3D image-guided robotic needle positioning system for small animal interventions

被引:8
|
作者
Bax, Jeffrey S. [1 ,2 ]
Waring, Christopher S. R. [1 ,2 ]
Sherebrin, Shi [2 ]
Stapleton, Shawn [3 ]
Hudson, Thomas J. [2 ]
Jaffray, David A. [3 ]
Lacefield, James C. [1 ,2 ]
Fenster, Aaron [1 ,2 ]
机构
[1] Univ Western Ontario, Biomed Engn Grad Program, London, ON N6A 5B9, Canada
[2] Univ Western Ontario, Robarts Res Inst, London, ON N6A 5K8, Canada
[3] Univ Toronto, Radiat Med Program, Princess Margaret Hosp, Toronto, ON M5G 2M9, Canada
基金
加拿大健康研究院; 加拿大自然科学与工程研究理事会;
关键词
medical robotics; image guidance; ULTRASOUND; GUIDANCE; SURGERY; MICE; INJECTIONS; ASSISTANT; BIOPSY;
D O I
10.1118/1.4771958
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This paper presents the design of a micro-CT guided small animal robotic needle positioning system. In order to simplify the robotic design and maintain a small targeting error, a novel implementation of the remote center of motion is used in the system. The system has been developed with the objective of achieving a mean targeting error of <200 mu m while maintaining a high degree of user friendliness. Methods: The robot is compact enough to operate within a 25 cm diameter micro-CT bore. Small animals can be imaged and an intervention performed without the need to transport the animal from one workspace to another. Not requiring transport of the animal reduces opportunities for targets to shift from their localized position in the image and simplifies the workflow of interventions. An improved method of needle calibration is presented that better characterizes the calibration using the position of the needle tip in photographs rather than the needle axis. A calibration fixture was also introduced, which dramatically reduces the time requirements of calibration while maintaining calibration accuracy. Two registration modes have been developed to correspond the robot coordinate system with the coordinate system of the micro-CT scanner. The two registration modes offer a balance between the time required to complete a registration and the overall registration accuracy. The development of slow high accuracy and fast low accuracy registration modes provides users with a degree of flexibility in selecting a registration mode best suited for their application. Results: The target registration error (TRE) of the higher accuracy primary registration was TREprimary = 31 +/- 12 mu m. The error in the lower accuracy combined registration was TREcombined = 139 +/- 63 mu m. Both registration modes are therefore suitable for small-animal needle interventions. The targeting accuracy of the robotic system was characterized using targeting experiments in tissue-mimicking gelatin phantoms. The results of the targeting experiments were combined with the known calibration and needle deflection errors to provide a more meaningful measure of the needle positioning accuracy of the system. The combined targeting errors of the system were 149 +/- 41 mu m and 218 +/- 38 mu m using the primary and combined registrations, respectively. Finally, pilot in vivo experiments were successfully completed to demonstrate the performance of the system in a biomedical application. Conclusions: The device was able to achieve the desired performance with an error of <200 mu m and improved repeatability when compared to other designs. The device expands the capabilities of image-guided interventions for preclinical biomedical applications. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4771958]
引用
收藏
页数:15
相关论文
共 50 条
  • [1] 3D Image-Guided Robotic System for Bone Fracture Reduction
    Lee, Seongpung
    Joung, Sanghyun
    Ha, Ho-Gun
    Lee, Jin-Han
    Park, Kyeong-Hyeon
    Kim, Shinyeol
    Nam, Kwonsun
    Lee, Jongsuk
    Lee, Hyun-Joo
    Oh, Chang-Wug
    Park, Ilhyung
    Hong, Jaesung
    [J]. IEEE ROBOTICS AND AUTOMATION LETTERS, 2022, 7 (02): : 4353 - 4360
  • [2] An ultra-high field strength MR image-guided robotic needle delivery system for in-bore small animal interventions
    Gravett, Matthew
    Cepek, Jeremy
    Fenster, Aaron
    [J]. MEDICAL PHYSICS, 2017, 44 (11) : 5544 - 5555
  • [3] Bridging 3D Slicer and ROS2 for Image-Guided Robotic Interventions
    Connolly, Laura
    Deguet, Anton
    Leonard, Simon
    Tokuda, Junichi
    Ungi, Tamas
    Krieger, Axel
    Kazanzides, Peter
    Mousavi, Parvin
    Fichtinger, Gabor
    Taylor, Russell H.
    [J]. SENSORS, 2022, 22 (14)
  • [4] Development of an image-guided robotic system for surgical positioning and drilling
    Tseng, Ching-Shiow
    Huang, Chiao-Chi
    Chen, Chen-San
    [J]. ROBOTICA, 2007, 25 : 375 - 383
  • [5] An Image-Guided Robot for Needle Insertion in Small Animal. Accurate Needle Positioning using Visual Servoing.
    Ayadi, A.
    Bayle, B.
    Graebling, P.
    Gangloff, J.
    [J]. 2008 IEEE/RSJ INTERNATIONAL CONFERENCE ON ROBOTS AND INTELLIGENT SYSTEMS, VOLS 1-3, CONFERENCE PROCEEDINGS, 2008, : 1453 - 1458
  • [6] Design and testing of a robotic system for MR image-guided interventions
    Christoforou, Eftychios G.
    Tsekos, Nikolaos V.
    Oezcan, Alpay
    [J]. JOURNAL OF INTELLIGENT & ROBOTIC SYSTEMS, 2006, 47 (02) : 175 - 196
  • [7] Design and Testing of a Robotic System for mr Image-guided Interventions
    Eftychios G. Christoforou
    Nikolaos V. Tsekos
    Alpay Özcan
    [J]. Journal of Intelligent and Robotic Systems, 2006, 47 : 175 - 196
  • [8] Registration of a needle-positioning robot to high-resolution 3-D ultrasound and computed tomography for image-guided interventions in small animals
    Waspe, Adam C.
    Lacefield, James C.
    Holdsworth, David W.
    Fenster, Aaron
    [J]. MEDICAL IMAGING 2008: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND MODELING, PTS 1 AND 2, 2008, 6918
  • [9] Image-guided robotic flexible needle steering
    Glozman, Daniel
    Shoham, Moshe
    [J]. IEEE TRANSACTIONS ON ROBOTICS, 2007, 23 (03) : 459 - 467
  • [10] Image-guided robotic interventions for prostate cancer
    Sridhar, Ashwin N.
    Hughes-Hallett, Archie
    Mayer, Erik K.
    Pratt, Philip J.
    Edwards, Philip J.
    Yang, Guang-Zhong
    Darzi, Ara W.
    Vale, Justin A.
    [J]. NATURE REVIEWS UROLOGY, 2013, 10 (08) : 452 - 462