoneutron dose equivalent per unit therapeutic X-ray dose of 18 MV, GE Saturne 20 linac in the treatment room using Monte Carlo (MC) MCNP linac head full simulation as well as thermoluminescence dosemeter measurements. This machine is one of the old linac models manufactured by General Electric Company; however, it is widely used in the developing countries because of low cost and simple maintenance for radiotherapy applications. The results showed a significant photoneutron dose from Saturne 20 linac head components especially at distances near the linac head (<150 cm). Results of this work could be used in several applications, especially designing bunker and entrance door shielding against neutrons produced by photoneutron reactions in GE Saturne 20. However, a detailed cost optimisation for a specific room would require a dedicated calculation.
引用
收藏
页码:356 / 363
页数:8
相关论文
共 50 条
  • [1] New aspect determination of photoneutron contamination in 18 MV medical linear accelerator
    Mohammadi, A.
    Afarideh, H.
    Davani, F. Abbasi
    Arbabi, A.
    RADIATION MEASUREMENTS, 2016, 95 : 55 - 61
  • [2] INVESTIGATION OF PHOTONEUTRON CONTAMINATION FROM THE 18-MV PHOTON BEAM IN A MEDICAL LINEAR ACCELERATOR
    Karaman, Onur
    Tanir, Ayse Gunes
    Karaman, Ceren
    MATERIALI IN TEHNOLOGIJE, 2019, 53 (05): : 699 - 704
  • [3] MCNPX Estimation of Photoneutron Dose to Eye Voxel Anthropomorphic Phantom From 18 MV Linear Accelerator
    Alghamdi, Ali A. A.
    DOSE-RESPONSE, 2023, 21 (02):
  • [4] Photoneutron Flux Measurement via NAA in a Radiotherapy Bunker with an 18 MV Linear Accelerator
    Gulumser, T.
    Cecen, Y.
    Yesil, A. H.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S764 - S765
  • [5] Effect of wedge filter and field size on photoneutron dose equivalent for an 18 MV photon beam of a medical linear accelerator
    Mesbahi, Asghar
    Keshtkar, Ahmad
    Mohammadi, Ehsan
    Mohammadzadeh, Mohammad
    APPLIED RADIATION AND ISOTOPES, 2010, 68 (01) : 84 - 89
  • [6] Photoneutron Flux Measurement via Neutron Activation Analysis in a Radiotherapy Bunker with an 18 MV Linear Accelerator
    Cecen, Yigit
    Gulumser, Tugce
    Yazgan, Cagri
    Dapo, Haris
    Ustun, Mahmut
    Boztosun, Ismail
    ICRS-13 & RPSD-2016, 13TH INTERNATIONAL CONFERENCE ON RADIATION SHIELDING & 19TH TOPICAL MEETING OF THE RADIATION PROTECTION AND SHIELDING DIVISION OF THE AMERICAN NUCLEAR SOCIETY - 2016, 2017, 153
  • [7] Characteristics of the photoneutron contamination of a medical linear accelerator: Yields and spectra vs. field size
    Garnica-Garza, HM
    MEDICAL PHYSICS, 2004, 31 (06) : 1868 - 1868
  • [8] Photoneutron dose levels in a linear accelerator room with lead shielding
    Li, Z
    Klein, E
    Esthappan, J
    Drzymala, R
    Purdy, J
    MEDICAL PHYSICS, 2002, 29 (06) : 1325 - 1325
  • [9] Evaluation of the photoneutron field produced in a medical linear accelerator
    Kim, H. S.
    Park, Y. H.
    Koo, B. C.
    Kwon, J. W.
    Lee, J.-S.
    Choi, H. S.
    RADIATION PROTECTION DOSIMETRY, 2007, 123 (03) : 323 - 328
  • [10] GEANT4 Simulation of Photoneutron Spectrum from Medical Linear Accelerator
    Chernyaev, Alexander
    Belikhin, Mikhail
    Lykova, Ekaterina
    Shcherbakov, Alexey
    QUANTUM BEAM SCIENCE, 2023, 7 (03)